Equinoxe 320-35-06Recalls & reported problems

Exactech, Inc. · Glenoid Plate · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862534903), received between Nov 2, 2021 and Apr 24, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Break1
  • Material Deformation1
  • Positioning Problem1

Patient problems

  • Failure of Implant2
  • Aseptic loosening1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2024Break; Material Deformation; Positioning Problem; Break; Material Deformation; Positioning Problem

    AS REPORTED, THE IMPLANT WAS OPENED AND LOADED WITH BONE GRAFT BY THE SURGICAL TECH BEFORE IMPACTION. DURING IMPACTION THE IMPLANT WOULD NOT FULLY SEAT AND WAS PULLED OUT TO REASSESS. UPON LOADING THE IMPLANT BACK ON THE INSERTER IT WAS NOTICED BY THE TECH THAT THE CAGE HAD WARPED. THE SURGEON THEN DECIDED TO WASTE THE IMPLANT AND PUT IN A NEW ONE. THE NEW BASEPLATE WAS IMPACTED WITH NO ISSUES. THIS EVENT IS NOT ASSOCIATED WITH A REVISION OF EXACTECH IMPLANTS. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177879 ↗
  • InjuryMay 23, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, APPROXIMATELY 2 YEARS POST OP INITIAL TSA, THIS 77 Y/O FEMALE PATIENT WAS REVISED DUE TO ASEPTIC GLENOID LOOSENING WITH INCREASED PAIN. SURGEON REVISED TO A HEMIARTHROPLASTY. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICES AND UNLIKELY RELATED TO PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES.

    Patient problems: Aseptic loosening; Failure of Implant; Pain

    MAUDE report 16987479 ↗
  • InjuryNov 2, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, APPROXIMATELY 1 YR. POSTOP THE IMPLANTED LTSA AND ENTERED THIS CLINICAL STUDY, THIS (B)(6) FEMALE PATIENT WHO HAD A HISTORY OF INFECTIONS IN PRIOR IMPLANTED DEVICES (MANUFACTURERS UNKNOWN) PRESENTED WITH ASEPTIC GLENOID LOOSENING. RADIOGRAPHS SHOWED ASEPTIC GLENOID LOOSENING AND THAT THE BONE GRAFT HAD NOT INCORPORATED. THE SURGEON OPTED TO DO A STAGED REVISION WITH BONE GRAFTING OF THE ILIAC CREST TO THE GLENOID FOLLOWED BY A REVISION TO A REVERSE TSA. THE FIRST PROCEDURE (REMOVAL OF COMPONENTS, BONE GRAFTING, AND PLACING ANTIBIOTIC SPACERS) WAS PERFORMED ON (B)(6) 2021. THE CONVERSION TO A REVERSE TSA WAS PERFORMED ON (B)(6) 2021. THE PATIENT WAS AN INPATIENT FOR 2 DAYS AND DISCHARGED HOME. THIS EVENT WAS REPORTED AS RESOLVED ON (B)(6) 2021. THE CASE REPORT FORM INDICATES THIS EVENT IS NOT RELATED TO DEVICES AND DEFINITELY RELATED TO THE PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES.

    Patient problems: Failure of Implant

    MAUDE report 12736810 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page