Skip to content

Equinoxe 320-35-07Recalls & reported problems

Exactech, Inc. · Glenoid Plate · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862534910), received between May 25, 2022 and Jan 12, 2026 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth4
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Mechanical Problem1
  • Use of Device Problem1

Patient problems

  • Pain3
  • Insufficient Information2
  • Bone Fracture(s)1
  • Failure of Implant1
  • Loss of Range of Motion1
  • Septic loosening1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 12, 2026Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A 71 YO FEMALE PATIENT, WHO HAD A LEFT RTSA, UNDERWENT A REVISION PROCEDURE APPROXIMATELY 4 YEARS 2 MONTHS POST THE INITIAL PROCEDURE. THE PATIENT COMPLAINED OF PAIN. THE BASEPLATE WAS OBSERVED TO BE DISLODGED FROM THE GLENOID ON CT. ONE OF THE SCREWS WAS OBSERVED TO BE BROKEN. THE HUMERAL LINER & TRAY WERE REMOVED. THE STEM WAS STILL WELL-FIXED DESPITE PROXIMAL HUMERAL BONE LOSS. THE GLENOSPHERE WAS REMOVED. THE BASEPLATE WITH ALL FOUR SCREWS WAS PULLED OUT AS ONE PIECE. ONE OF THE SCREWS WAS BROKEN. THE SURGEON FILLED THE BONE VOID WITH CAP & USED COMPETITOR¿S COMPONENTS FOR THE GLENOID BASEPLATE & GLENOSPHERE. THE SURGEON THEN IMPLANTED A NEW HUMERAL TRAY & LINER. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. AN X-RAY IMAGE WAS PROVIDED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN. IMAGES WERE PROVIDED. NO FURTHER INFORMATION. THIS IS 1 OF 2 REPORTS FOR THIS EVENT.

    Patient problems: Pain

    MAUDE report 24049022 ↗
  • InjurySep 11, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL REVERSE TOTAL SHOULDER REPLACEMENT ON THE LEFT SIDE. APPROXIMATELY 25 MONTHS POST-OP, THE PATIENT EXPERIENCED EARLY PAIN FROM POST-OP ALLODYNIA/PLEXOPATHY. SUBSEQUENTLY THE GLENOID AND HUMERUS BECAME SEPTICALLY LOOSE. AS A RESULT, APPROXIMATELY 30 MONTHS AFTER INITIAL REPLACEMENT, THE PATIENT UNDERWENT REVISION PROCEDURE. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Pain; Septic loosening

    MAUDE report 23027582 ↗
  • InjuryFeb 23, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, APPROXIMATELY 1 YEAR AND 5 MONTHS POST THE INITIAL LEFT TOTAL SHOULDER ARTHROPLASTY, THE PATIENT'S BASEPLATE LOOSENED AND THE PATIENT WAS REVISED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21442756 ↗
  • InjuryFeb 12, 2025Mechanical Problem; Use of Device Problem; Mechanical Problem; Use of Device Problem

    IT WAS REPORTED THAT THIS PATIENT'S LEFT KNEE WAS REVISED APPROXIMATELY 9 MONTHS POST INITIAL OPERATION. IT WAS STATED THAT THE IMPLANT HAD FAILED AND THE INFERIOR SCREW HAD BROKEN. UPON REVISION, FOUND OUT THAT THE IMPLANT WAS PUT IN UPSIDE DOWN. SURGEON REMOVED AS MUCH OF THE BROKEN SCREW AS HE COULD, BUT PART OF THE SCREW WAS STILL LEFT FRAGMENTED IN THE PATIENT. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Insufficient Information; Loss of Range of Motion; Pain

    MAUDE report 21364476 ↗
  • MalfunctionMar 19, 2023Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA CLINICAL STUDY, THAT INTRA-OP, THE 76 YO FEMALE PATIENT SUSTAINED A SCAPULAR FRACTURE OF THE TIP OF THE CORACOID THROUGH ANTERIOR SCREW, LIKELY FROM PULL OF ANTERIOR GLENOID RETRACTOR. THERE WAS NO ACTION TAKEN. THE PATIENT¿S OUTCOME WAS LAST KNOWN AS CONTINUING. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICE AND DEFINITELY RELATED TO PROCEDURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 16570566 ↗
  • InjuryMay 25, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, APPROXIMATELY 2 YEARS POST OP THE TSA, THIS (B)(6) FEMALE PATIENT WAS REVISED TO A STANDARD REVERSE DUE TO ASEPTIC GLENOID LOOSENING. BASEPLATE LOOSENED WHICH WAS FOLLOWED BY STEM LOOSENING. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICES AND/OR PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES.

    Patient problems: Failure of Implant

    MAUDE report 14498404 ↗

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page