Equinoxe 320-40-13Recalls & reported problems

Exactech, Inc. · Constrained Humeral Liner · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862535061), received between Nov 15, 2024 and Sep 25, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 6, 2024Open, Classified

    The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

    Z-1421-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn2
  • Detachment of Device or Device Component1
  • Unstable1

Patient problems

  • Joint Laxity2
  • Bone Notching1
  • Joint Dislocation1
  • Joint Impingement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2025Detachment of Device or Device Component; Naturally Worn; Detachment of Device or Device Component; Naturally Worn

    AS REPORTED, APPROXIMATELY 2 YEARS POST THE INITIAL RIGHT REVERSE TOTAL SHOULDER ARTHROPLASTY, THE PATIENT WAS REVISED DUE TO DISLOCATION. THE PATIENT INDICATED NO TRAUMA OCCURRED. THE SURGEON OBSERVED THE POLYETHYLENE COMPONENT DISSOCIATED FROM THE TRAY AND FLIPPED OVER. NO ABNORMAL WEAR WAS NOTED ON THE FACE OF THE POLYETHYLENE COMPONENT THAT MIGHT AFFECT THE LOCKING MECHANISM OF THE POLYETHYLENE LINER TO THE TRAY. THE SURGEON REPORTED THERE COULD HAVE POSSIBLY BEEN AN ISSUE WITH THE LOCKING MECHANISM OF THE POLY. THE SURGEON REPORTED IT WAS ALSO POSSIBLE THE POLY WASN'T IMPACTED FIRMLY ENOUGH ON THE INITIAL SURGERY. THE PATIENT WAS REVISED TO 40 +2.5 CONSTRAINED LINER. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO MEDICAL OR SURGICAL INTERVENTIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Joint Dislocation

    MAUDE report 23148676 ↗
  • InjuryAug 4, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL REVERSE TOTAL SHOULDER REPLACEMENT ON THE LEFT SIDE. SUBSEQUENTLY, THE PATIENT EXPERIENCED INSTABILITY AND BONE NOTCHING. AS A RESULT, APPROXIMATELY TWO YEARS AND THREE-MONTHS POST-SURGERY, THE PATIENT UNDERWENT A REVISION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS REPORT 1 OF 2 FOR THIS EVENT/PATIENT.

    Patient problems: Bone Notching; Joint Impingement; Joint Laxity

    MAUDE report 22698945 ↗
  • InjuryNov 15, 2024Unstable; Unstable

    AS REPORTED, THE PATIENT UNDERWENT A THIRD RIGHT SIDE SHOULDER REVISION SURGERY DUE TO INSTABILITY AFTER HER REVERSE REPLACEMENT. THE HUMERAL TRAY WAS REVISED FROM A 0 TO A +10. THE POLY WAS ALSO REVISED TO A CONSTRAINED LINER. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Joint Laxity

    MAUDE report 20694905 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →