Skip to content

Equinoxe 321-00-02Recalls & reported problems

Exactech, Inc. · Class 2

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862541246), received between May 14, 2021 and May 14, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. May 11, 2021Completed

    The HAT trial contains a retaining ring that has shown the potential to disengage during use.

    Z-1977-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2021Break; Break

    AS REPORTED, DURING THE PROCEDURE, THE +0 LATERALIZED TRAY THREADED ONTO THE FRACTURE STEM DURING TRAILING. DURING REMOVAL, THE RETAINING RING INSIDE THE TRIAL POPPED OFF AND FELL INTO THE PATIENT WOUND SITE. THE RETAINING RING WAS FOUND AND REMOVED FROM THE PATIENT WOUND SITE, BUT THE SURGEON COULD NOT RE-THREAD THE TRAIL WITHOUT THE RETAINING RING. THE SURGEON THEN TRAILED +5 LATERALIZED TRIAL AND WHEN TRYING TO THREAD TRIAL ONTO STEM THE RETAINING RING AGAIN POPPED OFF AND WOULD NOT THREAD ONTO TRIAL. THE SURGEON DECIDED TO STOP USING THE HAT SYSTEM AND USED FRACTURE STEM INSTEAD. ALL PARTS WERE RETRIEVED FROM THE PATIENT WOUND SITE AND WERE ACCOUNTED FOR. THERE WAS A 5-15 MINUTE DELAY AND THERE WERE NO ADVERSE EVENT RESULTING FORM THE DELAY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICE ARE RETURNING FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11828846 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page