Skip to content

Equinoxe 321-00-12Recalls & reported problems

Exactech, Inc. · Class 2

All models of Equinoxe (303)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862541260), received between May 14, 2021 and May 14, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. May 11, 2021Class IICompleted

    The HAT trial contains a retaining ring that has shown the potential to disengage during use.

    Z-1977-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2021Break; Break

    AS REPORTED, DURING THE PROCEDURE, THE +0 LATERALIZED TRAY THREADED ONTO THE FRACTURE STEM DURING TRAILING. DURING REMOVAL, THE RETAINING RING INSIDE THE TRIAL POPPED OFF AND FELL INTO THE PATIENT WOUND SITE. THE RETAINING RING WAS FOUND AND REMOVED FROM THE PATIENT WOUND SITE, BUT THE SURGEON COULD NOT RE-THREAD THE TRAIL WITHOUT THE RETAINING RING. THE SURGEON THEN TRAILED +5 LATERALIZED TRIAL AND WHEN TRYING TO THREAD TRIAL ONTO STEM THE RETAINING RING AGAIN POPPED OFF AND WOULD NOT THREAD ONTO TRIAL. THE SURGEON DECIDED TO STOP USING THE HAT SYSTEM AND USED FRACTURE STEM INSTEAD. ALL PARTS WERE RETRIEVED FROM THE PATIENT WOUND SITE AND WERE ACCOUNTED FOR. THERE WAS A 5-15 MINUTE DELAY AND THERE WERE NO ADVERSE EVENT RESULTING FORM THE DELAY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICE ARE RETURNING FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11828850 ↗

FDA clearance and classification

Cleared through 510(k) K191561 on Aug 20, 2019 (decision: Substantially Equivalent). Submitted by Exactech, Inc. as “Exactech Equinoxe Humeral Augmented Trays”. The FDA review took 68 days. FDA record ↗

The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.

Common questions about the Equinoxe 321-00-12

Has the Equinoxe 321-00-12 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the company has completed it but the FDA has not yet closed it. The recall, Z-1977-2021 (Class II), began on May 11, 2021: The HAT trial contains a retaining ring that has shown the potential to disengage during use. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Equinoxe 321-00-12?

The FDA's MAUDE database lists 1 adverse event report for this device, received on May 14, 2021. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Equinoxe 321-00-12?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Equinoxe 321-00-12 FDA approved?

It was cleared through 510(k) K191561 on Aug 20, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Equinoxe 321-00-12, and is it still sold?

The FDA's device database lists Exactech, Inc. as the company responsible for it (brand Equinoxe, model 321-00-12). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page