Equinoxe® 322-38-00Recalls & reported problems

Exactech, Inc. · Humeral Liner · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862593832), received between Feb 5, 2025 and Sep 8, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2025Class IIOpen, Classified

    Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.

    Z-2562-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated2

Patient problems

  • Joint Dislocation2
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 8, 2026Device Dislodged or Dislocated

    It was reported that the patient underwent a left shoulder replacement. Subsequently, the patient experienced a dislocation. As a result, a revision was performed during which the surgeon exchanged the existing liner for a thicker constrained liner. The patient was last known to be in stable condition following the event. No further information is available.

    Patient problems: Joint Dislocation

    MAUDE report 50670421 ↗
  • InjuryFeb 5, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A 70 YO FEMALE PATIENT, INITIAL REVERSE LEFT SHOULDER IMPLANTED IN (B)(6) 2024, UNDERWENT A REVISION PROCEDURE IN (B)(6) 2025, APPROXIMATELY 3 MONTHS POST THE INITIAL PROCEDURE. THE SURGEON EXPLANTED THE GLENOSPHERE, TRAY AND LINER AND WENT BACK IN WITH NEW ONES. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGES DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. X-RAYS WERE PROVIDED. THE EXPLANTED DEVICES ARE AVAILABLE FOR RETURN. A DEVICE IMAGE WAS PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Insufficient Information; Joint Dislocation

    MAUDE report 21316008 ↗

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →