Skip to content

Equinoxe® 322-38-13Recalls & reported problems

Exactech, Inc. · Humeral Liner, Constrained · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862593863), received between Sep 15, 2025 and Sep 15, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Naturally Worn” (1 report).

1 other recall cover devices cleared under the same FDA submission (K223833); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn1

Patient problems

  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 15, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT A 80 YO FEMALE PATIENT WHO HAD AN INITIAL RIGHT SHOULDER IMPLANTED, UNDERWENT A REVISION PROCEDURE APPROXIMATELY 9 MONTHS POST THE INITIAL PROCEDURE. THE PATIENT PRESENTED WITH PAIN. THEY HAD AN INFECTION. AN ANTIBIOTIC SPACER AND BEADS WERE INSERTED. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGES DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE PROVIDED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN DUE TO CHAIN OF COMMAND. DEVICE IMAGE WAS PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 23052969 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K223833. One submission often covers many models, so check the product and lot details in each record.

  • Z-2562-2025 · Aug 15, 2025 · Open, Classified
    Exactech Equinoxe, Reverse Shoulder Humeral Liners, 322-38-00 (145-DEG PE 38MM HUM LINER +0), REF: 322-38-03 (145-DEG PE 38MM HUM LINER +2.5), REF:322-42-00 (145-DEG PE 42MM HUM L

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page