Skip to content

Equinoxe® 322-40-03Recalls & reported problems

Exactech, Inc. · Humeral Liner · Class 2 · Implant

All models of Equinoxe (303)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862593887), received between Nov 5, 2024 and Mar 6, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

1 other recall cover devices cleared under the same FDA submission (K223833); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Insufficient Information2
  • Joint Laxity2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2025Unstable; Unstable

    IT WAS REPORTED THAT A FEMALE PATIENT, INITIAL RIGHT SHOULDER IMPLANTED IN (B)(6) 2024, UNDERWENT A REVISION PROCEDURE. DATE UNKNOWN. THE SURGEON REVISED THE PATIENT¿S REVERSE SHOULDER DUE TO INSTABILITY. HE REMOVED A 40+0 LINER AND REPLACED IT WITH A 40+2.5 CONSTRAINED LINER. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGES DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. X-RAYS WERE UNABLE TO BE OBTAINED. THE EXPLANTED DEVICE IS NOT AVAILABLE FOR RETURN AS THE HOSPITAL KEPT THE DEVICE. NO DEVICE IMAGES WERE OBTAINED. NO FURTHER INFORMATION.

    Patient problems: Insufficient Information; Joint Laxity

    MAUDE report 21539622 ↗
  • InjuryNov 5, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THIS PATIENT'S RIGHT SHOULDER WAS REVISED. THE PATIENT DESCRIBED FEELING ROCKING IN THE SHOULDER WITH RANGE OF MOTION. SURGEON REACHED OUT TO REQUEST INSTRUMENTS FOR EXTRACTION AND REVISION OF CURRENT IMPLANT. AXIAL VIEW DEMONSTRATED A MISALIGNED GLENOSHPERE WHICH APPEARED TO BE SITTING OFF THE BASEPLATE. NO GLENOSPHERE LOCKING SCREW WAS VISUALIZED ON X-RAY. CORONAL VIEW APPEARED NORMAL. UPON EXPOSURE OF THE JOINT, THE GLENOSPHERE WAS LOOSE AND THE LOCKING SCREW APPEARED TO BE BROKEN. SURGEON INSPECTED BASEPLATE AND NOTICED SCREW WAS BROKEN INTO THE CENTER HOLE, WHICH IT INSERTS INTO. SURGEON PROCEEDED TO EXTRACT LOCKING CAPS AND SCREWS USING THE HEX DRIVER AND SUCCESSFULLY EXTRACTED WITHOUT SCREWS BREAKING. ONCE SCREWS WHERE TAKEN OUT OF THE PATIENT, ANOTHER SURGEON ASSISTING ON THE CASE, EXTRACTED THE BASEPLATE VIA HOSPITAL INSTRUMENTS. DOCTORS TOOK THE LINER OFF VIA MALLET AND ELEVATOR, THEN REMOVED TORQUE SCREW WITH REMOVAL TOOL TO TAKE THE TRAY OFF. STEM WAS WELL FIXED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PRODUCT NOT RETURNING: PLEASE ASK HOSPITAL FOR RETURN OF PRODUCTS.

    Patient problems: Insufficient Information; Joint Laxity

    MAUDE report 20612253 ↗

FDA clearance and classification

Cleared through 510(k) K223833 on Sep 14, 2023 (decision: Substantially Equivalent). Submitted by Exactech, Inc. as “Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter Trays”. The FDA review took 266 days. FDA record ↗

The FDA regulates product code KWT (prosthesis, shoulder, non-constrained, metal/polymer cemented) under 21 CFR 888.3650 as a Class II device. 21 CFR 888.3650 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K223833. One submission often covers many models, so check the product and lot details in each record.

  • Z-2562-2025 · Aug 15, 2025 · Open, Classified
    Exactech Equinoxe, Reverse Shoulder Humeral Liners, 322-38-00 (145-DEG PE 38MM HUM LINER +0), REF: 322-38-03 (145-DEG PE 38MM HUM LINER +2.5), REF:322-42-00 (145-DEG PE 42MM HUM L

Common questions about the Equinoxe® 322-40-03

Has the Equinoxe® 322-40-03 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10885862593887), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the Equinoxe® 322-40-03?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Nov 5, 2024 and Mar 6, 2025. In those reports, the most common specific device problems are “Loosening of Implant Not Related to Bone-Ingrowth” (1), “Naturally Worn” (1) and “Unstable” (1); the most common patient problem is “Joint Laxity” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Equinoxe® 322-40-03?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Equinoxe® 322-40-03 FDA approved?

It was cleared through 510(k) K223833 on Sep 14, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Equinoxe® 322-40-03 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Equinoxe® 322-40-03, and is it still sold?

The FDA's device database lists Exactech, Inc. as the company responsible for it (brand Equinoxe®, model 322-40-03). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page