Skip to content

Equinoxe® 320-55-98Recalls & reported problems

Exactech, Inc. · Central Screw Baseplate One-Lock Kit · Class 2 · Implant

All models of Equinoxe (303)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862630384), received between Jun 1, 2026 and Jun 1, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Joint Dislocation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT A PATIENT, WHO HAD A LEFT RTSA, UNDERWENT A REVISION PROCEDURE. FOLLOWING THE SURGERY ALL WAS WELL WITH THE PATIENT. ALMOST A MONTH AFTER THE INITIAL SURGERY, SHE REACHED BACK TO PUT THE LEGS UP ON HER RECLINER AND FELT A POP. X-RAYS CONFIRMED THE GLENOSPHERE HAD MOVED AND WASN'T FULLY SEATED. THE LOOSE GLENOSPHERE WAS REMOVED ALONG WITH THE LOCKING PLATE, CSB GLENOSPHERE LOCKING SCREW AND LINER. WE THEN REVISED WITH A NEW 36 GLENOSPHERE, CSB GLENOSPHERE SCREW, LOCKING PLATE/RING AND 36 LINER. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. AN X-RAY WAS PROVIDED. THE EXPLANTED DEVICES ARE NOT ABLE TO BE RETURNED. THE HOSPITAL DOES NOT ALLOW REMOVAL. DEVICE IMAGES WERE PROVIDED. NO FURTHER INFORMATION. THIS IS 1 OF 2 REPORTS

    Patient problems: Joint Dislocation

    MAUDE report 25352853 ↗

FDA clearance and classification

Cleared through 510(k) K233482 on Jul 18, 2024 (decision: Substantially Equivalent). Submitted by Exactech, Inc. as “Equinoxe® Central Screw Baseplate System”. The FDA review took 266 days. FDA record ↗

The FDA regulates product code KWT (prosthesis, shoulder, non-constrained, metal/polymer cemented) under 21 CFR 888.3650 as a Class II device. 21 CFR 888.3650 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Equinoxe® 320-55-98

Has the Equinoxe® 320-55-98 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10885862630384), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Equinoxe® 320-55-98?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 1, 2026. That report lists device problem “Detachment of Device or Device Component” (1); patient problem “Joint Dislocation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Equinoxe® 320-55-98?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Equinoxe® 320-55-98 FDA approved?

It was cleared through 510(k) K233482 on Jul 18, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Equinoxe® 320-55-98 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Equinoxe® 320-55-98, and is it still sold?

The FDA's device database lists Exactech, Inc. as the company responsible for it (brand Equinoxe®, model 320-55-98). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page