Skip to content

RAPIDTRANSIT 601-251XRecalls & reported problems

Medos International Sàrl · RAPIDTRANSIT Infusion Catheter with 2 Markers Infusion Catheter Outer Diameter (Distal) 2.3F (0.75 mm) Usable Length 150 cm 2.8F (0.95 mm) .021 Inch (0.5 mm) 3cm 50cm Contents: 1 Infusion Catheter and 1 Shape Mandrel. Minimum Guiding Catheter I.D. / Dia. .042 Inch (1.1 mm) · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704028352), received between Mar 6, 2017 and Sep 14, 2018 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Physical Resistance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Physical Resistance1
  • Physical Resistance/Sticking1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 14, 2018Physical Resistance/Sticking; Physical Resistance/Sticking

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A STENT-ASSISTED COIL EMBOLIZATION OF A RECURRENT LEFT OPHTHALMIC ARTERY ANEURYSM, THE 4MM X 30MM ENTERPRISE 2 NO TIP (ENF403000/10689189) STENT WAS LOADED, FLUSHED, AND ADVANCED THROUGH THE 0.021¿ RAPID TRANSIT 150CM (601251X/30042646) DUAL MARKER MICROCATHETER, BUT RESISTANCE WAS ENCOUNTERED ABOUT HALFWAY THROUGH THE MICROCATHETER AND THE STENT COULD NO LONGER BE ADVANCED. THE STENT BECAME STUCK IN THE MICROCATHETER HUB DURING REMOVAL; THEREFORE, THE MICROCATHETER WAS REMOVED FROM THE PATIENT. THE MICROCATHETER AND STENT WERE REPLACED, AND TARGET POSITION WAS EASILY RE-OBTAINED. THE ENTERPRISE 2 STENT WAS IMPLANTED ACROSS THE NECK OF THE ANEURYSM AND FOUR COMPETITOR COILS WERE PLACED IN THE ANEURYSM. POST ANGIOGRAPHY REVEALED GOOD RESULT AND THE CASE WAS CONCLUDED. NO PATIENT COMPLICATIONS OCCURRED AS A RESULT OF THE EVENT. THE SURGERY WAS DELAYED 5-10 MINUTES DUE TO THE EVENT. THE PATIENT DID WELL POST-OP AND WAS DISCHARGED THE FOLLOWING DAY. THE RIGHT FEMORAL ARTERY WAS ACCESSED WITH A 6F SHEATH. THE TARGET ANEURYSM WAS PREVIOUSLY COILED WITH COMPETITOR COILS ONE YEAR PRIOR. THE VESSEL TORTUOSITY WAS LOW. NO VISIBLE PRODUCT DAMAG

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7876803 ↗
  • MalfunctionMar 6, 2017Physical Resistance; Physical Resistance

    THE CONTACT FROM THE FACILITY REPORTED THAT THERE WAS STRONG RESISTANCE FELT WHEN ADVANCING THE COIL THROUGH THE MIDDLE PORTION OF THE TRANSIT MICROCATHETER (601251X/17536303.). THE PROCEDURE WAS THE PRE STENT-GRAFTING COIL EMBOLIZATION OF AN ANEURYSM AT THE INTERNAL ILIAC ARTERY. THE PATIENT¿S INFORMATION WAS UNKNOWN. THE PATIENT¿S VESSEL LEVEL OF TORTUOUSNESS AND CALCIFICATION WERE UNKNOWN. NO REPORT OF THE OTHER DEVICE USED IN THE PROCEDURE IS AVAILABLE. THE TRANSIT (COMPLAINT PRODUCT) WAS ADVANCED TO THE SITE OF EMBOLIZATION. THE COIL (PUSHABLE COIL, DETAILS UNKNOWN) WAS INSERTED INTO THE CATHETER BUT THERE WAS STRONG RESISTANCE FELT AT THE MIDDLE PORTION. THE COIL WAS SUCCESSFULLY INSERTED SOMEHOW BUT THE PHYSICIAN FELT THAT IT MAY NOT BE SAFE TO USE. THE CATHETER WAS REPLACED WITH A COMPETITOR MICROCATHETER (DETAILS UNKNOWN.) THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT FURTHER ISSUES OR DELAY. THERE WAS ALSO NO PATIENT INJURY / COMPLICATION REPORTED. THE COMPLAINT PRODUCT WAS NEW AND STORED PER LABELING INSTRUCTIONS. THE PROCEDURE WAS CONDUCTED IN ACCORDANCE WITH THE IFU AND A CONSTANT FLUSH WAS MAINTAINED AT ALL TIMES. THERE WAS NO DELAY GREATER THAN 30 MINUTES DUE TO T

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6382983 ↗

FDA clearance and classification

Cleared through 510(k) K972518 on Nov 4, 1997 (decision: Substantially Equivalent). Submitted by Cordis Neurovascular, Inc. as “Transit/Rapidtransit/Speedster/Prowler Infusion Catheter”. The FDA review took 120 days. FDA record ↗

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the RAPIDTRANSIT 601-251X

Has the RAPIDTRANSIT 601-251X been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886704028352), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the RAPIDTRANSIT 601-251X?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Mar 6, 2017 and Sep 14, 2018. In those reports, the most common specific device problems are “Physical Resistance” (1) and “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the RAPIDTRANSIT 601-251X?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the RAPIDTRANSIT 601-251X FDA approved?

It was cleared through 510(k) K972518 on Nov 4, 1997. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the RAPIDTRANSIT 601-251X, and is it still sold?

The FDA's device database lists Medos International Sàrl as the company responsible for it (brand RAPIDTRANSIT, model 601-251X). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page