Skip to content

PROWLER SELECT 606-S155JXRecalls & reported problems

Medos International Sàrl · PROWLER SELECT LP ES TIP MARKERS 2 TIPSHAPE J Infusion Catheter 2.3F (0.75mm) 1.9F (0.65mm) .0165 Inch (0.4mm) 5cm 150cm Contents: 1 Infusion Catheter and 1 Shape Mandrel. Minimum Guiding Catheter I.D. / Dia. .035 Inch (.9mm) · Class 2

All models of PROWLER SELECT (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704028789), received between Jul 30, 2018 and Mar 2, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K021591); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Crack1
  • Physical Resistance/Sticking1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2020Crack; Crack

    THE HEALTHCARE PROFESSIONAL REPORTED THAT THE TIP OF THE 150CM X 5CM, 2 MARKERS, J TIP PROWLER SELECT LPES MICROCATHETER (606S155JX / 30187163) WAS CRACKED. THERE WAS NO REPORT OF ANY PATIENT INJURY OR COMPLICATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9775775 ↗
  • MalfunctionJul 30, 2018Physical Resistance/Sticking; Physical Resistance/Sticking

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE COIL EMBOLIZATION PROCEDURE TARGETING AN INTERNAL CAROTID ARTERY (ICA) ANEURYSM, THE ORBIT GALAXY COMPLEX XTRASOFT COIL 2MM X 2CM (640CX0202 / 30006827) COULD NOT BE ADVANCED THROUGH THE PROWLER SELECT LPES, 150CM X 5CM, 2 MARKERS, J TIP MICROCATHETER (606S155JX / 17680649). THERE WAS NO KINK OR BENT OBSERVED ON THE MICROCATHETER PRIOR TO THE REPORTED IMPEDED ISSUE ENCOUNTERED WITH THE MICROCATHETER. THE DEVICE WAS PREPPED AND USED PER THE INSTRUCTION FOR USE (IFU) WITH ADEQUATE AND CONTINUOUS FLUSH MAINTAINED THROUGH THE CATHETER. THE PHYSICIAN REPLACED BOTH THE COIL AND THE MICROCATHETER AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WAS NO PROCEDURAL DELAY AND THERE WAS NO REPORT OF ANY PATIENT CONSEQUENCE OR ADVERSE EVENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7733329 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K021591. One submission often covers many models, so check the product and lot details in each record.

  • Z-1482-06 · Aug 11, 2006 · Terminated
    Cordis Neurovascular Pre-Shaped PROWLER Infusion Catheters.
  • Z-1483-06 · Aug 11, 2006 · Terminated
    Cordis Neurovascular Pre-Shaped Prowler Select Infusion Catheters.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page