PROWLER SELECT 606-S255MXRecalls & reported problems
Medos International Sàrl · PROWLER SELECT PLUS TIP MARKERS 2 TIPSHAPE 90 Infusion Catheter 2.8F (0.95mm) 2.3F (0.75mm) .021 Inch (0.5mm) 5cm 150cm Contents: 1 Infusion Catheter and 1 Shape Mandrel. Minimum Guiding Catheter I.D. / Dia. .042 Inch (1.1mm) · Class 2
All models of PROWLER SELECT (9)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704028871), received between Dec 30, 2016 and Jul 21, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Physical Resistance” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K021591); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Physical Resistance1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 21, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
AS REPORTED BY THE FIELD, A PROWLER SELECT PLUS 5CM 90 TIP MICROCATHETER (606S255MX, 30118145) WAS USED TO IMPLANT A PULSE RIDER (PRODUCT CODE/ LOT NUMBER UNKNOWN). HOWEVER, RESISTANCE BECAME STRONGER AND STRONGER AFTER THE PULSERIDER REACHED THE HUB. IT GOT STUCK ON THE 30 CM FROM THE DISTAL END. THE PULSERIDER WAS REMOVED AND THE COMPLAINT MICROCATHETER (MC) WAS FLUSHED. THE PHYSICIAN TRIED AGAIN BUT THE SAME ISSUE CONTINUED. THE COMPLAINT MICROCATHETER WAS REPLACED WITH ANOTHER MICROCATHETER. THE PULSERIDER WAS ABLE TO BE SMOOTHLY DELIVERED THROUGH THE REPLACED DEVICE AND IMPLANTED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WAS NO EVIDENCE OF PHYSICAL MATERIAL WITHIN THE DEVICE. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE PATIENT. THERE WAS NO DAMAGE NOTED ON THE MC DURING MANIPULATION OR UPON REMOVAL FROM THE PATIENT. A CONTINUOUS FLUSH ON THE MICROCATHETER WAS MAINTAINED. NO EXCESSIVE FORCE WAS APPLIED TO THE DEVICE. THE TARGET LESION BEING TREATED WAS THE BASILAR ARTERY TOP (BA). NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12203527 ↗MalfunctionDec 30, 2016Physical Resistance; Physical Resistance
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING COIL EMBOLIZATION FOR THE DIREMPTION OF THE INTERNAL CAROTID ARTERY, AN ENTERPRISE (ENF403900/10571456) WAS DELIVERED INTO THE PROWLER SELECT PLUS (606S255MX/16066293), BUT IT GOT STUCK AT APPROXIMATELY 15CM FROM ITS TIP. THE ENTERPRISE WAS REMOVED FROM THE MICROCATHETER AND THE CATHETER WAS REPLACED WITH ANOTHER ONE. THE SAME ENTERPRISE WAS USED SUCCESSFULLY AFTER REPLACED THE MICROCATHETER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT FURTHER ISSUES; HOWEVER, DUE TO THE EVENT IT WAS DELAYED FOR 30 MINUTES. THERE WAS ALSO NO PATIENT INJURY / COMPLICATION REPORTED. THE COMPLAINT PRODUCT WAS NEW AND STORED PER LABELING INSTRUCTIONS. THE PROCEDURE WAS CONDUCTED IN ACCORDANCE WITH THE IFU AND THE CONSTANT FLUSH HAD BEEN MAINTAINED AT ALL TIME. NO VISIBLE DEFECT/DAMAGE WAS NOTED ON THE PRODUCTS PRIOR TO OR AFTER THE EVENT. IT WAS REPORTED THAT THE PRODUCT WILL BE RETURNED FOR INVESTIGATION. THE PATIENT¿S VESSEL WAS MODERATELY TORTUOUS AND MODERATELY CALCIFIED. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6215893 ↗
FDA clearance and classification
Cleared through Special 510(k) K021591 on May 22, 2002 (decision: Substantially Equivalent). Submitted by Cordis Neurovascular, Inc. as “Prowler Select (10, 14, And Plus) Infusion Catheters With And Without Pre-Shaped Tips”. The FDA review took 7 days. FDA record ↗
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K021591. One submission often covers many models, so check the product and lot details in each record.
Common questions about the PROWLER SELECT 606-S255MX
Has the PROWLER SELECT 606-S255MX been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886704028871), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the PROWLER SELECT 606-S255MX?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 30, 2016 and Jul 21, 2021. In those reports, the most common specific device problems are “Physical Resistance” (1) and “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PROWLER SELECT 606-S255MX?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PROWLER SELECT 606-S255MX FDA approved?
It was cleared through Special 510(k) K021591 on May 22, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the PROWLER SELECT 606-S255MX, and is it still sold?
The FDA's device database lists Medos International Sàrl as the company responsible for it (brand PROWLER SELECT, model 606-S255MX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page