Medos International Sàrl · PROWLER SELECT PLUS TIP MARKERS 2 TIPSHAPE STRAIGHT Infusion Catheter 2.8F (0.95mm) 2.3F (0.75mm) .021 Inch (0.5mm) 15cm 170cm Contents: 1 Infusion Catheter and 1 Shape Mandrel. Minimum Guiding Catheter I.D. / Dia. .042 Inch (1.1mm) · Class 2

All models of PROWLER SELECT (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886704028895), received between Oct 1, 2019 and Jun 30, 2023 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K021591); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Contamination /Decontamination Problem1
  • Peeled/Delaminated1
  • Physical Resistance/Sticking1

Patient problems

  • No Consequences Or Impact To Patient2
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE EVENT WAS REPORTED VIA THE EXCELLENT STUDY, THE 60-YEAR-OLD FEMALE PATIENT WITH A HISTORY OF CONGESTIVE HEART FAILURE AND HYPERTENSION PRESENTED WITH AN UNWITNESSED, NON-WAKE-UP STROKE ON (B)(6) 2023. SHE WAS LAST SEEN WELL THE DAY BEFORE, ON (B)(6) 2023. THE PATIENT PRESENTED TO THE TREATING HOSPITAL ON (B)(6) 2023. INTRAVENOUS TISSUE PLASMINOGEN ACTIVATOR (TPA) WAS NOT ADMINISTERED AT THE TIME OF STROKE PRESENTATION. THE SUSPECTED ORIGINAL OF THE EMBOLISM DOCUMENTED IN THE CASE REPORT FORM (CRF) IS OF ¿UNDETERMINED ETIOLOGIES.¿ THE PATIENT¿S BASELINE A NIH STROKE SCALE (NIHSS) SCORE WAS 16 AND A MODIFIED RANKIN SCALE (MRS) SCORE OF 0. ON (B)(6) 2023, THE PATIENT UNDERWENT AN ENDOVASCULAR MECHANICAL THROMBECTOMY. A 5MM X 22MM EMBOTRAP III REVASCULARIZATION DEVICE (ET307522 / 22L259AV) WAS USED WITH A 170CM X 15CM 2 MARKERS PROWLER SELECT PLUS MICROCATHETER (606S272X / 30561949), A FLOWGATE2¿ BALLOON GUIDE CATHETER (STRYKER), AND AN AXS CATALYST® 6 DISTAL ACCESS CATHETER (STRYKER). THE PRE-PASS MODIFIED THROMBOLYSIS IN CEREBRAL INFARCTION (MTICI) SCORE WAS 0. THE FIRST PASS TARGETING AN OCCLUSION AT THE TERMINAL BIFURCATION OF THE LEFT INTERNAL CAROTID ARTERY (ICA) ¿ ¿CAROTID

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17241111 ↗
  • MalfunctionMar 28, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE FIELD, AT THE END OF THE PROCEDURE, A PROWLER SELECT PLUS 170/15 CM MICROCATHETER ( 606S272X, 30858782) SHOWED A GEL-LIKE SUBSTANCE AT THE PROXIMAL END OF THE HEMOSTATIC VALVE. THERE WAS NO PATIENT INJURY REPORTED. IN ORDER TO FURTHER INVESTIGATE THIS MATTER, THE CUSTOMER HAD THE PROWLER SELECT PLUS MICROCATHETER, THE ASSOCIATED HEMOSTATIC VALVES, AND A BRAND-NEW CATHETER EXAMINED IN A SUITABLE INSTITUTE. ADDITIONAL INFORMATION RECEIVED OF THE DEVICE INVOLVED CONCLUDED THE FOLLOWING: THE FT-IR SPECTRA OF THE DRYING RESIDUES OF THE CLOUDY APPEARING LIQUIDS AT THE END OF THE VALVE OF THE USED MICRO CATHETER SHOW A SIGNIFICANT ACCORDANCE TO THE SPECTRUM OF THE OUTER SURFACE OF THE DISTAL END AS WELL AS THE ONE OF THE OUTER SURFACE OF THE LIGHT BLUE COLOURED PART OF THE UNUSED CATHETER (P1). THE SPECTRA ALSO SHOW A SIGNIFICANT ACCORDANCE TO DATABASE SPECTRA OF POLYVINYLPYRROLIDONE. THE RESIDUES IN THE SAMPLE PROBABLY CONSIST OF POLYVINYLPYRROLIDONE. THE FT-IR SPECTRUM OF THE WHITISH PARTICLE IN THE LIQUID AT THE END OF THE VALVE OF THE SAMPLE SHOWS A VERY HIGH ACCORDANCE TO DATABASE SPECTRA OF DIFFERENT PROTEINS. THE PARTICLE IS VERY LIKELY FORMED BY PROTEIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629159 ↗
  • MalfunctionSep 8, 2020Peeled/Delaminated; Peeled/Delaminated

    AS REPORTED BY THE FIELD, DURING A MECHANICAL THROMBECTOMY OF THE LEFT CEREBRAL MEDIA ARTERY, THE INNER LINER OF A 170/15 CM PROWLER SELECT PLUS (606S272X, UNKNOWN LOT) WAS ¿RUBBED OFF¿ AFTER TWO MANEUVERS WITH STENT RETRIEVERS (FIRST, EMBOTRAP 2 5X33MM AND SECOND, SOLITAIRE X 4X40MM). THE MICROCATHETER WAS REPLACED, NOT BY A PROWLER SELECT PLUS. THE SURGERY WAS DELAYED BY THREE MINUTES DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCES REPORTED. THERE WAS NO PERFORMANCE ISSUES WITH THE EMBOTRAP. A CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE MICROCATHETER. THERE WAS RESISTANCE BETWEEN THE MICROCATHETER AND THE STENT RETRIEVERS AFTER THE INNER LINER WAS GONE. THE FRICTION BETWEEN THE STENT RETRIEVER AND THE MICROCATHETER INCREASED. THAT WAS THE CAUSE OF THE MICROCATHETER BEING REPLACED. THERE WAS NO REPORT OF ANY COATING FLAKES ENDING UP IN THE PATIENT. ADDITIONAL INTERVENTION WAS NOT NEEDED TO REMOVE THE DEVICE FROM THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10503583 ↗
  • MalfunctionOct 1, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A MECHANICAL THROMBECTOMY IT WAS NOT POSSIBLE TO INSERT A 5X33 EMBOTRAP 2 (ET007533, 18M017AV) INTO A 170CM PROWLER SELECT PLUS MICROCATHETER (606S272X, 30066730). INSTEAD OF SMOOTH INSERTION AS USUAL THE EMBOTRAP 2 WAS DESTROYED DURING ATTEMPT. A MEDTRONIC SOLITAIRE MICROCATHETER (MC) WAS USED INSTEAD BUT PROBLEM WAS PERSISTENT, AND THE SOLITAIRE MC WAS ALSO DESTROYED. A NEW (170CM) PROWLER SELECT PLUS (606S272X, 30066730) WAS USED AND PLACED IN THE TARGET VESSEL. A SECOND 5X33 EMBOTRAP 2 (ET007533, 18M017AV) WAS DESTROYED DUE TO THE SAME PROBLEM. EVENTUALLY IT WAS WITH DIFFICULTIES THAT IT WAS POSSIBLE TO INSERT A PRESET (PHENOX) THROMBECTOMY DEVICE INTO THE PROWLER SELECT PLUS. THE PROCEDURE WAS FINISHED WITH THIS SETUP. THERE WAS A SURGERY DELAYED OF 30 MINUTES DUE TO THE REPORTED EVENTS. IT IS UNKNOW IF THE PATIENT¿S CONDITION WORSENED AS A RESULT OF THE SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9140871 ↗

FDA clearance and classification

Cleared through Special 510(k) K021591 on May 22, 2002 (decision: Substantially Equivalent). Submitted by Cordis Neurovascular, Inc. as “Prowler Select (10, 14, And Plus) Infusion Catheters With And Without Pre-Shaped Tips”. The FDA review took 7 days. FDA record ↗

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K021591. One submission often covers many models, so check the product and lot details in each record.

  • Z-1482-06 · Aug 11, 2006 · Terminated
    Cordis Neurovascular Pre-Shaped PROWLER Infusion Catheters.
  • Z-1483-06 · Aug 11, 2006 · Terminated
    Cordis Neurovascular Pre-Shaped Prowler Select Infusion Catheters.

Common questions about the PROWLER SELECT 606-S272X

Has the PROWLER SELECT 606-S272X been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886704028895), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the PROWLER SELECT 606-S272X?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 1, 2019 and Jun 30, 2023. In those reports, the most common specific device problems are “Contamination /Decontamination Problem” (1), “Peeled/Delaminated” (1) and “Physical Resistance/Sticking” (1); the most common patient problem is “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PROWLER SELECT 606-S272X?

None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PROWLER SELECT 606-S272X FDA approved?

It was cleared through Special 510(k) K021591 on May 22, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the PROWLER SELECT 606-S272X, and is it still sold?

The FDA's device database lists Medos International Sàrl as the company responsible for it (brand PROWLER SELECT, model 606-S272X). Its record shows it as in commercial distribution.