Skip to content

CODMAN EDS 3 82-1731Recalls & reported problems

Cerenovus, Inc. · CODMAN EDS 3 CSF External Drainage System · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886704040910), received between Jan 24, 2019 and Feb 15, 2019 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Disconnection” (6 reports).

6 other recalls cover devices cleared under the same FDA submission (K061568, K162437); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Disconnection6
  • Crack1
  • Failure to Disconnect1
  • Fluid/Blood Leak1
  • Obstruction of Flow1

Patient problems

  • No Consequences Or Impact To Patient6
  • Air Embolism1
  • Cerebrospinal Fluid Leakage1
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 15, 2019Disconnection; Disconnection

    TUBING BECAME DISCONNECTED FROM BURETTE CHAMBER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8345871 ↗
  • MalfunctionFeb 15, 2019Disconnection; Disconnection

    STOPCOCK DISCONNECTED FROM TUBING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8345844 ↗
  • MalfunctionFeb 15, 2019Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak

    EDS III LEAKED AT 3-WAY STOP COCK.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8345691 ↗
  • MalfunctionFeb 15, 2019Disconnection; Disconnection

    EDS III LEAKED AT CONNECTION AND WOULD NOT DRAIN INTO THE COLLECTION BAG.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8345599 ↗
  • MalfunctionFeb 15, 2019Obstruction of Flow; Obstruction of Flow

    INADEQUATE DRAINAGE AT DRIP CHAMBER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8345407 ↗
  • MalfunctionFeb 15, 2019Disconnection; Disconnection

    DISCONNECTION OF MALE TERMINATION WITH SYSTEM WAS FRACTURED, CAUSING DRAINAGE SYSTEM TO BECOME DISCONNECTED AT THE VENTRICULAR CATHETER FROM THE TUBING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8345394 ↗
  • MalfunctionFeb 15, 2019Disconnection; Disconnection

    EDS3 DID NOT ADEQUATELY DRAIN, SO ANOTHER WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 8345306 ↗
  • InjuryJan 29, 2019Crack; Crack

    CONNECTION TO THE PATIENT BECAME CRACKED, ALLOWING AIR INTO THE SYSTEM AND INTERCRANIAL VENTRICLES.

    Patient problems: Air Embolism

    MAUDE report 8287371 ↗
  • MalfunctionJan 24, 2019Failure to Disconnect; Failure to Disconnect

    EDS3 LEAKED FLUID FROM THE THREE WAY TAP FOLLOWING PATIENT BEING ROLLED.

    Patient problems: Cerebrospinal Fluid Leakage

    MAUDE report 8274764 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K061568, K162437. One submission often covers many models, so check the product and lot details in each record.

  • Z-0634-2024 · Nov 28, 2023 · Open, Classified
    External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx only
  • Z-0635-2024 · Nov 28, 2023 · Open, Classified
    EDS 3 CSF External Drainage System, REF: NT821731C, STERILEEO, Rx only
  • Z-1143-2021 · Jan 12, 2021 · Terminated
    Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebrospinal fluid (CFS) from the cerebral ventricles or the lumbar subarachnoid
  • Z-0090-2015 · Sep 15, 2014 · Terminated
    Codman EDS 3 Drainage System (no Ventricular Catheter) (External Drainage System) Catalog Number: 82-1731 Indicated for draining cerebrospinal fluid (CFS) and other fluids of simil
  • Z-0091-2015 · Sep 15, 2014 · Terminated
    Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar p
  • Z-0089-2015 · Sep 15, 2014 · Terminated
    Codman EDS 3 Drainage System with Ventricular Catheter (External Drainage System) Catalog Number: 82-1730 Indicated for draining cerebrospinal fluid (CFS) and other fluids of simil

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates