CODMAN HAKIM SIPHONGUARD 82-3146Recalls & reported problems

Cerenovus, Inc. · CODMAN HAKIM Programmable Valve Right Angle Valve with SIPHONGUARD Device and Unitized Distal Catheter Programmable in steps of 10mm H2O (98 Pa): 30mm H2O to 200mm H2O (294 Pa - 1960 Pa) Includes: Programmable Valve, 14cm Ventricular Catheter, 120cm Peritoneal Catheter, Information Manual and Priming Adapter · Class 2 · Implant

All models of CODMAN HAKIM SIPHONGUARD (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886704041535), received between Jan 20, 2022 and Nov 26, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Difficult to Program or Calibrate1
  • Infusion or Flow Problem1
  • Obstruction of Flow1

Patient problems

  • Insufficient Information3
  • Cerebral Ventriculomegaly1
  • Headache1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2024Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 20787402 ↗
  • InjuryJun 14, 2024Break; Break

    A PHYSICIAN REPORTED A HAKIM VALVE (ID 823182) WAS IMPLANTED DUE TO UNKNOWN REASON VIA VENTRICULOPERITONEAL (VP) SHUNT ON UNKNOWN DATE WITH UNKNOWN SETTING. THE DEVICE BROKE DUE TO PATIENT FALL. THEREFORE, THE VALVE WAS REMOVED AND REPLACED ON UNKNOWN DATE. ACCORDING TO INFORMATION PROVIDED, IT IS UNKNOWN HOW THE VALVE FAILURE WAS DISCOVERED. IT IS ALSO UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS AND SYMPTOMS DUE TO PRODUCT FAILURE, AND IF X-RAY TAKEN TO ASSURE NO PARTS LEFT IN PATIENT. CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19539032 ↗
  • InjuryJun 8, 2022Infusion or Flow Problem; Infusion or Flow Problem

    A PHYSICIAN REPORTED A HAKIM VALVE ((B)(4)) WAS IMPLANTED VIA V-P SHUNT ON (B)(6) 2020 WITH (B)(4). DUE TO UNDERDRAINAGE SYMPTOMS THE PATIENT VISITED THE HOSPITAL. THE FLOW OF THE VALVE WAS CONFIRMED, BUT NO IMPROVEMENT WAS SEEN EVEN IF THE PRESSURE WAS LOWERED. THE VALVE WAS REMOVED AND REPLACED ON (B)(6) 2022. THE PATIENT RECOVERED.

    Patient problems: Insufficient Information

    MAUDE report 14632479 ↗
  • InjuryJan 20, 2022Obstruction of Flow; Obstruction of Flow

    A PHYSICIAN REPORTED A HAKIM VALVE (ID (B)(4)) AS IMPLANTED VIA V-P SHUNT IN (B)(4) 2021 WITH SETTING 110MMH2O. ON (B)(6) 2021, THE PATIENT HAD A HEADACHE AND AN X-RAY WAS PERFORMED. VENTRICULAR ENLARGEMENT WAS CONFIRMED, SO THE VALVE SET PRESSURE WAS LOWERED TO 50 MMH2O, BUT THE VENTRICLES DID NOT SHRINK. ON (B)(4) 2021 A SHUNT CONTRAST WAS PERFORMED AND THE VALVE SYSTEM WAS SUSPECTED TO BE OBSTRUCTED; THEREFORE, ON (B)(6) 2021, THE VALVE SYSTEM WAS REMOVED AND REPLACED. THE PATIENT RECOVERED.

    Patient problems: Cerebral Ventriculomegaly; Headache

    MAUDE report 13308157 ↗