Skip to content

CODMAN HAKIM 82-3148Recalls & reported problems

Cerenovus, Inc. · CODMAN HAKIM Programmable Valve Right Angle Valve and Unitized Distal Catheter Programmable in steps of 10mm H2O (98 Pa): 30mm H2O to 200mm H2O (294 Pa - 1960 Pa) Includes: Programmable Valve, 14cm Ventricular Catheter, 120cm Peritoneal Catheter, Information Manual and Priming Adapter · Class 2 · Implant

All models of CODMAN HAKIM (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704041542), received between Mar 2, 2023 and Mar 24, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Obstruction of Flow1
  • Pressure Problem1

Patient problems

  • Insufficient Information2
  • Cerebral Ventriculomegaly1
  • Intracranial Hemorrhage1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 24, 2024Pressure Problem; Pressure Problem

    A PHYSICIAN REPORTED A SHUNT MALFUNCTIONING ON A HAKIM VALVE (ID 823148) AFTER DOING MAGNETIC RESONANCE IMAGING (MRI). A 80 YEAR-OLD MALE PATIENT WITH A CASE OF NORMAL PRESSURE HYDROCEPHALUS FOR WHICH PROGRAMMABLE VENTRICULOPERITONEAL (VP) SHUNT WAS DONE IN (B)(6) 2018. THE PATIENT WAS DOING ALRIGHT BUT READMITTED IN THE HOSPITAL IN (B)(6) 2023 FOR A SUSPECTED ISCHEMIC STROKE. A MAGNETIC RESONANCE IMAGING (MRI) BRAIN (PAIN) WAS PERFORMED 1.5 T MACHINE. THE SHUNT VALVE OPENING PRESSURE WAS KEPT DURING SURGERY AT 120MM H20, AND A PATIENT WAS DISCHARGED AFTER IT WAS RESET TO 120 MM H20. ON (B)(6) 2023, HE WAS ADMITTED AGAIN WITH BILATERAL MASSIVE SUBDURAL HEMORRHAGE FOR WHICH EMERGENCY SURGERY WAS DONE TO EVACUATE THE BLOOD IN THE ODD HOUR. IT WAS PRESUMED THAT THERE WAS OVERDRAINAGE OF CEREBROSPINAL FLUID (CSF), THEREFORE SHUNT TUBE WAS TIED AT THE CHEST WALL SO THAT NO CEREBROSPINAL FLUID (CSF) CAN FLOW SINCE NO TECHNICIAN WAS AVAILABLE TO RESET THE SHUNT. A PATIENT IMPROVED LATER AND WAS DISCHARGED FROM THE HOSPITAL. ON (B)(6), 2024, HE WAS ADMITTED AGAIN WITH DROWSINESS AND A COMPUTED TOMOGRAPHY (CT) BRAIN SHOWED VENTRICULOMEGALY. AN X-RAY PERFORMED TO SEE THE VALVE OPENING PRESSU

    Patient problems: Cerebral Ventriculomegaly; Intracranial Hemorrhage

    MAUDE report 18966522 ↗
  • InjuryJan 17, 2024Obstruction of Flow; Obstruction of Flow

    A PHYSICIAN REPORTED A HAKIM VALVE (ID 823148) WAS IMPLANTED IN A PATIENT WITH COMMUNICATING HYDROCEPHALUS VIA LUMBOPERITONEAL (LP) SHUNT ON UNKNOWN DATE WITH UNKNOWN SETTING. ON (B)(6) 2023, A REVISION SURGERY WAS PERFORMED DUE TO SUSPECTED OBSTRUCTION OF THE VALVE. THE VALVE AND VENTRICULAR CATHETER WERE EXPLANTED AND REPLACED ON (B)(6) 2023. ACCORDING TO INFORMATION PROVIDED, IT IS UNKNOWN IF THE VENTRICULAR CATHETER IS A COMPONENT OF THE HAKIM VALVE.

    Patient problems: Insufficient Information

    MAUDE report 18530289 ↗
  • InjuryMar 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    A FACILITY REPORTED A HAKIM VALVE (ID 823148) WAS IMPLANTED IN (B)(6) 2022 FOR MANAGEMENT OF CEREBROSPINAL FLUID RETENTION. THE PATIENT WAS ADMITTED IN EMERGENCY ROOM ON (B)(6) 2023 WITH DIAGNOSIS OF RIGHT CEREBELLOPONTINE ANGLE TUMOR WITH HYDROCEPHALUS, STATUS POST LEFT VENTRICULAR-PERITONEAL (VP) SHUNT INSERTION WITH SHUNT MALFUNCTION. THE PRESSURE SETTING WAS SET MANY TIMES BUT THERE WAS NO RESPONSE. THE VALVE WAS REMOVED ON UNKNOWN DATE, HOWEVER, THERE IS NO INFORMATION IF THE DEVICE WAS REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 16469989 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates