CODMAN HAKIM 82-3807Recalls & reported problems
Cerenovus, Inc. · CODMAN HAKIM Precision Fixed Pressure Valve In-Line Valve and Unitized Distal Catheter Operating Pressure: 40mm H2O Plus or Minus 10mm H2O (392 Pa Plus or Minus 98 Pa) Low Range Includes: Precision Valve, 14cm Ventricular Catheter, 120cm Unitized Peritoneal Catheter, Information Manual, Right Angle Adapter and Priming Adapter · Class 2 · Implant
All models of CODMAN HAKIM (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704041733), received between Jul 28, 2022 and Sep 1, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Migration or Expulsion of Device1
Patient problems
- Hydrocephalus1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 1, 2022Break; Break
A FACILITY REPORTED A HAKIM VALVE (ID 823807) WAS IMPLANTED ON (B)(6) 2022. THE UNITIZED SHUNT IS BROKEN AT THE POINT WHERE DISTAL CATHETER IS CONNECTED TO THE VALVE. IT WAS CLEARLY SEEN ON THE X-RAY THAT THE DISTAL CATHETER WAS NO LONGER CONNECTED TO THE VALVE. THE PATIENT BEGAN TO EXHIBIT CLINICAL SIGNS OF HYDROCEPHALUS ON (B)(6) 2022 AND A RADIOGRAPH WAS PERFORMED WHICH SHOWED THE SHUNT DEFECT. INFORMATION RECEIVED ON (B)(6) 2022 INDICATES THE CARE PLAN IS TO SEE IF THE PATIENT CAN BE MEDICALLY MANAGED BEFORE DECIDING TO DO ANOTHER SURGERY AND REPLACE THE VALVE. IN REGARDS THE CLINICAL SIGNS, MEDICAL STAFF IS STILL WAITING TO SEE IF MEDICAL MANAGEMENT IS SUFFICIENT.
Patient problems: Hydrocephalus
MAUDE report 15337402 ↗InjuryJul 28, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
2 OF 3 REPORTS. OTHER MFG REPORT NUMBERS: 3013886523-2022-00349, 3013886523-2022-00351. A FACILITY REPORTED 3 CATHETERS WERE IMPLANTED RECENTLY IN 3 DIFFERENT PEDIATRIC PATIENTS. AFTER FEW DAYS THEY HAD THE NEED THE REMOVE THE PERITONEAL CATHETER (NOT RELATED TO VALVE SYSTEM MALFUNCTION) AND THEY WEREN¿T ABLE TO PULL AWAY THE CATHETER (THAT WAS LEFT IN THE PERITONEAL SPACE. THE CATHETER WAS LEFT WITH ITS ENTIRE LENGTH WITHOUT LOOP CONFIGURATION). IN THE MOST SEVERE CASE, THE CATHETER WAS KNOT TOGETHER WITH AN INTESTINAL LOOP AND REQUIRE A SURGERY TO CUT THE LOOP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15126920 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page