ENVOY 671-258-95DBRecalls & reported problems

Medos International Sàrl · ENVOY DA XB Guiding Catheter MPD 95cm 6F (2.0mm) 0.071 Inches (1.8mm) · Class 2

All models of ENVOY (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886704075110), received between Feb 26, 2020 and Apr 20, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Crack” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack3
  • Burst Container or Vessel2
  • Fluid/Blood Leak1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 20, 2026Material Separation

    A HEALTHCARE PROFESSIONAL REPORTED THAT WHILE PREPPING AN ENVOY DA XB, 6F, 95 CM CATHETER (PRODUCT CODE: 67125895DB, LOT NUMBER: 31618625), THERE WAS SMALL ¿FLAKING¿ AT TIP. IT APPEARED TO BE THE COATING. THERE WAS NO PATIENT INJURY REPORTED. THERE WERE NO PROCEDURAL DELAYS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926617 ↗
  • MalfunctionMar 24, 2026Crack

    A HEALTHCARE PROFESSIONAL REPORTED THAT THERE WAS A HUB DEFECT ON AN ENVOY DA XB, 6F, 95 CM, MPD (PRODUCT CODE: 67125895DB, LOT NUMBER: 31674324). THE WIRE COULD NOT BE INSERTED AND THE DEVICE WAS UNUSABLE. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 24-MAR-2026 CONFIRMED THAT THE ¿HUB DEFECT¿ REFERRED TO A CRACK OR FRACTURE AT THE HUB. NO DEVICE FRAGMENTS REMAINED IN THE PATIENT. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WAS NO FOLLOW-UP INTERVENTIONS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24676554 ↗
  • MalfunctionJul 24, 2025Crack; Crack

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PROCEDURE ON (B)(6) 2025, THE HUB OF THE ENVOY DA XB, 6F, 95CM, MPD (67125895DB / 31502173) WAS BROKEN WHEN IT WAS FIRST OPENED. THE PHYSICIAN REPLACED IT WITH ANOTHER ENVOY DA XB, 6F, 95CM, MPD (67125895DB) AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO NEGATIVE IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22599541 ↗
  • MalfunctionMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE, THERE WAS A LEAK AT THE HUB OF THE ENVOY DA XB, 6F, 95CM, MPD (67125895DB / 31393518). MULTIPLE ROTATING HEMOSTASIS VALVES (RHVS) WERE TRIED IN ATTEMPT TO ALLEVIATE THE REPORTED LEAK, BUT THE CATHETER HUB KEPT LEAKING. THE LEAKING STOPPED WHEN THE PHYSICIAN REPLACED THE ENVOY CATHETER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY DELAY. NO INTERVENTION WAS REQUIRED. THERE WAS NO NEGATIVE IMPACT TO THE PATIENT. ON 21-MAY-2025, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE INTENDED PROCEDURE WAS EMBOLIZATION OF MIDDLE MENINGEAL ARTERY (MMA) FOR SUBDURAL HEMATOMA (SDH). THE PROCEDURE WAS TARGETING THE COMMON CAROTID ARTERY. THE REPLACEMENT DEVICE WAS ANOTHER ENVOY DA XB, 6F, 95CM, MPD (67125895DB).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22075417 ↗
  • MalfunctionApr 16, 2020Crack; Crack

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION, THE TIP OF A 6F, 95CM ENVOY DA XB CATHETER (67125895DB, LOT UNKNOWN) CRACKED. THERE WAS NO PATIENT INJURY REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9967435 ↗
  • MalfunctionFeb 26, 2020Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION TO AN ANEURYSM OF THE RIGHT POSTERIOR COMMUNICATING ARTERY (PCOM), AN ECLIPSE2L BALLOON WAS FIRST PLACED AND THEN A HEADWAY MICROCATHETER (MC) WAS USED AS USUAL. HOWEVER, THE MICROCATHETER DID NOT PASS THROUGH TWO 6F ENVOY DA 95 CM GUIDE CATHETERS (67125895DB, LOT UNKNOWN). IT SHOWED TO BE BROKEN FROM THE INSIDE AT THE HUBS, FROM THE OUTSIDE NO DEFECTS WERE DETECTABLE. THE DAMAGE SEEMINGLY AT THE END OF THE CATHETERS. BOTH COMPLETE SYSTEMS HAD TO BE REMOVED AND RE-LOADED. THERE WAS AN HOUR PROCEDURE DELAY DUE TO THE EVENT, HOWEVER, THE PATIENT DID NOT SUFFER ANY DAMAGE FROM THE PROLONGATION OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED BY THE USE OF A COMPETITOR¿S DEVICE. NO EXCESSIVE FORCE HAD BEEN APPLIED TO THE DEVICES. NO DEVICE FRAGMENTS REMAIN IN THE PATIENT. A PICTURE OF ONE DEVICE WAS RECEIVED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9758001 ↗
  • MalfunctionFeb 26, 2020Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION TO AN ANEURYSM OF THE RIGHT POSTERIOR COMMUNICATING ARTERY (PCOM), AN ECLIPSE2L BALLOON WAS FIRST PLACED AND THEN A HEADWAY MICROCATHETER (MC) WAS USED AS USUAL. HOWEVER, THE MICROCATHETER DID NOT PASS THROUGH TWO 6F ENVOY DA 95 CM GUIDE CATHETERS (67125895DB, LOT UNKNOWN). IT SHOWED TO BE BROKEN FROM THE INSIDE AT THE HUBS, FROM THE OUTSIDE NO DEFECTS WERE DETECTABLE. THE DAMAGE SEEMINGLY AT THE END OF THE CATHETERS. BOTH COMPLETE SYSTEMS HAD TO BE REMOVED AND RE-LOADED. THERE WAS AN HOUR PROCEDURE DELAY DUE TO THE EVENT, HOWEVER, THE PATIENT DID NOT SUFFER ANY DAMAGE FROM THE PROLONGATION OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED BY THE USE OF A COMPETITOR¿S DEVICE. NO EXCESSIVE FORCE HAD BEEN APPLIED TO THE DEVICES. NO DEVICE FRAGMENTS REMAIN IN THE PATIENT. A PICTURE OF ONE DEVICE WAS RECEIVED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9757943 ↗