CODMAN CERTAS 82-8800PLRecalls & reported problems
Medos International Sàrl · CODMAN CERTAS Plus Programmable Valve Inline Valve Only Includes: Priming Adapter and Information Manual · Class 2 · Implant
All models of CODMAN CERTAS (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 1
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886704079002), received between Dec 21, 2018 and Feb 27, 2024 — classified as 1 death, 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Infusion or Flow Problem1
- Insufficient Device Problem Information1
- Obstruction of Flow1
- Output Problem1
- Self-Activation or Keying1
Patient problems
- Balance Problems1
- Cerebral Edema1
- Cerebral Ventriculomegaly1
- Cerebrospinal Fluid Leakage1
- Dizziness1
- Headache1
- Hydrocephalus1
- Lethargy1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryFeb 27, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
InjuryAug 24, 2022Output Problem; Output Problem
A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) WAS IMPLANTED ON (B)(6), 2019. ON (B)(6), 2022 THE PATIENT ATTEMPTED TO HAVE SHUNT REPROGRAMMED FROM A SETTING 5 TO 6. THEY ATTEMPTED TO REPROGRAM TO 6 AND BOTH THE MANUAL AND ELECTRONIC PROGRAM CONFIRMED CHANGE. THE PATIENT WAS SENT DOWN TO X-RAY TO CONFIRM BUT X RAY CONTINUED TO SHOW VALVE WAS AT A 5. THEY SET SHUNT AT 7 AS THAT WAS CLOSEST TO 6 BUT ON AUGUST 05, 2022, PATIENT PRESENTED WITH SHUNT FAILURE SYMPTOMS, THEREFORE, THE VALVE WAS EXPLANTED AND REPLACED WITH A NEW VALVE. THE PATIENT WAS HAVING SIGNS OF VENTRICULAR ENLARGEMENT AND SEVERE LETHARGY WHEN UNDERDRAINING.
Patient problems: Cerebral Ventriculomegaly; Lethargy
MAUDE report 15284170 ↗InjuryMay 23, 2022Fluid/Blood Leak; Fluid/Blood Leak
A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) WAS IMPLANTED VIA L-P SHUNT ON (B)(6) 2022. A POCKET OF FLUID AROUND THE SHUNT VALVE WAS FOUND AND THE GRAVITY COMPENSATOR WAS LEAKING, THEREFORE THE VALVE WAS REMOVED ON (B)(6) 2022. BASED ON INFORMATION PROVIDED, IT IS UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS AND SYMPTOMS DUE TO THE VALVE ISSUE.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 14477792 ↗InjuryMar 1, 2022Obstruction of Flow; Obstruction of Flow
A PHYSICIAN REPORTED A CERTAS VALVE (ID 828800) WAS IMPLANTED IN A (B)(6) FEMALE PATIENT VIA "L-P" SHUNT ON (B)(6) 2021 WITH SETTING 2. THE VALVE WAS USED WITH THE SILASCON LUMBAR CATHETER (MANUFACTURED BY KANEKA, PRODUCT CODE: 702-JJ). A SHUNT DYSFUNCTION WAS SUSPECTED AND SHUNT CONTRAST EXAMINATION CONFIRMED A PROXIMAL OBSTRUCTION ON (B)(6) 2022. THE PATIENT HAD REPEATED FALLS AND IT IS UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS AND SYMPTOMS. THE VALVE AND THE LUMBAR CATHETER WERE REMOVED ON (B)(6) . THE VALVE WAS REPLACED ON (B)(6) . THE PATIENT RECOVERED.
Patient problems: Balance Problems
MAUDE report 13636305 ↗DeathFeb 10, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A FACILITY REPORTED A VALVE (UNKNOWN MANUFACTURER) HAS BEEN IMPLANTED FOR 40 YEARS AND NEVER REVISED. THE PATIENT IS A (B)(6) FEMALE SINCE BIRTH PRESENTED WITH PROXIMAL VP SHUNT OBSTRUCTION. THE PATIENT PRESENTED WITH HYDROCEPHALUS AND HEADACHES FOR 72 MONTHS AND ON (B)(6) 2022, A CERTAS PLUS VALVE (ID (B)(6)) WAS IMPLANTED WITH SETTING 5. AT THE TIME OF THE VALVE IMPLANTATION, THE PATIENT HAD A HEADACHE, BUT SHE WAS NEUROLOGICALLY INTACT AND THERE WERE NO ISSUES DURING VALVE INSERTION. THE PATIENT WAS MONITORED IN ICU FOR 24 HOURS AND ON (B)(6) 2022, THE SHUNT SETTINGS CHANGED FROM 5 TO 3 AND THE PATIENT CODED (TACHYCARDIC AND WAS FOUND UNRESPONSIVE IN BED). THE PATIENT SUFFERED CARDIAC ARREST AT APPROXIMATELY 9:00P.M. ON (B)(6) 2022, A REVISION SURGERY WAS ATTEMPTED UNSUCCESSFULLY DUE TO COLLAPSED VENTRICLES DUE TO MASSIVE BRAIN EDEMA. THE PATIENT PASSED AWAY ON (B)(6) 2022.
Patient problems: Cerebral Edema; Headache; Hydrocephalus
MAUDE report 13506957 ↗MalfunctionJan 6, 2022Infusion or Flow Problem; Infusion or Flow Problem
A FACILITY REPORTED A CERTAS VALVE WOULD NOT DRAIN. NO PATIENT INJURY REPORTED. THE VALVE WAS USED TO TREAT NORMAL PRESSURE HYDROCEPHALUS. NO ADDITIONAL INFORMATION WAS PROVIDED AFTER SEVERAL ATTEMPTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13174743 ↗MalfunctionDec 21, 2018Self-Activation or Keying
THE PATIENT HAD A PROGAV SHUNT VALVE FOR VPS. IT WAS DISCOVERED THAT THIS VALVE CHANGED SETTING ON ITS OWN, WITHOUT BEING PROGRAMMED. WE WERE ALSO UNABLE TO RE-PROGRAM THE VALVE IN THE CLINICAL SETTING. IT WAS CONCLUDED THAT VALVE WAS MALFUNCTIONING AND PATIENT WAS TAKEN TO THE OPERATING ROOM TO REMOVE PROGAV VALVE AND EXCHANGE FOR A CERTAS PLUS VALVE.
Patient problems: No Code Available
MAUDE report 8189267 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page