CODMAN CERTAS 82-8810PLRecalls & reported problems
Medos International Sàrl · CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only Includes: Priming Adapter and Information Manual · Class 2 · Implant
All models of CODMAN CERTAS (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886704081890), received between Dec 30, 2021 and Aug 20, 2024 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Pressure Problem2
- Inappropriate or Unexpected Reset1
- Obstruction of Flow1
- Output Problem1
- Positioning Failure1
Patient problems
- Insufficient Information2
- Irritability2
- Ambulation or Postural Difficulties1
- Cerebral Ventriculomegaly1
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
- Vomiting1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2024Positioning Failure; Positioning Failure
MEDWATCH #: (B)(4). A FACILITY REPORTED A CERTAS VALVE (ID 828810PL) WAS IMPLANTED ON (B)(6) 2021 WITH SETTING AT 3. IN (B)(6) 2023, THE VALVE WAS REPROGRAMMED FROM 3 TO 5 DUE TO A PROGRESSIVE DECREASE IN VENTRICLE SIZE. THE PATIENT DID WELL WITH NO RECURRENT SEIZURES. ON (B)(6) 2024, DURING A CLINIC VISIT, IT WAS FOUND THAT THE VALVE COULD NO LONGER BE ADJUSTED FROM ITS SETTING OF 3 AND THE MAGNETIC RESONANCE IMAGING (MRI) SHOWED DILATATION OF THE LEFT LATERAL VENTRICLE, WHICH WAS HIGHER THAN WHAT WAS SEEN ON IMAGING IN (B)(6) 2023. THE VENTRICLE SIZE IN (B)(6) 2024 WAS LARGER THAN HER PREVIOUS VENTRICLES SIZE WHEN HER PREVIOUS INTENTIONAL SETTING WAS AT 3 IN (B)(6) OF 2023. BETWEEN (B)(6) 2024 AND (B)(6) 2024, THE PATIENT HAD WORSENING OF ABILITY TO PERFORM MOTOR SKILLS (STOPPED WALKING IN (B)(6) 2024), RUBBING EARS AND RUBBING HEAD (NON-VERBAL PATIENT), WAKING FROM SLEEP CRYING, INABILITY TO SLEEP IN RECUMBENT, INCREASED DAYTIME IRRITABILITY, AND PROGRESSIVE VENTRICULAR DILATION DUE TO ABNORMAL INTRACRANIAL PRESSURE. THE VALVE WAS REMOVED AND REPLACED. THE PATIENT SYMPTOMS IMPROVED, IS NOW WALKING AGAIN AND NO LONGER RUBBING EARS OR HEAD.
Patient problems: Ambulation or Postural Difficulties; Cerebral Ventriculomegaly; Irritability
MAUDE report 20004135 ↗InjuryJun 9, 2023Obstruction of Flow; Obstruction of Flow
A FACILITY REPORTED A CERTAS VALVE (ID 828810PL) WAS IMPLANTED ON (B)(6) 2021 WITH SETTING 7 OF 38MMH20. DURING A CLINIC CHECK UP THE VALVE WAS FOUND TO BE "STUCK". THERE WERE SEVERAL ATTEMPTS TO CHANGE/UNSTICK THE VALVE TO THE DESIRED SETTING. ON (B)(6) 2023 , THE DECISION WAS MADE TO EXPLANT THE VALVE AND TEST THE COMPONENTS TO FIND THE CAUSE OF THE ISSUES. VENTRICULAR CATHETER WAS CHECKED AND WAS CONFIRMED OPERATIONAL WITH AN ICP OF 16 BASED ON MANOMETER READING. PERITONEAL CATHETER WAS CHECKED AND CONFIRMED OPERATIONAL FROM MANOMETER. THE VALVE WAS EXPLANTED TESTED AT 110 MMH2O. A NEW VALVE WAS IMPLANTED AT SETTING 4 = 110 MMH2O.
Patient problems: Insufficient Information
MAUDE report 17100880 ↗MalfunctionJan 13, 2023Inappropriate or Unexpected Reset
PATIENT WAS BROUGHT TO THE OPERATING ROOM FOR REVISION OF VENTRICULAR SHUNT. A PRE MRI SHUNT DIAL XRAY WAS PERFORMED AND SETTING NOTED AT 4. POST MRI SHUNT DIAL X-RAY NOTED A CHANGE IN SETTING TO 6. NEUROSURGERY UNABLE TO ADJUST VALVE. CODMAN COMPANY REP PRESENT AND UNABLE TO ADJUST VALVE. SHUNT HAD TO BE SURGICALLY REMOVED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 16156729 ↗MalfunctionSep 14, 2022Pressure Problem; Pressure Problem
A FACILITY REPORTED A CERTAS VALVE (ID 828810PL) WAS IMPLANTED IN A 17 YEAR OLD MALE PATIENT ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT HAD AN MRI. IMMEDIATELY FOLLOWING THE MRI THE VALVE WAS CHECKED WITH THE CERTAS MANUAL VERIFICATION TOOL. IT WAS VERIFIED THAT THE VALVE WAS SET TO SETTINGS 6. THE PATIENT WAS THEN SENT TO X-RAY TO CONFIRM AND IT WAS CONFIRMED THAT THE VALVE WAS AT SETTING 6. THEN THEY REPROGRAMMED THE PATIENT BACK TO SETTING 5 AND WAS CONFIRMED WITH THE MANUAL VERIFICATION TOOL AND X-RAY. IT WAS STATED THAT THE PATIENT DID NOT EXPERIENCE ANY SIGNS OR SYMPTOMS OF UNDER DRAINAGE BECAUSE OF THE HIGHER SETTING. THE VALVE REMAINS IMPLANTED AT THE TIME OF THIS REPORT. THE PATIENT IS DOING FINE WITH CERTAS PLUS VALVE CURRENTLY SET TO SETTING 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15420403 ↗InjuryMar 29, 2022Pressure Problem; Pressure Problem
A FACILITY REPORTED A CERTAS VALVE WAS IMPLANTED WITH SETTING 5 ON (B)(6) 2020 TO TREAT HYDROCEPHALUS. DUE TO SOME BLOOD IN THE VENTRICLES, THE SETTING WAS CHANGED TO 7 ON THE SAME DAY. ON (B)(6) 2020, THE SETTING WAS CHANGED TO 5 WITH DIGITAL AND MAGNETIC CONFIRMATION. IN (B)(6) 2022, THE PATIENT'S PARENTS CALLED INFORMING THE PATIENT HAD IRRITABILITY WHICH PROGRESSED INTO VOMITING AND IT WAS FOUND THE VALVE WAS SET AT SETTING 8 (SUPPOSED TO BE IN 5). IT IS UNKNOWN IF THE PATIENT HAD MAGNETIC EXPOSURE, MRIS, ETC. THE VALVE WAS ADJUSTED TO SETTING 5, CONFIRMED ON THE MACHINE AND X-RAY. PATIENT IMPROVED.
Patient problems: Irritability; Vomiting
MAUDE report 13938519 ↗InjuryDec 30, 2021Output Problem; Output Problem
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page