CODMAN CERTAS SIPHONGUARD 82-8814PLRecalls & reported problems
Medos International Sàrl · CODMAN CERTAS SIPHONGUARD Plus Programmable Valve Inline Small Valve Only with SIPHONGUARD Includes: Priming Adapter and Information Manual · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886704081937), received between Feb 16, 2022 and Jul 11, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Obstruction of Flow” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Obstruction of Flow2
- Break1
- Fluid/Blood Leak1
- Fracture1
- Pressure Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Insufficient Information1
- Lethargy1
- Vomiting1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 11, 2023Obstruction of Flow; Obstruction of Flow
InjuryFeb 22, 2023Break; Pressure Problem; Break; Pressure Problem
N/A.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 16419214 ↗InjuryMay 26, 2022Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
A PHYSICIAN REPORTED A CERTAS VALVE (UNKNOWN ID) WAS IMPLANTED IN (B)(6) 2022. THE PATIENT WAS SEEN FOR FOLLOW UP ON (B)(6) 2022; THE VENTRICLES WERE SLIT-LIKE SO THE VALVE SETTING WAS INCREASED FROM 1 TO 3. ON (B)(6) 2022 THE PATIENT WAS SEEN WITH SHUNT MALFUNCTION SYSTEM. THE PATIENT WAS LETHARGIC, VOMITING, NOT EATING, AND NOT SEEMING HIMSELF. IMAGING INCREASED VENTRICLES. THEREFORE, THE VALVE WAS REMOVED AND REPLACED ON (B)(6) 2022. UPON VP SHUNT REVISION PERFORMED: ON INITIAL INCISION GENERALIZED CSF LEAKAGE WAS NOTED AROUND THE VALVE. EXPLORING VALVE SITE NOTED THAT THE VALVE ITSELF WAS FRACTURED, WITH NOTABLE CSF LEAKING, AND APPEARED TO BE COMPLETELY SHEARED OFF AT SIPHONGUARD AREA OF THE VALVE.
Patient problems: Lethargy; Vomiting
MAUDE report 14519063 ↗InjuryFeb 16, 2022Obstruction of Flow; Obstruction of Flow
A PHYSICIAN REPORTED A CERTAS VALVE (ID 828814) WAS IMPLANTED VIA V-P SHUNT ON (B)(6) 2022 WITH UNKNOWN SETTING. AFTER THE VALVE WAS IMPLANTED THERE WAS NO CHANGE IN THE VENTRICLES. ON (B)(6) 2022 THE SET PRESSURE COULD NOT BE CHANGED FROM SETTING 6 TO 1. ON (B)(6) 2022 ANOTHER ATTEMPT WAS MADE TO CHANGE THE SET PRESSURE USING A NEODYMIUM MAGNET, BUT THE PRESSURE COULD NOT BE CHANGED EITHER. AFTER ADDING A CONTRAST MEDIUM AND CONFIRMING PATENCY, VALVE OBSTRUCTION WAS CONFIRMED. THE VALVE WAS REMOVED AND REPLACED ON (B)(6) 2022.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13543067 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page