CEREGLIDE NIC71125URecalls & reported problems
Cerenovus, Inc. · CEREGLIDE 71 Intermediate Catheter 125 cm 0.071 in (1.80 mm) 0.0820 in (2.08 mm) 0.0825 in (2.10 mm) · Class 2
All models of CEREGLIDE (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886704085423), received between Sep 11, 2024 and Feb 4, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Deformation Due to Compressive Stress2
- Fluid/Blood Leak2
- Crack1
- Difficult to Advance1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2026Material Twisted/Bent; Physical Resistance/Sticking
A HEALTHCARE PROFESSIONAL REPORTED AN ISSUE OCCURRING DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE INVOLVING THE 125 CM CEREGLIDE 71 CATHETER (PRODUCT CODE: NIC71125U, LOT: 31722200) USED COAXIALLY WITH A NON¿J&J MICROCATHETER. DURING ADVANCEMENT, THE CEREGLIDE CATHETER BECAME IMPEDED AT THE C1¿C2 SEGMENT OF THE INTERNAL CAROTID ARTERY AND WAS UNABLE TO REACH THE INTENDED TARGET SITE. THE PHYSICIAN WITHDREW BOTH THE CEREGLIDE CATHETER AND THE MICROCATHETER. UPON REMOVAL, THE CEREGLIDE CATHETER WAS OBSERVED TO BE KINKED/BENT. THE PHYSICIAN REPLACED THE CEREGLIDE DEVICE WITH AN ALTERNATIVE BRAND CATHETER AND COMPLETED THE PROCEDURE SUCCESSFULLY USING THE ORIGINAL MICROCATHETER. NO DELAY IN THERAPY OR PROCEDURAL COMPLICATION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 02 FEB 2026 INDICATED THAT THE MICROCATHETER USED WAS IDENTIFIED AS A MEDTRONIC ECHELON10. VESSEL TORTUOSITY WAS PRESENT. NO PHYSICAL MATERIAL OR FOREIGN MATTER WAS IDENTIFIED INSIDE THE DEVICE. THERE WAS NO ALLEGATION OF PATIENT INJURY RELATED TO A PRODUCT ISSUE. NO PATIENT INJURY OR ADVERSE CLINICAL OUTCOME WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24260262 ↗MalfunctionNov 17, 2025Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A THROMBECTOMY PROCEDURE USING A 125M CEREGLIDE 71 CATHETER (NIC71125U / 31464782), THE FIRST ASPIRATION ATTEMPT WAS SUCCESSFUL. HOWEVER, AFTER REMOVING THE CEREGLIDE CATHETER TO ATTEMPT A SECOND ASPIRATION, THERE WERE TWO FLATTENED AREAS AT THE PROXIMAL END. IT WAS REPORTED THAT ¿EVEN A MICROWIRE COULD NOT PASS THROUGH.¿ THE PROCEDURE WAS CONTINUED USING A 5MM X 22MM EMBOTRAP III REVASCULARIZATION DEVICE AND A PHENOM¿ 27 MICROCATHETER (MEDTRONIC) TO RETRIEVE THE THROMBUS. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23576990 ↗MalfunctionAug 25, 2025Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED VIA A HEALTHCARE PROFESSIONAL THAT THIS WAS A MECHANICAL THROMBECTOMY OF AN OCCLUSION AT THE INTERNAL CAROTID ARTERY TO TREAT CEREBRAL INFARCTION, AFTER THE FIRST PASS WITH A 125CM CEREGLIDE 71 CATHETER (NIC71125C, 31526795), THE CEREGLIDE WAS PULLED FROM THE PATIENT'S BODY, AND FOUND A FLATTENED AREA OF APPROXIMATELY 2 CM AT THE TIP OF THE CATHETER. THE CEREGLIDE 71 WAS REPLACED WITH ANOTHER NEW ONE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO NEGATIVE IMPACT TO THE PATIENT. IT IS UNKNOWN IF A CONTINUOUS FLUSH WAS PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22883552 ↗MalfunctionJun 10, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE PHYSICIAN OPENED THE PACKAGE OF A 125M CEREGLIDE 71 CATHETER (NIC71125U / 31464782), BEFORE INTRODUCING THE CEREGLIDE CATHETER, THE PHYSICIAN NOTICED BRISK SALINE LEAKING AROUND THE HUB. THE DEVICE WAS NOT INTRODUCED INTO THE PATIENT. THERE WAS NO NEGATIVE PATIENT IMPACT. ON 09-JUN-2025, ADDITIONAL INFORMATION WAS RECEIVED. THE INFORMATION INDICATED THAT THE PROCEDURE DATE WAS ON (B)(6) 2025. THE DEVICE WAS REPLACED WITH ANOTHER 125M CEREGLIDE 71 CATHETER (NIC71125U) AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY ISSUE. THERE WAS A DELAY OF ¿MAYBE 1-2 MINUTES MAX,¿ BUT IT WAS NOT CLINICALLY SIGNIFICANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22181039 ↗MalfunctionJan 2, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
AS REPORTED BY THE FIELD, A 125CM CEREGLIDE 71 CATHETER (NIC71125C, 31395151) WAS PUSHED AT THE ACCESS, TILL THE VERY END TOUCHING THE Y ADAPTOR. PROBABLY DUE TO THE MANIPULATION AND CONTACT TO THE ACCESS SYSTEM THE PROXIMAL ¿NECK¿ OF THE CEREGLIDE SHAFT WAS BROKEN AND AIR WAS (WOULD) ENTER THE SYSTEM WHILE ASPIRATION THROUGH A WHOLE. THEY DID NOT CONTINUE AND CHANGED THE CATHETER TO ANOTHER COMPETITIVE BRAND AND CONTINUED SUCCESSFULLY WITH THE PROCEDURE, NO SIGNIFICANT DELAY, NO ADDITIONAL RISK, NO HARM HAPPENED TO THE PATIENT. THE PHYSICIAN ADMITTED THAT HE MIGHT SHOULD HAVE USED A LONGER CEREGLIDE TO AVOID IRRITATION AT THE PROXIMAL NECK, BUT HE STATED THAT THIS IS UNEXPECTED THAT THIS WOULD CAUSE DAMAGE, SINCE OTHER ASPIRATION CATHETERS WITHSTAND THIS TYPE OF MANIPULATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21055922 ↗InjurySep 11, 2024Difficult to Advance; Difficult to Advance
IT WAS REPORTED, VIA ELECTRONIC MAIL, THAT A 125CM CEREGLIDE 71 CATHETER (NIC71125C / 31332257) WAS USED FOR A THROMBECTOMY PROCEDURE. DURING THE PROCEDURE, THE USER REPORTED THE CEREGLIDE 71 NEED TO BE REPLACED WITH THEIR STANDARD CATHETER, BECAUSE IT WAS NOT POSSIBLE TO REACH ABOVE THE OPHTHALMIC LIP WITH CEREGLIDE. THE EVENT WAS INITIALLY REPORTED AS SUCH, ¿THE CATHETER NEEDED TO BE CHANGED FROM CEREGLIDE TO THEIR STANDARD CATHETER, BECAUSE NOT POSSIBLE TO REACH ABOVE THE OPHTHALMIC LIP WITH CEREGLIDE, THIS WORKED AFTER CHANGING TO STANDARD CATHETER.¿ ON 26-AUG-2024, ADDITIONAL INFORMATION WAS RECEIVED. SUMMARY: PER THE INFORMATION, THE EVENT DID NOT RESULT IN ANY CONSEQUENCE OR INJURY TO THE PATIENT. REGARDING IF THE EVENT RESULTED IN ANY UNDUE PROCEDURAL DELAY OR CONSEQUENCE THAT COULD BE CONSIDERED CLINICALLY SIGNIFICANT, IT WAS REPORTED, ¿SLIGHT DELAY IN RECANALIZATION OF INTRACRANIAL OCCLUSION.¿ RELEVANT ANATOMICAL INFORMATION WAS REPORTED AS ¿SEVERE CAROTID TORTUOSITY.¿ THE DEVICE DID NOT APPEAR DAMAGED AT ANY TIME. A CONTINUOUS FLUSH WAS MAINTAINED. THE PATIENT IS A 39-YEAR-OLD FEMALE. NO FURTHER INFORMATION WAS MADE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20191229 ↗
FDA clearance and classification
Cleared through 510(k) K221930 on Mar 9, 2023 (decision: Substantially Equivalent). Submitted by Cerenovus, Inc. as “CEREGLIDE 71 Intermediate Catheter”. The FDA review took 251 days. FDA record ↗
Cleared through 510(k) K221934 on Mar 9, 2023 (decision: Substantially Equivalent). Submitted by Cerenovus, Inc. as “CEREGLIDE 71 Intermediate Catheter; Cerenovus Aspiration Tubing Set”. The FDA review took 251 days. FDA record ↗
The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.
Common questions about the CEREGLIDE NIC71125U
Has the CEREGLIDE NIC71125U been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886704085423), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CEREGLIDE NIC71125U?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Sep 11, 2024 and Feb 4, 2026. In those reports, the most common specific device problems are “Deformation Due to Compressive Stress” (2), “Fluid/Blood Leak” (2) and “Crack” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CEREGLIDE NIC71125U?
None of the reports are classified as deaths, 1 as an injury and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CEREGLIDE NIC71125U FDA approved?
It was cleared through 510(k) K221930 on Mar 9, 2023; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the CEREGLIDE NIC71125U, and is it still sold?
The FDA's device database lists Cerenovus, Inc. as the company responsible for it (brand CEREGLIDE, model NIC71125U). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, THROMBUS RETRIEVER devices
- Penumbra System 5MAXJET7KIT-BPenumbra, Inc. · Catheter, Thrombus Retriever1 recall14 deaths221 reports
- Penumbra System 3MAXCPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER1 death241 reports
- EMBOTRAP ET309537Neuravi Limited · Catheter, Thrombus Retriever13 deaths177 reports
- ACE 5MAXACE068KITPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER170 reports
- Embotrap ET009533Neuravi Limited · Catheter, Thrombus Retriever16 deaths153 reports
- Penumbra System 5MAXACE132Penumbra, Inc. · CATHETER, THROMBUS RETRIEVER3 deaths112 reports
- Penumbra System 5MAXJET7KIT-APenumbra, Inc. · Catheter, Thrombus Retriever1 death94 reports
- Trevo XP 90185Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death40 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page