Medos International Sàrl · DISTAL RIB CLEAR CANNULA WITH OBTURATOR 7mm x 75mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886705003563), received between Feb 19, 2025 and Feb 19, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 5 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product5
  • Packaging Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem

    THIS IS REPORT 7 OF 7 FOR (B)(4). IT WAS REPORTED THAT PRIOR TO A SHOULDER ARTHROSCOPY PROCEDURE, A TOTAL OF SEVEN CANNULA DEVICES WERE FOUND TO BE COMPLETELY DAMAGED. THE REPORTER NOTED THAT THE BOXES WERE BENT AND COMPROMISED, AND THAT THE STERILITY OF CERTAIN CANNULAS HAD BEEN AFFECTED DUE TO TEARS. WHILE THE OUTER IMPLANT BOXES WERE DAMAGED, THE PACKAGING INSIDE REMAINED STERILE. THE REPORTER STATED THAT THEY CHOSE TO RETAIN THE IMPLANTS SINCE THEY WERE UNDAMAGED. THE CANNULAS WERE DISCARDED AFTER BEING TAKEN OUT OF TRUNK STOCK. THERE WAS A SLIGHT DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21414640 ↗
  • MalfunctionFeb 19, 2025Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem

    THIS IS REPORT 4 OF 7 FOR (B)(4). IT WAS REPORTED THAT PRIOR TO A SHOULDER ARTHROSCOPY PROCEDURE, A TOTAL OF SEVEN CANNULA DEVICES WERE FOUND TO BE COMPLETELY DAMAGED. THE REPORTER NOTED THAT THE BOXES WERE BENT AND COMPROMISED, AND THAT THE STERILITY OF CERTAIN CANNULAS HAD BEEN AFFECTED DUE TO TEARS. WHILE THE OUTER IMPLANT BOXES WERE DAMAGED, THE PACKAGING INSIDE REMAINED STERILE. THE REPORTER STATED THAT THEY CHOSE TO RETAIN THE IMPLANTS SINCE THEY WERE UNDAMAGED. THE CANNULAS WERE DISCARDED AFTER BEING TAKEN OUT OF TRUNK STOCK. THERE WAS A SLIGHT DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21414595 ↗
  • MalfunctionFeb 19, 2025Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem

    THIS IS REPORT 5 OF 7 FOR (B)(4). IT WAS REPORTED THAT PRIOR TO A SHOULDER ARTHROSCOPY PROCEDURE, A TOTAL OF SEVEN CANNULA DEVICES WERE FOUND TO BE COMPLETELY DAMAGED. THE REPORTER NOTED THAT THE BOXES WERE BENT AND COMPROMISED, AND THAT THE STERILITY OF CERTAIN CANNULAS HAD BEEN AFFECTED DUE TO TEARS. WHILE THE OUTER IMPLANT BOXES WERE DAMAGED, THE PACKAGING INSIDE REMAINED STERILE. THE REPORTER STATED THAT THEY CHOSE TO RETAIN THE IMPLANTS SINCE THEY WERE UNDAMAGED. THE CANNULAS WERE DISCARDED AFTER BEING TAKEN OUT OF TRUNK STOCK. THERE WAS A SLIGHT DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21414594 ↗
  • MalfunctionFeb 19, 2025Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem

    THIS IS REPORT 1 OF 7 FOR (B)(4). IT WAS REPORTED THAT PRIOR TO A SHOULDER ARTHROSCOPY PROCEDURE, A TOTAL OF SEVEN CANNULA DEVICES WERE FOUND TO BE COMPLETELY DAMAGED. THE REPORTER NOTED THAT THE BOXES WERE BENT AND COMPROMISED, AND THAT THE STERILITY OF CERTAIN CANNULAS HAD BEEN AFFECTED DUE TO TEARS. WHILE THE OUTER IMPLANT BOXES WERE DAMAGED, THE PACKAGING INSIDE REMAINED STERILE. THE REPORTER STATED THAT THEY CHOSE TO RETAIN THE IMPLANTS SINCE THEY WERE UNDAMAGED. THE CANNULAS WERE DISCARDED AFTER BEING TAKEN OUT OF TRUNK STOCK. THERE WAS A SLIGHT DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21414431 ↗
  • MalfunctionFeb 19, 2025Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem

    THIS IS REPORT 3 OF 7 FOR (B)(4). IT WAS REPORTED THAT PRIOR TO A SHOULDER ARTHROSCOPY PROCEDURE, A TOTAL OF SEVEN CANNULA DEVICES WERE FOUND TO BE COMPLETELY DAMAGED. THE REPORTER NOTED THAT THE BOXES WERE BENT AND COMPROMISED, AND THAT THE STERILITY OF CERTAIN CANNULAS HAD BEEN AFFECTED DUE TO TEARS. WHILE THE OUTER IMPLANT BOXES WERE DAMAGED, THE PACKAGING INSIDE REMAINED STERILE. THE REPORTER STATED THAT THEY CHOSE TO RETAIN THE IMPLANTS SINCE THEY WERE UNDAMAGED. THE CANNULAS WERE DISCARDED AFTER BEING TAKEN OUT OF TRUNK STOCK. THERE WAS A SLIGHT DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21414422 ↗

Other CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices

All CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices →