Skip to content

NA 214116Recalls & reported problems

Medos International Sàrl · THREADED CLEAR CANNULA WITH OBTURATOR 7mm x 75mm · Class 1

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886705003570), received between Dec 5, 2017 and Jan 7, 2026 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Delivered as Unsterile Product1
  • Detachment of Device or Device Component1
  • Device Contaminated During Manufacture or Shipping1
  • Leak/Splash1
  • Mechanical Jam1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2026Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, WHEN THE SURGEON INSERTED THE CLEARCANNULA-THREADED 7.0MMX75MM DEVICE INTO THE BODY, THE TIP OF THE DEVICE CHIPPED OFF AND BROKE, AND FRAGMENTS OF THE DEVICE FELL INTO THE JOINT. IT WAS REPORTED THAT THE FRAGMENTS WERE RETRIEVED AND BECAUSE THEY ARE NOT MADE OF MATERIAL THAT SHOWS UP ON X-RAYS, IT WAS CONFIRMED THAT THE SHAPE OF THE BROKEN BODY AND THE FRAGMENTS MATCHED. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS 30-MINUTE SURGICAL DELAY DUE TO THE EVENT. THERE WAS NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23991469 ↗
  • MalfunctionJun 23, 2025Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR (ARCR) PROCEDURE, WHEN THE SURGEON INSERTED THE ANCHOR DEVICE THROUGH THE CLEARCANNULA-THREADED 7.0MMX75MM CANNULA DEVICE, THE MIDDLE LAYER OF THE THREE-LAYER SEAL WAS DAMAGED, CAUSING THE ANCHOR TO SWERVE AWAY FROM THE INSIDE OF THE CANNULA DEVICE. THIS EVENT OCCURRED ALTHOUGH THE SURGEON USED THE DEVICE ACCORDING TO THE SURGICAL TECHNIQUE. IT WAS REPORTED THAT THE BROKEN PART/FRAGMENT WAS REMOVED, AND NOTHING WAS LEFT IN THE BODY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES, OR SURGICAL DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22310306 ↗
  • MalfunctionJan 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE SURGEON USED THE CLEARCANNULA-THREADED 7.0MMX75MM 5PK-ST DEVICE AS USUAL. HOWEVER, THE DISK-SHAPED PLASTIC COMPONENT CAME OFF. IT WAS REPORTED THAT THE SURGEON, WHO HAS USED THIS DEVICE FOR MANY YEARS, NOTED THAT THEY HAD NEVER ENCOUNTERED SUCH DAMAGE BEFORE. THERE WAS THIRTY-MINUTE DELAY TO THE SURGICAL PROCEDURE. A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY SUCCESSFULLY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21264273 ↗
  • MalfunctionJan 8, 2025Break; Leak/Splash; Break; Leak/Splash

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY PROCEDURE, IT WAS DISCOVERED THAT THE CLEARCANNULA-THREADED 7.0MMX75MM 5PK -ST DEVICE HAD A LEAK IN THE SEALING HOLE OF THE THREADED CANNULA. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE WAS FRACTURED BROKEN (2+ PIECES). IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21103380 ↗
  • MalfunctionMay 22, 2024Mechanical Jam; Mechanical Jam

    IT WAS REPORTED BY THE SALES REP IN JAPAN THAT DURING AN ARCR SURGICAL PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT WHEN THE SURGEON ATTEMPTED TO USE THE CLEAR CANNULA SYSTEM 7.0 X 75 MM ¿ THREADED DEVICE, THE INNER TUBE DID NOT GO IN SMOOTHLY AND GOT STUCK. THE SURGEON TRIED PULLING IT OUT AND REINSERTING IT SEVERAL TIMES; HOWEVER, IT DID NOT IMPROVE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A THIRTY MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19369804 ↗
  • MalfunctionDec 29, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED BY THE AFFILIATE IN RUSSIA THAT THE CLEAR CANNULA SYSTEM 7.0 X 75 MM ¿ THREADED DEVICE WAS NOT WAS NOT PACKED PROPERLY AS IT LET IN AIR; AND THEREFORE WAS NOT STERILE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16075203 ↗
  • MalfunctionFeb 2, 2022Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THERE WAS A SMALL BLACK SPOT INSIDE THE PACKAGE OF THE CLEAR CANNULA-THREADED 7.0MMX75MM 5PK -ST DEVICE. THE PACKAGE WAS NOT OPENED AND ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13426443 ↗
  • MalfunctionSep 24, 2018Break; Break

    IT WAS REPORTED BY THE SALES REPRESENTATIVE THAT DURING A SHOULDER SCOPE PROCEDURE THE PURPLE TROCAR ON THE CUSTOMER'S 7.0 X 75MM FULLY THREADED CANNULA WHEN PULLING OUT FELL ON THE FLOOR AND BROKE. THE SALES REPRESENTATIVE STATED THAT THE DEVICE BROKE OFF BETWEEN THE HANDLE AND STEAM ONTO THE FLOOR. THE SALES REPRESENTATIVE CONFIRMED THAT THE DEVICE DID NOT BREAK IN THE PATIENT. THE SALES REPRESENTATIVE REPORTED THAT THE SURGEON COMPLETED THE PROCEDURE WITH ANOTHER LIKE DEVICE WITH NO PATIENT CONSEQUENCES OR DELAYS. THE SALES REPRESENTATIVE WAS NOT PRESENT FOR THE CASE THEREFORE COULD NOT PROVIDE ANY FURTHER INFORMATION. THE DEVICE WILL BE RETURNING FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7902837 ↗
  • InjuryDec 5, 2017Break; Break

    THE SALES REP REPORTED VIA PHONE THAT AT THE END OF A SHOULDER ARTHROSCOPY THE SURGEON NOTICED THAT THERE WAS A CLEAR PLASTIC PIECE IN THE PATIENTS JOINT SPACE. THE SURGEON LOOKED AT THE 7.0X75MM FULLY THREADED CANNULA THAT WAS USED IN THE CASE AND NOTICED TAHT IT WAS DAMAGED. THE SURGEON WENT BACK INTO THE PATIENT AND USED A SHAVER TO REMOVE THE PLASTIC PIECE. THE SURGEON WAS CONFIDENT THAT HE HAD REMOVED THE PIECE BUT THE FIRST ASSIST STATED THAT THE PIECE WAS STILL IN THE PATIENT. THE SURGEON FELT CONFIDENT IT WAS REMOVED AND CLOSE THE PATIENT UP WITHOUT ANY ISSUES. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7084294 ↗

Other CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices

All CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.