NA 220450Recalls & reported problems
Depuy Mitek, LLC · Cortical Screw 4.5mm x 50mm · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886705007301), received between Apr 19, 2021 and Apr 19, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT AFTER AN UNKNOWN PROCEDURE ON (B)(6) 2020, IT WAS OBSERVED THAT A SCREW DEVICE BROKE INSIDE THE PATIENT'S BODY AND WAS REMOVED. THERE WAS NO SURGICAL DELAY NOR ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11688645 ↗
FDA clearance and classification
Cleared through 510(k) K934366 on Oct 7, 1994 (decision: Substantially Equivalent for Some Indications). Submitted by Mitek Surgical Products, Inc. as “Mitek Cancellous Fixation Screw, Cortical Fixation Screw, Soft Tissue Washer”. The FDA review took 395 days. FDA record ↗
The FDA regulates product code JDW (pin, fixation, threaded) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 220450
Has the NA 220450 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705007301), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 220450?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 19, 2021. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 220450?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 220450 FDA approved?
It was cleared through 510(k) K934366 on Oct 7, 1994. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 220450 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the NA 220450, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand NA, model 220450). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PIN, FIXATION, THREADED devices
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall2 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- APEX 5038-1-080Stryker GmbH · PIN, FIXATION, THREADED48 reports
- Agee Digit Widget® DWD-232Hand Biomechanics Lab Inc · Pin, Fixation, Threaded32 reports
- SPATIAL FRAME 71070720Smith & Nephew, Inc. · PIN, FIXATION, THREADED15 reports
- APEX 5038-2-110Stryker GmbH · PIN, FIXATION, THREADED11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page