Skip to content

NA SD314.230Recalls & reported problems

Synthes GmbH · CANNULATED 4.0MM HEXAGONAL SCREWDRIVER/190MM LONG W/QC · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982004833), received between Dec 11, 2024 and Sep 11, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent6

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED ON AN UNKNOWN DATE, THE LOCKING TOWERS WERE STRIPPED AND THE DEPTH GAUGES WERE BENT. THE FNS BOLT AND SCREWS DIDN¿T LOCK, DIDN¿T BILL. THE SOFT TISSUE REDUCTION DID NOT SNAP IN. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23027695 ↗
  • MalfunctionSep 11, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED ON AN UNKNOWN DATE, THE LOCKING TOWERS WERE STRIPPED AND THE DEPTH GAUGES WERE BENT. THE FNS BOLT AND SCREWS DIDN¿T LOCK, DIDN'T BILL. THE SOFT TISSUE REDUCTION DID NOT SNAP IN. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23027662 ↗
  • MalfunctionSep 11, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED ON AN UNKNOWN DATE, THE LOCKING TOWERS WERE STRIPPED AND THE DEPTH GAUGES WERE BENT. THE FNS BOLT AND SCREWS DIDN¿T LOCK, DIDN¿T BILL. THE SOFT TISSUE REDUCTION DID NOT SNAP IN. THERE WAS NO PATIENT CONSEQUENCES. (B)(4) IS RELATED TO (B)(4). (B)(4) WAS CREATED TO CAPTURE THE ADDITIONAL IPS FROM MOTHER COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23026319 ↗
  • MalfunctionDec 11, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THREE (3) SCREWDRIVERS WERE STRIPPED AFTER POST OP INSPECTION. FOUR (4) TFNA HYBRID HANDLES WERE REQUESTED TO BE REPLACED. FOUR (4) AIMING ARMS WERE REQUESTED TO BE REPLACED AFTER LAG SCREW MISSING NAIL. THERE WAS NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901686 ↗
  • MalfunctionDec 11, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THREE (3) SCREWDRIVERS WERE STRIPPED AFTER POST OP INSPECTION. FOUR (4) TFNA HYBRID HANDLES WERE REQUESTED TO BE REPLACED. FOUR (4) AIMING ARMS WERE REQUESTED TO BE REPLACED AFTER LAG SCREW MISSING NAIL. THERE WAS NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901612 ↗
  • MalfunctionDec 11, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THREE (3) SCREWDRIVERS WERE STRIPPED AFTER POST OP INSPECTION. FOUR (4) TFNA HYBRID HANDLES WERE REQUESTED TO BE REPLACED. FOUR (4) AIMING ARMS WERE REQUESTED TO BE REPLACED AFTER LAG SCREW MISSING NAIL. THERE WAS NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901510 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page