NA 02.212.056Recalls & reported problems
Synthes GmbH · 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886982055217), received between Aug 11, 2022 and Aug 11, 2022 — classified as 10 injuries.
The most frequently reported specific device problem is “Break” (10 reports).
1 other recall cover devices cleared under the same FDA submission (K000684); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break10
Patient problems
- Failure of Implant10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 6 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214548 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 5 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214547 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 8 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214520 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 3 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214514 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 9 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214497 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 10 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214491 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 4 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214473 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 2 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214464 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM THIS IS REPORT 7 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214439 ↗InjuryAug 11, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT SCREWS FROM A LOCKING COMPRESSION (LCP) EXTRAARTICULAR DISTAL HUMERUS PLATING FAILED POSTOPERATIVELY. BASED ON THE X-RAY, IT IS EVIDENT THAT THE SCREW HEADS WERE BROKEN AND THE IMPLANT FAILED. ACCORDING TO SURGEON, THE PATIENT DOES NOT WISH TO UNDERGO A REVISION AT THIS STAGE. THE INITIAL SURGERY WAS PERFORMED SIX WEEKS AGO, AND THE PATIENT WAS DOING WELL. ACCORDING TO THE PATIENT, NO EXTERNAL FORCES EXERTED ON THE IMPLANT CAUSED IT TO MOVE. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) 3.5MM LOCKING SCREW SLF-TPNG WITH STARDRIVE RECESS 56MM. THIS IS REPORT 1 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15214408 ↗
FDA clearance and classification
Cleared through 510(k) K000684 on Apr 28, 2000 (decision: Substantially Equivalent). Submitted by Synthes (Usa) as “Small Fragment Dynamic Compression Locking (Dcl) System”. The FDA review took 59 days. FDA record ↗
The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K000684. One submission often covers many models, so check the product and lot details in each record.
- Z-1800-2015 · May 13, 2015 · Terminated3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated for fixation of fractures, osteotomies, and non-unions of the clavicle,
Common questions about the NA 02.212.056
Has the NA 02.212.056 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982055217), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the NA 02.212.056?
The FDA's MAUDE database lists 10 adverse event reports for this device, received on Aug 11, 2022. In those reports, the most common specific device problem is “Break” (10); the most common patient problem is “Failure of Implant” (10). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 02.212.056?
None of the reports are classified as deaths, 10 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 02.212.056 FDA approved?
It was cleared through 510(k) K000684 on Apr 28, 2000. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 02.212.056 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the NA 02.212.056, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 02.212.056). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page