MULTILOC 03.019.010Recalls & reported problems
Synthes GmbH · EXTRACTION SHAFT F/4.5MM TI MULTILOC SCREWS/QUICK COUPLING · Class 1
All models of MULTILOC (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982069474), received between Apr 16, 2020 and Apr 14, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 14, 2023Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY FOR FRACTURE OF RIGHT HUMERAL SHAFT IN (B)(6) 2022. IN THE SURGERY, THE NAIL IN QUESTION WAS INTENTIONALLY IMPLANTED EXCESSIVELY TO PREVENT IMPINGEMENT, AND A 10 MM END CAP WAS USED. THE PROXIMAL PART OF THE NAIL WAS LOCKED WITH THE 2 SCREWS IN QUESTION, AND THE DISTAL PART WAS LOCKED WITH 3 LOCKING SCREWS. THE POSTOPERATIVE COURSE WAS GOOD, AND THE IMPLANTS USED IN THE INITIAL SURGERY WERE TO BE REMOVED ON (B)(6) 2023. IN THE REMOVAL SURGERY, OPERATIONS WERE PROCEEDED, AND THE ISSUES WERE OCCURRED AS FOLLOWS: AFTER REMOVING THE END CAP, THE SURGEON HAD A DIFFICULTY IN CONNECTING THE EXTRACTION SCREW TO THE NAIL BECAUSE THE DISK-SHAPED OVERHANG AROUND THE CONNECTION PART OF THE EXTRACTION SCREW INTERFERED WITH THE BONE AND SOFT TISSUE AROUND THE EXTRACTION SCREW. THE SURGEON ATTEMPTED TO EXTRACT THE 2 SCREWS IN QUESTION, BUT THE SURGEON HAD A DIFFICULTY IN EXTRACTING THEM DUE TO BONE INGROWTH AND THE DEFORMITY OF THE SCREW HEADS. THE 3 LOCKING SCREWS AT THE DISTAL PART OF THE NAIL WERE EXTRACTED SUCCESSFULLY WITH NO PROBLEM. EVENTUALLY, THE NAIL WAS SUCCEED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16747030 ↗MalfunctionApr 16, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT UNDERWENT A REMOVAL SURGERY WITH THE EXTRACTION SHAFT. DURING THE REMOVAL OF A SCREW THE SURGEON HAD DIFFICULTY IN REMOVING THE SCREW BECAUSE HE COULD NOT ENGAGE THE EXTRACTION SHAFT TO THE SCREW HEAD. THE SURGEON MANAGED TO REMOVE THE SCREW WITH THE EXTRACTION SCREW AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITH A THIRTY (30) MINUTE DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREW (PART#UNKNOWN, LOT#UNKNOWN, QUANTITY 1) THIS REPORT IS FOR ONE (1) EXTRACTION SHAFT F/4.5MM TI MULTILOC SCREWS/QUICK COUPLING. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9968116 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page