NA 03.110.007Recalls & reported problems

Synthes GmbH · STARDRIVE SCREWDRIVER T8 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes StarDrive screwdriver is a surgical screwdriver used with screws. The FDA records hold 97 reports, with no deaths and no injuries; many describe the tip stripping, twisting or breaking, and the handle or parts coming off. Reports fell to 3 in the last 12 months from 11 in the prior 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
106
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 106 adverse event reports that list this exact device (DI 10886982071859), received between Feb 6, 2017 and Sep 17, 2026 — classified as 106 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (43 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 8 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 7 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent43
  • Break33
  • Material Deformation18
  • Device-Device Incompatibility15
  • Contamination /Decontamination Problem4
  • Bent3
  • Crack3
  • Dull, Blunt3

Patient problems

  • No Clinical Signs, Symptoms or Conditions76
  • No Patient Involvement14
  • No Known Impact Or Consequence To Patient7
  • Insufficient Information4
  • No Consequences Or Impact To Patient4
  • No Code Available1

Recent adverse event reports

The 25 most recent of 106 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809135 ↗
  • MalfunctionSep 16, 2026Material Twisted/Bent

    It was reported several instruments were worn.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787732 ↗
  • MalfunctionSep 16, 2026Material Twisted/Bent

    It was reported several instruments were worn.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787311 ↗
  • MalfunctionSep 16, 2026Material Twisted/Bent

    It was reported several instruments were worn.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787301 ↗
  • MalfunctionSep 16, 2026Material Twisted/Bent

    It was reported several instruments were worn.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787278 ↗
  • MalfunctionSep 16, 2026Material Twisted/Bent

    It was reported several instruments were worn.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785380 ↗
  • MalfunctionSep 16, 2026Material Twisted/Bent

    It was reported several instruments were worn.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785360 ↗
  • MalfunctionAug 19, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50326418 ↗
  • MalfunctionAug 14, 2026Break; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50257536 ↗
  • MalfunctionApr 28, 2026Material Twisted/Bent

    IT WAS REPORTED THAT THE SCREWDRIVER TIP WAS STRIPPED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25021301 ↗
  • MalfunctionMar 5, 2026Break; Use of Device Problem

    IT WAS REPORTED THAT THE HANDLE OF THE SCREWDRIVER FROM 2.4 VA DISTAL RADIUS SET WAS BROKEN AND PIECE WAS MISSING. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24519222 ↗
  • MalfunctionNov 13, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT TIP OF DEVICE IS STRIPPED. NO OPERATION INVOLVED. NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23544353 ↗
  • MalfunctionSep 15, 2025Break; Break

    IT WAS REPORTED ON (B)(6) 2025, THE FORCEP'S SCREW FELL OUT, CAUSING THE DEVICE TO SPLIT IN TWO PIECES. ALSO, SILVER CAP ON BACK OF THE SCREWDRIVER FELL OUT. NO OPERATION INVOLVED. NO CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23052569 ↗
  • MalfunctionMay 29, 2025Break; Crack; Break; Crack

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE WOODEN SCREWDRIVER HANDLE WAS CRACKED AND BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22103912 ↗
  • MalfunctionMay 27, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ONE DRIVER HAD A STRIPPED TIP AND THE OTHER ONE WAS MISSING THE SILVER KNOB FROM THE BACK OF THE HANDLE. THERE WAS NO OPERATION INVOLVED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22087066 ↗
  • MalfunctionMay 12, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2025 THE INSTRUMENT TIP WAS TWISTED. DOESN'T WORK WITH SCREWS. NO OPERATION INVOLVED. THERE WAS NO PATIENT CONSEQUENCES. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22002708 ↗
  • MalfunctionMay 12, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON APRIL 18,2025 THE INSTRUMENT TIP WAS TWISTED. DOESN'T WORK WITH SCREWS. NO OPERATION INVOLVED. THERE WAS NO PATIENT CONSEQUENCES. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22002689 ↗
  • MalfunctionApr 1, 2025Break; Break

    IT WAS REPORTED THAT THE SCREWDRIVER CAME INTO THE DECONTAMINATION AREA WITH SEPARATION. IT WAS IDENTIFIED IN THE DECONTAMINATION AREA POST OPERATIVELY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21746219 ↗
  • MalfunctionJan 23, 2025Break; Break

    IT WAS REPORTED THAT THE HANDLE IS BROKEN AND MISSING A PIECE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21218547 ↗
  • MalfunctionJan 17, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS BROKE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21178173 ↗
  • MalfunctionDec 13, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024. THE SCREWDRIVER RECESS STRIPPED DURING EXTRACTION OF SCREW STUCK IN A PLATE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH AN UNKNOWN DELAY.

    Patient problems: Insufficient Information

    MAUDE report 20924368 ↗
  • MalfunctionDec 13, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOW PROCEDURE ON AN UNKNOWN DATE IN 2024. THE SCREWDRIVER RECESS STRIPPED DURING EXTRACTION OF SCREW STUCK IN A PLATE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH AN UNKNOWN DELAY.

    Patient problems: Insufficient Information

    MAUDE report 20924361 ↗
  • MalfunctionDec 4, 2024Break; Break

    IT WAS REPORTED THAT ON NOVEMBER 21, 2024, THE BACK PLATE OF THE DEVICE FALLS OUT. THERE WAS NO PATIENT OR OPERATION INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20843263 ↗
  • MalfunctionSep 27, 2024Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ORTHOPEDIC PROCEDURE ON (B)(6) 2024, THE T8 SCREWDRIVER WAS FAILING TO RETAIN SCREWS. UPON FURTHER INSPECTION, THE TIP HAD DEVELOPED A TWIST. ANOTHER DRIVER WAS OPENED TO COMPLETE THE CASE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. NO PATIENT CONSEQUENCE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20320765 ↗
  • MalfunctionSep 18, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    DURING ROUTINE MEDICAL DEVICE REPROCESSING, MEDICAL DEVICE REPROCESSING STAFF NOTICED THAT THE SHAFT OF THE COMPLAINED ITEM BECAME LOOSE AT THE CONNECTION POINT TO THE REST OF THE INSTRUMENT. ALSO, STAFF IDENTIFIED THAT STAR-DRIVE TIP OF THE ITEM BECAME DEFORMED AND WOULD NO LONGER FUNCTION TO FIT INTO A T8 SCREW HEAD RECESS. A THIRD DISCOVERY WAS MADE BY STAFF THAT THE TIP OF ITEM HAD BECOME STRIPPED AND WOULD NO LONGER PROPERLY ENGAGE THE SCREW HEAD RECESS OF A STAR-DRIVE SCREW. THESE 3 INSTRUMENTS WERE DEEMED TO BE NO LONGER FUNCTIONAL AND WERE SUBSEQUENTLY DISCARDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) STARDRIVE SCREWDRIVER T8. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20246617 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page