NA 03.117.998Recalls & reported problems
Synthes GmbH · STRAIGHT/CURVED POINTED REDUCTION FORCEPS · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982073297), received between Jun 28, 2019 and Feb 5, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Contamination /Decontamination Problem1
- Crack1
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2024Break; Crack; Contamination /Decontamination Problem; Device-Device Incompatibility; Break; Crack; Contamination /Decontamination Problem; Device-Device Incompatibility
IT WAS REPORTED THAT THE DEVICES WERE UNABLE TO BE STERILIZED/ASSEMBLED. THEY WERE CHIPPING AND CRACKING. THEY ARE UNABLE TO BE USED. THIS REPORT IS FOR ONE (1) STRAIGHT/CURVED POINTED REDUCTION FORCEPS THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18641772 ↗MalfunctionJun 28, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT AN ELBOW SURGERY WHEN THE SPIN DOWN NUT GOT STUCK OF THE STRAIGHT/CURVED POINTED REDUCTION FORCEPS ON THREAD. SURGEON HAD TO CUT THE THREAD TO ALLOW THE CLAMP TO BE REMOVED FROM THE PATIENT. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A MINIMAL DELAY OF ONE (1) MINUTE. THERE WERE NO FRAGMENTS GENERATED. PATIENT STATUS WAS UNKNOWN. THE INSTRUMENT WAS RECEIVED BROKEN AT THE SPINDLE, IN ADDITION THE SPINDLE IS BENT PREVENTING THE SPEED NUT FROM ADVANCING. THIS REPORT IS FOR A REDUCTION FORCEPS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8744013 ↗
Other FORCEPS devices
- NA 314.291Synthes GmbH · FORCEPS177 reports
- NA 399.99Synthes GmbH · FORCEPS111 reports
- NA PDL114Synthes GmbH · FORCEPS81 reports
- GEM COUPLER Forceps GEM4183CSynovis Micro Companies Alliance, Inc · FORCEPS1 recall4 reports
- AESCULAP BJ080RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ130RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ132RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ045RAesculap AG · FORCEPS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page