Skip to content

NA 03.306.003Recalls & reported problems

Synthes GmbH · LOWER SIDE ARM FOR PELVIC C-CLAMP II · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982077707), received between Apr 21, 2023 and Apr 17, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility5
  • Break1
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2024Break; Crack; Device-Device Incompatibility; Break; Crack; Device-Device Incompatibility

    IT WAS REPORTED THAT ON UNKNOWN DATE THE ITEM WAS UNABLE TO BE STERILIZED OR ASSEMBLED DUE TO CHIPPING AND CRACKING. IT WAS UNABLE TO USED AND NEEDED REPLACEMENT AS SOON AS POSSIBLE. THIS REPORT IS FOR ONE LOWER SIDE ARM FOR PELVIC C-CLAMP II. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19128333 ↗
  • MalfunctionMay 8, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM FRANCE REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT OVER THE COURSE OF MULTIPLE PROCEDURES, THE SYSTEM DID NOT SLIDE AS USUAL. PROGRESSIVELY WITH USE, A PART OF THE SYSTEM GOT JAMMED. SO THERE WAS A FORCED USE ON ONE SIDE. DURING ONE EVENT, AT THE REMOVAL OF THE SYSTEM, THE SIDE THAT USED TO SLIDE NO LONGER SLIDES. THE CONNECTION BETWEEN THE INTERIOR AND EXTERIOR RAILS HAS BEEN DIFFICULT. THE THREADED SLEEVE GOT STUCK IN THE LOWER ARM. THE METAL INSERT SEPARATED FROM THE LOWER ARM. THE SECOND THREADED SLEEVE ALSO SEIZED UP. TO DATE, THE EQUIPMENT IS UNUSABLE. THE SURGEONS HAD TO FORCE TO RELEASE ALL THE PARTS THAT WERE BLOCKED. A SOLUTION HAD TO BE FOUND TO REMOVE THE MATERIAL BLOCKED IN THE PATIENT'S PELVIS. THE SURGEONS HAVE ADOPTED STRATEGIES ALLOWING THE PERFORMANCE OF THE INTERVENTION WITH NO PATIENT CONSEQUENCE. THIS REPORT IS FOR A LOWER SIDE ARM FOR PELVIC C-CLAMP II. THIS IS REPORT 9 OF 11 FOR (B)(4) .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16889227 ↗
  • MalfunctionMay 8, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM FRANCE REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT OVER THE COURSE OF MULTIPLE PROCEDURES, THE SYSTEM DID NOT SLIDE AS USUAL. PROGRESSIVELY WITH USE, A PART OF THE SYSTEM GOT JAMMED. SO THERE WAS A FORCED USE ON ONE SIDE. DURING ONE EVENT, AT THE REMOVAL OF THE SYSTEM, THE SIDE THAT USED TO SLIDE NO LONGER SLIDES. THE CONNECTION BETWEEN THE INTERIOR AND EXTERIOR RAILS HAS BEEN DIFFICULT. THE THREADED SLEEVE GOT STUCK IN THE LOWER ARM. THE METAL INSERT SEPARATED FROM THE LOWER ARM. THE SECOND THREADED SLEEVE ALSO SEIZED UP. TO DATE, THE EQUIPMENT IS UNUSABLE. THE SURGEONS HAD TO FORCE TO RELEASE ALL THE PARTS THAT WERE BLOCKED. A SOLUTION HAD TO BE FOUND TO REMOVE THE MATERIAL BLOCKED IN THE PATIENT'S PELVIS. THE SURGEONS HAVE ADOPTED STRATEGIES ALLOWING THE PERFORMANCE OF THE INTERVENTION WITH NO PATIENT CONSEQUENCE. THIS REPORT IS FOR A LOWER SIDE ARM FOR PELVIC C-CLAMP II. THIS IS REPORT 1 OF 11 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16887767 ↗
  • MalfunctionApr 21, 2023Device-Device Incompatibility; Device-Device Incompatibility

    THERE WAS NO CHANGE OF SURGICAL TECHNIQUE. THE SYSTEM WAS USED FOR RETAINING BONES DURING PELVIS SURGERIES. AFTER SOME USES, THE SYSTEM DID NOT USUALLY SLIDE AS USUAL. PROGRESSIVELY WITH USES, A PART OF THE SYSTEM WAS JAMMED. SO THERE WAS A FORCED USE ON ONE SIDE. DURING THIS EVENT, AT THE REMOVAL OF THE SYSTEM, THE SIDE THAT USED TO SLIDE NO LONGER. THE SURGEONS HAD ADOPTED STRATEGIES ALLOWING THE PERFORMANCE OF THE INTERVENTION WITH NO PATIENT CONSEQUENCE. THIS IS REPORT 13 OF 13 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16791939 ↗
  • MalfunctionApr 21, 2023Device-Device Incompatibility; Device-Device Incompatibility

    THERE WAS NO CHANGE OF SURGICAL TECHNIQUE. THE SYSTEM WAS USED FOR RETAINING BONES DURING PELVIS SURGERIES. AFTER SOME USES, THE SYSTEM DID NOT USUALLY SLIDE AS USUAL. PROGRESSIVELY WITH USES, A PART OF THE SYSTEM WAS JAMMED. SO THERE WAS A FORCED USE ON ONE SIDE. DURING THIS EVENT, AT THE REMOVAL OF THE SYSTEM, THE SIDE THAT USED TO SLIDE NO LONGER. THE SURGEONS HAD ADOPTED STRATEGIES ALLOWING THE PERFORMANCE OF THE INTERVENTION WITH NO PATIENT CONSEQUENCE. THIS IS REPORT 3 OF 13 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16791458 ↗

Other COMPONENT, TRACTION, INVASIVE devices

All COMPONENT, TRACTION, INVASIVE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page