NA 03.311.032Recalls & reported problems
Synthes GmbH · SMOOTH WIRE/1.8MM DIAMETER 400MM LONG · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982077974), received between Jul 16, 2019 and Dec 12, 2020 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K051600, K092190, K161417); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Consequences Or Impact To Patient1
- Swelling1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE. DURING THE PROCEDURE, A TENSIONER HAD AN UNKNOWN ALLEGATION. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. THE PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS COMPLAINT INVOLVES 1 DEVICE. THIS REPORT IS FOR 1 SMOOTH WIRE/1.8MM DIAMETER 400MM LONG THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10994640 ↗MalfunctionOct 25, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2019, DURING A FOOT SURGERY, THE TWO (2) TENSIONERS WOULD NOT RELEASE REDUCTION WIRE DURING THE APPLICATION OF A DISTRACTION OSTEOGENESIS (DO) RING FIXATOR FRAME. IT CAME APART DURING THE ATTEMPTS TO RELEASE THE WIRE FROM THE TENSIONER INTRAOPERATIVELY. THERE ARE TWO TENSIONERS THAT FAILED DURING THE PROCEDURE. ANOTHER ONE (1) TENSIONER COMES TO A HARD STOP WHEN COMPLETELY RELEASING TENSION. IT IS SUPPOSED TO CLICK AND CONTINUE TO ROTATE WHEN ALL TENSION HAS BEEN RELEASED REACHING ITS ENDPOINT HOWEVER, IT DID NOT ¿CLICK¿ WHEN THE HANDLE WAS FULLY OPENED. THE SAME TYPE OF TENSIONER WAS USED TO COMPLETE THE PROCEDURE. NO ISSUE WITH WIRE. TENSIONER¿S FAILED ON BOTH REDUCTION WIRES AND SMOOTH WIRES. RIGHT LOWER EXTREMITY MULTIPLANAR RING EXTERNAL FIXATOR. WIRE DID NOT APPEAR TO BE THE ISSUE. TENSIONER MALFUNCTION OCCURRED ON SEVERAL DIFFERENT WIRES. TWO OF THE TENSIONERS TERMINALLY IMPINGED ON THE WIRE. IN AN EFFORT TO RELEASE THE TENSIONER FROM THE WIRE THE PHYSICIAN CONTINUED TO TURN THE HANDLE COUNTER CLOCKWISE. TWO OF THE TENSIONERS BECAME DISASSEMBLED. THERE WAS NO SURGICAL DELAY. THE PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCE. CONCOMIT
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9238160 ↗InjuryJul 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT A MAXFRAME ADJUSTMENT. THE SURGEON WANTED TO ADD A FOOT SUPPORT KIT AND CHANGE A WIRE. IT WAS NOTED THAT THE SKIN INCISION WAS NOT RELAXED ENOUGH INITIALLY ALONG WITH INITIAL SWELLING MAY HAVE CAUSED REDNESS DUE TO TENTING OF SKIN. THAT IS WHY THE SMOOTH WIRE WAS REMOVED AND PLACED IT WITH NEW REDUCTION WIRE. WIRE WAS INTACT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. PATIENT STATUS IS UNKNOWN. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available; Swelling
MAUDE report 8793479 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051600, K092190, K161417. One submission often covers many models, so check the product and lot details in each record.
- Z-1833-2014 · Apr 10, 2014 · TerminatedSynthes External Fixation Systems (Small, Medium, Large and DO), bone fixation appliance components.
- Z-1324-2009 · Feb 9, 2009 · TerminatedSquare Nut MR Safe, Catalog Number 03.311.060. The Square Nut MR Safe is used in conjunction with Synthes Distraction Osteogenesis Ring Systems. It can be used to secure componen
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-13040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page