NA 03.607.513Recalls & reported problems

Synthes GmbH · CABLE CUTTER · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes cable cutter is a surgical instrument used to cut cerclage wire during orthopedic procedures. The 35 reports mostly describe the tool breaking or cracking, often mid-operation, with the cutting tip snapping off or the front cutter splitting into pieces, and once failing to cut at all. No deaths or injuries were reported, and most cases describe no patient impact. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10886982080585), received between Jan 25, 2017 and Feb 6, 2026 — classified as 35 malfunctions.

The most frequently reported specific device problem is “Break” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break32
  • Crack2
  • Failure to Cut1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Known Impact Or Consequence To Patient7
  • No Consequences Or Impact To Patient3
  • No Patient Involvement2
  • No Code Available1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2026Break

    IT WAS REPORTED THAT THE CABLE FRONT CUTTER BROKE INTRAOPERATIVELY AND THEY HAD TO USE THE SIDE CUTTER INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24279996 ↗
  • MalfunctionDec 17, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23829089 ↗
  • MalfunctionJun 26, 2025Break; Break

    IT WAS REPORTED WHEN THE CERCLAGE WIRE WAS BEING CUT, THE CUTTING TIP FELL OFF/BROKE OFF. NO ADVERSE PATIENT CONSEQUENCES REPORTED, NO SURGICAL DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336530 ↗
  • MalfunctionDec 9, 2024Break; Crack; Break; Crack

    ADDITIONAL INFORMATION RECEIVED STATES THAT THIS WAS A PRE-OPERATIVE EVENT. THE INSTRUMENT WAS CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20879173 ↗
  • MalfunctionJun 12, 2024Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TIP OF THE FRONT CUTTER IS BROKEN IN TWO PIECES. THERE WAS NO SURGICAL DELAY OR PATIENT HARM REPORTED. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19521360 ↗
  • MalfunctionJun 11, 2024Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2024, THE STERILE TECHNICIAN DISCOVERED THAT THE CABLE CUTTER WAS BROKEN. THIS REPORT INVOLVES ONE (1) CABLE CUTTER. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19508970 ↗
  • MalfunctionMay 28, 2024Break; Break

    IT WAS REPORTED THAT ON MARCH 7, 2024, THE ITEM WAS BROKEN. IT WAS DISCOVERED BY THE STERILE TECHNICIAN. NO FURTHER DETAILS WERE PROVIDED. THIS REPORT IS FOR A CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19406435 ↗
  • MalfunctionApr 22, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT IS DAMAGED/BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19153276 ↗
  • MalfunctionDec 12, 2023Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2023, THE TIP OF THE CABLE CUTTER IN QUESTION BROKE OFF. THERE WAS NO REPORTED PATIENT OR PROCEDURE INTERACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18313151 ↗
  • MalfunctionDec 5, 2023Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ITEM BROKE WHEN IT WAS BEING TESTED PREOPERATIVELY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265138 ↗
  • MalfunctionSep 29, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, A PORTION OF ONE OF THE CUTTER JAWS WAS BROKEN OFF. THIS WAS DISCOVERED DURING LOAN KIT INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT INVOLVES ONE CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15507947 ↗
  • MalfunctionSep 15, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CUTTING PLIERS BROKE. THERE WAS A SURGICAL DELAY OF FIFTEEN (15) MINUTES. THE METAL PARTS WERE REMOVED FROM THE SURGICAL WOUND. THIS REPORT INVOLVES ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15427453 ↗
  • MalfunctionAug 9, 2022Break; Break

    DEVICE REPORT FROM GERMANY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, A PIECE OF THE HEAD OF THE CUTTER BROKE OFF WHILE CUTTING OFF CERCLAGE. THE OPERATION WAS COMPLETED WITHOUT ANY ADVERSE EFFECTS, THERE WAS NO PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15191358 ↗
  • MalfunctionAug 9, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN KOREA, SOUTH AS FOLLOWS: IT WAS REPORTED ON JULY 14, 2022, THE PRODUCT WAS BROKEN IN THE PACKAGE. THIS REPORT IS FOR ONE (1) FRONT CUTTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15191091 ↗
  • MalfunctionAug 3, 2022Crack; Crack

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, CRACK WAS FOUND IN CABLE CUTTER. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15157715 ↗
  • MalfunctionMay 24, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE JAW OF CABLE CUTTER HAS SNAPPED OFF. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER. THIS COMPLAINT INVOLVES ONE(1) DEVICE FRONT CUTTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14491452 ↗
  • MalfunctionMar 24, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT A CABLE SYSTEM FRONT CUTTER BROKE DURING CASE. THIS EVENT INVOLVES ONE (1) DEVICE. THIS IS REPORT 1 OF 1 FOR COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13872239 ↗
  • MalfunctionNov 23, 2021Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6), 2021, THE SURGEON FAILED TO FOLLOW INSTRUCTIONS TO USE THE FRONT CUTTER AND THE INSTRUMENT WAS DAMAGED. A PIECE NEAR THE NOSE OF THE INSTRUMENT CHIPPED OFF. DEVICE FRAGMENTS WERE EASILY REMOVED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR A CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12860892 ↗
  • MalfunctionJun 25, 2021Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, ONE SIDE OF THE JAW AREA HAS SNAPPED OF THE FRONT CUTTER. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. THIS REPORT IS FOR ONE (1). THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12070196 ↗
  • MalfunctionMay 25, 2021Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, THE TIP OF THE CABLE CUTTER BROKE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11879294 ↗
  • MalfunctionMar 18, 2021Break; Break

    IT WAS REPORTED THAT ON UNKNOWN DATE, THE CABLE CUTTER WERE FOUND IN SET TO BE BROKEN. THIS WAS DISCOVERED WHEN THEY OPENED THE SET FOR THE CASE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11516017 ↗
  • MalfunctionMar 18, 2021Break; Break

    IT WAS REPORTED THAT ON UNKNOWN DATE, THE CABLE CUTTER WERE FOUND IN SET TO BE BROKEN. THIS WAS DISCOVERED WHEN THEY OPENED THE SET FOR THE CASE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11516016 ↗
  • MalfunctionSep 29, 2020Failure to Cut; Failure to Cut

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2020, THAT THE PATIENT UNDERWENT FOR AN UNKNOWN SURGERY. DURING THE SURGERY, THE END OF THE CABLE CUTTER WAS WRAPPED OFF AND CABLE DID NOT CUT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH 5 MINUTES DELAY. ALL FRAGMENTS WERE REMOVED WITHOUT ANY ADDITIONAL INTERVENTION. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: UNKNOWN CABLE (PART#: UNKNOWN, LOT#: UNKNOWN, QUANTITY#: 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10601108 ↗
  • MalfunctionAug 20, 2019Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CABLE CUTTER BROKE. THERE WAS NO REPORTED PATIENT CONSEQUENCE. THIS REPORT IS FOR A CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8910360 ↗
  • MalfunctionJan 31, 2019Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIP OF THE FRONT CUTTER BROKE AFTER TWO (2) MONTHS. IT WAS UNKNOWN WHEN WAS THE ISSUE WAS DISCOVERED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) CABLE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8292673 ↗

Other CUTTER, WIRE devices

All CUTTER, WIRE devices →