Skip to content

NA 05.001.123Recalls & reported problems

Synthes GmbH · DRILL/BURR ATTACHMENT FOR 2.35MM SHAFTS · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10886982129383), received between Dec 11, 2017 and Dec 16, 2025 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Excessive Heating” (6 reports).

6 other recalls cover devices cleared under the same FDA submission (K043310); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excessive Heating6
  • Overheating of Device4
  • Complete Loss of Power1
  • Device Operates Differently Than Expected1
  • Device Stops Intermittently1
  • Fail-Safe Did Not Operate1
  • Noise, Audible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient3
  • No Known Impact Or Consequence To Patient2
  • Burn(s)1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT THE DRILL ATTACHMENT AND THE RECIPROCATING SAW ATTACHMENT WERE HEATING UP. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN A SURGICAL PROCEDURE. THERE WAS NO DELAY REPORTED. IT WAS REPORTED THAT THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23822317 ↗
  • MalfunctionDec 28, 2022Noise, Audible; Excessive Heating; Noise, Audible; Excessive Heating

    IT WAS REPORTED FROM THE CZECH REPUBLIC THAT THE ATTACHMENT DEVICE WAS OVERHEATING AND PRODUCED A STRANGE NOISE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16059357 ↗
  • MalfunctionJun 29, 2022Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS IS REPORT 2 OF 2 OF THE SAME EVENT: IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT WHEN THE ELECTRIC PEN DRIVE DEVICE, DRILL/BURR ATTACHMENT DEVICE AND THE CABLE DEVICE WERE PLUGGED IN, THE UNIT STARTED RUNNING. IT WAS REPORTED THAT THERE WAS A SLIGHT DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14852218 ↗
  • MalfunctionAug 9, 2021Excessive Heating; Excessive Heating

    THIS IS EVENT 2 OF 2 OF THE SAME EVENT. IT WAS REPORTED FROM (B)(6) THAT THE ATTACHMENT DEVICES WERE HEATING UP WHILE IN USE WITH THE ELECTRIC PEN DRIVE DEVICE. IT WAS FURTHER REPORTED THAT IT WAS UNKNOWN IF THE ATTACHMENTS WERE CAUSING THE HEAT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12297724 ↗
  • MalfunctionApr 29, 2021Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating

    THIS IS REPORT 2 OF 3 FOR THE SAME EVENT. IT WAS REPORTED FROM THE (B)(6) THAT THE ELECTRIC PEN DRIVE DEVICE STOPPED WHILE IN USE AND OVERHEATED, AND THE DRILL/BURR ATTACHMENT DEVICE AND RECIPROCATING SAW ATTACHMENT DEVICE WERE NOT WORKING AND OVERHEATED. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11748136 ↗
  • MalfunctionMar 16, 2021Excessive Heating; Excessive Heating

    THIS IS REPORT 2 OF 3 OF THE SAME EVENT: IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ELECTRIC PEN DRIVE DEVICE, DRILL/BURR ATTACHMENT DEVICE, CABLE DEVICE, AND RECIPROCATING SAW ATTACHMENT DEVICE HEATED UP WHILE BEING USED TOGETHER. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11499753 ↗
  • MalfunctionJan 31, 2020Excessive Heating; Excessive Heating

    IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE ATTACHMENT DEVICE HAD ABNORMALLY HIGH TEMPERATURE. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF THERE WAS A SPARE DEVICE AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9653407 ↗
  • InjurySep 17, 2019Overheating of Device; Overheating of Device

    THIS IS REPORT 2 OF 4 FOR THE SAME EVENT: IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ¿PLUG¿ WAS STUCK IN THE HANDPIECE DEVICE WHILE BEING USED TOGETHER WITH THE DRILL ATTACHMENT J-LATCH COUPLING DEVICE, DRILL/BURR ATTACHMENT DEVICE, AND CABLE DEVICE. ACCORDING TO THE REPORTER, THE ISSUE WAS OBSERVED SEVERAL TIMES DURING PRE-SURGERY. IT WAS FURTHER REPORTED THAT THE PATIENT HAD TO BE HOSPITALIZED LONGER DUE TO A BURNT LIP (INFECTION AND PAIN) WHICH RESULTED IN SWELLING AND POST-SURGICAL PAIN. THERE WAS A FIFTEEN-MINUTE DELAY TO THE SURGICAL PROCEDURE. A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WAS MEDICAL INTERVENTION AND PROLONGED HOSPITALIZATION ASSOCIATED WITH THIS EVENT. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2019. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Burn(s); Pain; Unspecified Infection

    MAUDE report 9078430 ↗
  • MalfunctionAug 12, 2019Overheating of Device; Overheating of Device

    THIS IS EVENT 2 OF 3 OF THE SAME EVENT. IT WAS REPORTED THAT THE MOTOR DEVICE, BURR ATTACHMENT DEVICE, RECIPROCATING SAW ATTACHMENT DEVICE AND CABLE TO CONSOLE DEVICE ALL OVERHEATED AND SHORTED OUT WHILE BEING USED TOGETHER. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8884144 ↗
  • MalfunctionJun 24, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE DID NOT ROTATE AND BECAME HOT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR WHETHER A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8727288 ↗
  • MalfunctionJul 10, 2018Overheating of Device; Overheating of Device

    IT WAS REPORTED FROM (B)(6) THAT THE ENGINE OF THE DRILL/BURR ATTACHMENT DEVICE WAS BURNT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2018. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7673198 ↗
  • MalfunctionDec 11, 2017Device Stops Intermittently; Device Operates Differently Than Expected; Device Stops Intermittently; Device Operates Differently Than Expected

    THIS IS REPORT 3 OF 5 FOR THE SAME EVENT. IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE ELECTRIC PEN DRIVE DEVICE HAD A BAD CONTACT AND WOULD SOMETIMES STOP WHEN USED WITH THREE ATTACHMENT DEVICES, A HAND SWITCH DEVICE, AND A HANDPIECE DEVICE. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT AS AN UNSPECIFIED SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS REPORTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO PATIENT OR USER INJURIES REPORTED. IT WAS REPORTED THERE WAS NO MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7104068 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K043310. One submission often covers many models, so check the product and lot details in each record.

  • Z-0681-2017 · Nov 1, 2016 · Terminated
    Synthes 0.76 mm Drill Bit/Stryker J-Latch/with 5mm Stop/44.5 mm Non-Sterile
  • Z-2784-2016 · Jun 24, 2016 · Terminated
    Adaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in
  • Z-2783-2016 · Jun 24, 2016 · Terminated
    Light Adaptor for Small Battery Drive and Small Battery Drive II.
  • Z-2199-2016 · May 31, 2016 · Terminated
    Part # 313.250, 1.1mm Drill Bit, Mini QC with 4mm Stop/44.5mm, lot # 4326989, mfg. 9/28/2001 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments ar
  • Z-1313-2015 · Dec 19, 2014 · Terminated
    Synthes Electric Pen Drive (EPD) System, The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticati
  • Z-0615-2014 · Apr 15, 2013 · Terminated
    Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and met

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →