NA 05.001.241Recalls & reported problems

Synthes GmbH · LID FOR RECON SAGITTAL SAW FOR TRAUMA RECON SYSTEM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982130167), received between Feb 1, 2018 and Jul 17, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fail-Safe Did Not Operate” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fail-Safe Did Not Operate4
  • Fail-Safe Problem1
  • Premature Separation1
  • Unintended System Motion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2025Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE LID DEVICE HAD AN UNINTENDED/UNEXPECTED DISENGAGEMENT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FUNCTION OF MODE SWITCH LEVER. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22530872 ↗
  • MalfunctionFeb 14, 2024Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED FROM GERMANY THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE LID FOR RECON SAGITTAL SAW DEVICE MOVING PARTS DID NOT MOVE SMOOTHLY, THE MODE SWITCH WAS STUCK, FAILED LEAK TIGHTNESS TEST AND RUN IN THE LOCKED POSITION. THE DEVICE ALSO FAILED PRETESTS FOR LEAKAGE TEST USING BUBBLE EMISSION TECHNIQUE, CHECK FUNCTION OF MODE SWITCH LEVER, CHECK THE MODE SWITCH LOCKING FUNCTION WITHOUT HANDPIECE, CHECK TORQUE OF MODE SWITCH LEVER, CHECK THE MODE SWITCH LOCKING FUNCTION WITH HANDPIECE, CHECK THE LID LOCKING FUNCTION WITH HANDPIECE AND CHECK IN ¿LOCK¿ MODE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18705408 ↗
  • MalfunctionMay 5, 2021Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LID RECON SAGITTAL SAW DEVICE LOCKING MECHANISM DID NOT FUNCTION, HAD COMPONENT DAMAGE, RAN IN LOCKED POSITION AND THE HOUSING WAS WORN. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK THE LID LOCKING FUNCTION WITH HANDPIECE, CHECK THE MODE SWITCH LOCKING FUNCTION WITH HANDPIECE, CHECK THE MODE SWITCH LOCKING FUNCTION WITHOUT HANDPIECE AND GENERAL CONDITION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11776784 ↗
  • MalfunctionDec 16, 2020Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LID DEVICE LOCKING MECHANISM DID NOT FUNCTION, THE DEVICE RAN IN THE LOCKED POSITION AND LOCKING MECHANISM WAS STIFF/STUCK. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK IN THE LOCK POSITION, CHECK THE LID LOCKING FUNCTION WITH THE HANDPIECE, CHECK THE MODE SWITCH LOCKING FUNCTION WITH THE HANDPIECE AND CHECK THE MODE SWITCH LOCKING FUNCTION WITHOUT HANDPIECE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNSPECIFIED MALFUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11022874 ↗
  • MalfunctionOct 12, 2020Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LID DEVICE RAN IN LOCKED POSITION AND HAD A COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK IN "LOCK" MODE AND CHECK THE FUNCTION WITH POWER MODULE AND HANDPIECE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AND THE SURGERY WAS COMPLETED AS INTENDED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10667901 ↗
  • MalfunctionFeb 1, 2018Unintended System Motion; Fail-Safe Problem; Unintended System Motion; Fail-Safe Problem

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LOCKING MECHANISM WAS DEFECTIVE ON THE LID DEVICE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FUNCTION ¿LOCK/UNLOCK¿ WITH SAFETY BUTTON, CHECK THE LOCKING FUNCTION WITH HANDPIECE AND CHECK THE FUNCTION WITH POWER MODULE AND HANDPIECE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7234874 ↗

Other MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices

All MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices →