Norian 07.704.110SRecalls & reported problems
Synthes GmbH · NORIAN DRILLABLE FAST SET PUTTY 10CC-STERILE · Class 2 · Implant
All models of Norian (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982131652), received between Feb 13, 2018 and Apr 19, 2023 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Chemical Problem” (4 reports).
5 other recalls cover devices cleared under the same FDA submission (K073303, K102722); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Chemical Problem4
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2023Chemical Problem; Chemical Problem
IT WAS REPORTED THAT ON (B)(6) 2023, THE NORIAN FAST SET PUTTY SOLIDIFIED TOO QUICK. IT COULD NOT INSERT THROUGH DELIVERY DEVICE. COMPETITOR'S EQUIPMENT WAS USED. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) NORIAN DRILLABLE FAST SET PUTTY 10CC-S. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16773283 ↗MalfunctionSep 14, 2020Chemical Problem; Chemical Problem
IT WAS REPORTED THAT ON AUGUST 18, 2020, THE SURGEON WANTED TO USE THE NORIAN FAST SET PUTTY FOR AN ANKLE CASE. THEY MIXED THE PUTTY ON THE BACK TABLE FOR NINETY (90) SECONDS. THE SURGEON PUT THE NORIAN IN THE DESIRED LOCATION, BUT AFTER SIX (6) MINUTES IT DID NOT CHANGE CONSISTENCY. THEY DRIPPED WARM SALINE ON THE NORIAN AND LET THE TOURNIQUET DOWN TO TRY AND CREATE A WARM AND WET ENVIRONMENT. AFTER THIRTY (30) MINUTES, THE NORIAN WAS STILL SOFT AND EVENTUALLY WASHED AWAY. THE TEAM SPENT TIME TROUBLESHOOTING TO TRY AND COME UP WITH WAYS TO ENHANCE THE ENVIRONMENT NEEDED FOR THE NORIAN PUTTY. THERE WAS A FORTY FIVE (45) MINUTE SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) NORIAN DRILLABLE FAST SET PUTTY 10 CC STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10530935 ↗MalfunctionAug 3, 2018Chemical Problem; Chemical Problem
IT WAS REPORTED THAT TWO (2) NORIAN DRILLABLE FAST SET PUTTY SOLIDIFIED TOO QUICKLY WHILE USING THE DELIVERY SYSTEM TO DELIVER THE BIOMATERIAL DURING A PROCEDURE ON (B)(6) 2018. THE SURGICAL TEAM WAS NOT ABLE TO GET IT OUT OF THE SYRINGE AS IT BECAME CLOGGED OR BLOCKED. THE ISSUE HAPPENED TWICE. THE PRODUCT WAS REPLACED WITH NORIAN DRILLABLE INJECT TO COMPLETE THE PROCEDURE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY RELATED TO THE INCIDENT REPORTED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: GRAFT DELIVERY DEVICE (PART 03.950.001S, LOT NUMBER UNKNOWN, QUANTITY 2). THIS REPORT IS FOR ONE (1) NORIAN DRILLABLE FAST SET PUTTY 10CC. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7747169 ↗MalfunctionAug 3, 2018Chemical Problem; Chemical Problem
IT WAS REPORTED THAT TWO (2) NORIAN DRILLABLE FAST SET PUTTY SOLIDIFIED TOO QUICKLY WHILE USING THE DELIVERY SYSTEM TO DELIVER THE BIOMATERIAL DURING A PROCEDURE ON (B)(6) 2018. THE SURGICAL TEAM WAS NOT ABLE TO GET IT OUT OF THE SYRINGE AS IT BECAME CLOGGED OR BLOCKED. THE ISSUE HAPPENED TWICE. THE PRODUCT WAS REPLACED WITH NORIAN DRILLABLE INJECT TO COMPLETE THE PROCEDURE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY RELATED TO THE INCIDENT REPORTED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: GRAFT DELIVERY DEVICE (PART 03.950.001S, LOT NUMBER UNKNOWN, QUANTITY 2). THIS REPORT IS FOR ONE (1) NORIAN DRILLABLE FAST SET PUTTY 10CC. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7747151 ↗InjuryFeb 13, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2018, PATIENT UNDERWENT INITIAL SURGERY FOR OPEN REDUCTION INTERNAL FIXATION OF PROXIMAL HUMERUS FRACTURE. DURING THE SURGERY, NORIAN DRILLABLE FAST SET PUTTY 10CC-STERILE WAS PULLED OUT FROM PYXIS SYSTEM AND TEMPERATURE INDICATOR TURNED BLACK. THERE WAS FIVE MINUTES DELAY IN PROCEDURE AS SURGEON HAS TO USE NORIAN DRILLABLE FAST SET PUTTY 3 CC STERILE AND NORIAN DRILLABLE FAST SET PUTTY 5 CC STERILE TO COMPLETE THE PROCEDURE SUCCESSFULLY. SURGEON MENTIONED THAT HE NEEDED MORE NORIAN BUT COMPLETED SURGERY WITH 3CC AND 5CC NORIAN PUTTY. PATIENT WAS REPORTED AS STABLE POST OPERATIVELY. ADDITIONALLY, IT WAS NOTED THAT 3 CC, 5 CC, AND 10 CC NORIAN WERE ALL STORED IN SAME PYXIS SYSTEM AT SAME TEMPERATURE. THIS REPORT IS FOR ONE (1) NORIAN DRILLABLE FAST SET PUTTY 10CC STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7266631 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K073303, K102722. One submission often covers many models, so check the product and lot details in each record.
- Z-0617-2018 · Aug 10, 2017 · TerminatedNorian Drillable Inject
- Z-0618-2018 · Aug 10, 2017 · TerminatedNorian Drillable Inject
- Z-1942-2016 · May 11, 2016 · TerminatedNorian Drillable Inject 3 CC-Sterile; catalog # 07.704.003S Intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony str
- Z-1944-2016 · May 11, 2016 · TerminatedNorian Drillable Inject 10 CC-Sterile; catalog # 07.704.010S Intended for bony voids or defects of the extremities and pelvis that are not instrinsic to the stability of the bony s
- Z-1943-2016 · May 11, 2016 · TerminatedNorian Drillable Inject 5 CC-Sterile; catalog # 07.704.005S Intended for bony voids or defects of the extremities and pelvis that are not instrinsic to the stability of the bony st
Other FILLER, BONE VOID, CALCIUM COMPOUND devices
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44150Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND3 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42275Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42280Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42200Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44145Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- MAGNIFUSE® Bone Graft 7509014Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43102Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- HydroSet XT 897005Stryker Corporation · FILLER, BONE VOID, CALCIUM COMPOUND1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page