Skip to content

NA 202.210Recalls & reported problems

Synthes GmbH · 2.7MM LOCKING SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 10MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982143945), received between Feb 5, 2018 and Oct 23, 2018 — classified as 3 injuries.

6 other recalls cover devices cleared under the same FDA submission (K033995, K050110, K063049, K111540, K113364); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Hypersensitivity/Allergic reaction2
  • Reaction2
  • No Code Available1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 23, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT INITIAL SURGERY ON (B)(6) 2018 TO TREAT AN ANKLE FRACTURE. X-RAY TAKEN ON AN UNKNOWN POSTOPERATIVE DATE REVEALED MALREDUCTION. THE INITIAL PLATE PLACEMENT PUT THE FIBULA IN VALGUS. PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2018 FOR A HARDWARE REMOVAL. THE PLATE WAS REMOVED, AND A 1/3 TUBULAR WAS USED FOR FINAL FIXATION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) 2.7MM LOCKING SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 10MM. THIS IS REPORT 7 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: No Code Available

    MAUDE report 7996613 ↗
  • InjuryFeb 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN (B)(6) 2016, PATIENT UNDERWENT SURGERY TO TREAT A LEFT ANKLE FRACTURE. PATIENT WAS IMPLANTED WITH ONE (1) 2.7MM/3.5MM LCP LATERAL DISTAL PLATE AND TEN (10) SCREWS. IT WAS UNKNOWN TO THE SURGEON IF THE PATIENT WAS ALLERGIC TO METAL AT THE TIME OF THE INITIAL SURGERY. SUBSEQUENTLY THE PATIENT FELT SOME SOFT TISSUE IRRITATION AND THIS WAS DISCOVERED TO BE A REACTION TO METAL ALLERGY. THE SURGEON THEN DECIDED TO REMOVE ALL IMPLANTED DEVICES AND THE CONSTRUCT WITH ALL DEVICES INTACT WAS REMOVED ON (B)(6) 2018. INTRA-OPERATIVE X-RAYS TAKEN SUGGESTED THAT THE FRACTURE HAD COMPLETELY HEALED. THE REMOVAL OF PLATE AND SCREWS CONSTRUCT WAS COMPLETED SUCCESSFULLY AND WITHOUT ANY COMPLICATIONS. NO ADDITIONAL INTERVENTIONS WERE REQUIRED AND THERE WAS NO SURGICAL DELAY. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION. THIS REPORT IS FOR ONE (1) 2.7MM LOCKING SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 10MM. THIS IS REPORT 4 OF 11 FOR (B)(4).

    Patient problems: Hypersensitivity/Allergic reaction; Reaction

    MAUDE report 7242216 ↗
  • InjuryFeb 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN (B)(6) 2016, PATIENT UNDERWENT SURGERY TO TREAT A LEFT ANKLE FRACTURE. PATIENT WAS IMPLANTED WITH ONE (1) 2.7MM/3.5MM LCP LATERAL DISTAL PLATE AND TEN (10) SCREWS. IT WAS UNKNOWN TO THE SURGEON IF THE PATIENT WAS ALLERGIC TO METAL AT THE TIME OF THE INITIAL SURGERY. SUBSEQUENTLY THE PATIENT FELT SOME SOFT TISSUE IRRITATION AND THIS WAS DISCOVERED TO BE A REACTION TO METAL ALLERGY. THE SURGEON THEN DECIDED TO REMOVE ALL IMPLANTED DEVICES AND THE CONSTRUCT WITH ALL DEVICES INTACT WAS REMOVED ON (B)(6) 2018. INTRA-OPERATIVE X-RAYS TAKEN SUGGESTED THAT THE FRACTURE HAD COMPLETELY HEALED. THE REMOVAL OF PLATE AND SCREWS CONSTRUCT WAS COMPLETED SUCCESSFULLY AND WITHOUT ANY COMPLICATIONS. NO ADDITIONAL INTERVENTIONS WERE REQUIRED AND THERE WAS NO SURGICAL DELAY. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION. THIS REPORT IS FOR ONE (1) 2.7MM LOCKING SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 10MM. THIS IS REPORT 10 OF 11 FOR (B)(4).

    Patient problems: Hypersensitivity/Allergic reaction; Reaction

    MAUDE report 7242183 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033995, K050110, K063049, K111540, K113364. One submission often covers many models, so check the product and lot details in each record.

  • Z-2389-2016 · Jun 28, 2016 · Terminated
    Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone,
  • Z-0323-2016 · Nov 2, 2015 · Terminated
    2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone frag
  • Z-0167-2016 · Oct 6, 2015 · Terminated
    2.7 mm Locking Screw SLF-TPNG with T8 Stardrive Recess 12mm. The Synthes modular Mini Fragment LCP System is intended for fixation of fractures, osteotomies, nonunions, replantati
  • Z-1274-2014 · Jan 7, 2014 · Terminated
    Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bon
  • Z-0380-2014 · Jul 31, 2013 · Terminated
    2.4 and 2.7 MM Locking Screw SLF-TPNG with T8 Stardrive Recess 12 MM Fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragmen
  • Z-2278-2013 · Mar 8, 2013 · Terminated
    Synthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus, olecranon, and ulna in adults

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources and updates