LCP 241.266Recalls & reported problems
Synthes GmbH · 3.5MM LCP® POSTEROLATERAL DISTAL HUMERUS PLATE 7H-RIGHT · Class 2 · Implant
All models of LCP (367)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982166272), received between Mar 8, 2018 and Mar 8, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
1 other recall cover devices cleared under the same FDA submission (K033995); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
Patient problems
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 8, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT A PATIENT WAS IMPLANTED WITH A 3.5MM LOCKING COMPRESSION PLATE (LCP) POSTEROLATERAL DISTAL HUMERUS PLATE, 3.5MM CORTEX SCREWS (QUANTITY OF 3), 2.7MM LOCKING SCREWS (QUANTITY OF 2) AND 3.5MM LOCKING SCREWS (QUANTITY OF 4) FOR AN ORIF (OPEN REDUCTION INTERNAL FIXATION) OF A HUMERUS IN (B)(6) 2017. A NONUNION WAS NOTED ON AN UNKNOWN DATE. X-RAYS WERE TAKEN ON AN UNKNOWN DATE AND SHOWED THAT THE (TWO) 2.7MM LOCKING SCREWS WERE NOT CONNECTED TO THE PLATE. BOTH SCREWS WERE STUCK IN THE BONE FULLY INTACT, CAUSING MINIMAL FIXATION. THE X-RAYS ALSO SHOWED THAT THE THREADS ON ONE OF THE 2.7MM LOCKING SCREW HEADS STRIPPED. PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2018. ALL HARDWARE WAS EASILY REMOVED. PATIENT WAS REVISED TO A VARIABLE ANGLE (VA) PLATE, SCREWS AND VIVIGEN FORMABLE BONE GRAFTS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS REPORTED AS STABLE. CONCOMITANT DEVICES REPORTED: 3.5MM CORTEX SCREW, SELF-TAPPING 20MM (PART 204.820, LOT NUMBER UNKNOWN, QUANTITY 1); 3.5MM CORTEX SCREW, SELF-TAPPING 24MM (PART 204.824 LOT NUMBER UNKNOWN, QUANTITY 1); 3.5MM CORTEX SCREW, SELF-TAPPING 28MM (PART 204.828 LOT NUMBER UNKNOWN,
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 7325339 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033995. One submission often covers many models, so check the product and lot details in each record.
- Z-2278-2013 · Mar 8, 2013 · TerminatedSynthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus, olecranon, and ulna in adults
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page