NA 310.19Recalls & reported problems

Synthes GmbH · 2.0MM DRILL BIT/QC/100MM · Class 1

Adverse event reports
21
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 10886982186102), received between Aug 18, 2017 and Dec 10, 2024 — classified as 3 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut15
  • Break6
  • Scratched Material4
  • Entrapment of Device3
  • Activation, Positioning or Separation Problem1
  • Material Fragmentation1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Foreign Body In Patient3
  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2024Failure to Cut; Scratched Material; Failure to Cut; Scratched Material

    IT WAS REPORTED THAT DURING ROUTINE INSPECTION, IT WAS DETERMINED THAT THE INSTRUMENT WAS DULL & GAUGED DUE TO REGULAR USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20895615 ↗
  • MalfunctionDec 6, 2024Failure to Cut; Scratched Material; Failure to Cut; Scratched Material

    IT WAS REPORTED THAT DURING ROUTINE INSPECTION, IT WAS DETERMINED THAT SEVERAL OF THE INSTRUMENTS WERE DULL & GAUGED DUE TO REGULAR USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20872150 ↗
  • MalfunctionDec 6, 2024Failure to Cut; Scratched Material; Failure to Cut; Scratched Material

    IT WAS REPORTED THAT DURING ROUTINE INSPECTION, IT WAS DETERMINED THAT SEVERAL OF THE INSTRUMENTS WERE DULL & GAUGED DUE TO REGULAR USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20871155 ↗
  • MalfunctionDec 6, 2024Failure to Cut; Scratched Material; Failure to Cut; Scratched Material

    IT WAS REPORTED THAT DURING ROUTINE INSPECTION, IT WAS DETERMINED THAT SEVERAL OF THE INSTRUMENTS WERE DULL & GAUGED DUE TO REGULAR USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20870025 ↗
  • MalfunctionMar 18, 2024Break; Entrapment of Device; Activation, Positioning or Separation Problem; Material Integrity Problem; Break; Entrapment of Device; Activation, Positioning or Separation Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, A BROKEN DRILL LEFT IN THE PATIENTS HIP. THIS REPORT IS FOR ONE (1) UNKNOWN DRILL BITS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 18924597 ↗
  • InjuryJul 27, 2023Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2023, A DRILL 2.0MM BIT, BROKE INTRAOPERATIVELY INSIDE PATIENT. PART OF THE DRILL BIT WAS REMOVED AND THE FAR SHAFT OF THE BIT WAS RETAINED INSIDE PATIENT. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE 2.0MM DRILL BIT/QC/100MM FOR (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 17405016 ↗
  • MalfunctionNov 29, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN SURGERY, THE DRILL BIT BROKE. THE DRILL WAS USED AS A ROTATIONAL REFERENCE MARKER. THE SURGEON TRIED PASSING A FLEXIBLE REAMER PAST. THE DRILL BIT BROKE INTO TWO SEGMENTS WHEN THE REAMER ENGAGED. BOTH SEGMENTS WERE RETRIEVED. AND, REAMING ROD PUSHER WAS STUCK WHEN USED TO CLEAR OUT BONE FROM A CANNULATED DRILL. THE ROD PUSHER WAS PULLED OUT WITH FORCE, REVEALING A BENT WIRE SHAFT. THERE WAS A SURGICAL DELAY OF FIFTEEN MINUTES. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12882708 ↗
  • MalfunctionSep 1, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT ALL OF THE DRILL BITS ARE DULL. IT WAS UNKNOWN IF THERE WAS ANY PROCEDURE INVOLVED. THERE WERE NO PATIENT CONSEQUENCES ARE REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR (1)2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 2 OF 10 FOR (B)(4). RELATED PRODUCT COMPLAINT: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12410611 ↗
  • MalfunctionSep 1, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT ALL OF THE DRILL BITS ARE DULL. IT WAS UNKNOWN IF THERE WAS ANY PROCEDURE INVOLVED. THERE WERE NO PATIENT CONSEQUENCES ARE REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR (1)2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 1 OF 10 FOR (B)(4). RELATED PRODUCT COMPLAINT: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12410539 ↗
  • MalfunctionAug 30, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT ALL OF THE DRILL BITS ARE DULL. IT WAS UNKNOWN IF THERE WAS ANY PROCEDURE INVOLVED. THERE WERE NO PATIENT CONSEQUENCES ARE REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES THIRTY (30) DEVICES. THIS REPORT IS FOR (1)2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 20 OF 20 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12389134 ↗
  • MalfunctionAug 30, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT ALL OF THE DRILL BITS ARE DULL. IT WAS UNKNOWN IF THERE WAS ANY PROCEDURE INVOLVED. THERE WERE NO PATIENT CONSEQUENCES ARE REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES 30 DEVICES. THIS REPORT IS FOR (1) 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 19 OF 20 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12389130 ↗
  • MalfunctionAug 30, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT ALL OF THE DRILL BITS ARE DULL. IT WAS UNKNOWN IF THERE WAS ANY PROCEDURE INVOLVED. THERE WERE NO PATIENT CONSEQUENCES ARE REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES 30 DEVICES. THIS REPORT IS FOR (1) 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 4 OF 20 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12388755 ↗
  • MalfunctionJun 7, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ALL THE DRILL BITS ARE DULL, AND NEED TO BE REPLACED. THIS COMPLAINT INVOLVES SIX (6) DEVICES. THIS REPORT IS FOR ONE (1) 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 6 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11948894 ↗
  • MalfunctionJun 4, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ALL THE DRILL BITS ARE DULL, AND NEED TO BE REPLACED. THIS COMPLAINT INVOLVES 1 DEVICES. THIS REPORT IS FOR 1 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 5 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11942608 ↗
  • MalfunctionJun 4, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ALL THE DRILL BITS ARE DULL, AND NEED TO BE REPLACED. THIS COMPLAINT INVOLVES 6 DEVICES. THIS REPORT IS FOR 1 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 4 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11942602 ↗
  • MalfunctionJun 4, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ALL THE DRILL BITS ARE DULL, AND NEED TO BE REPLACED. THIS COMPLAINT INVOLVES 6 DEVICES. THIS REPORT IS FOR 1 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 3 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11942594 ↗
  • MalfunctionJun 4, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ALL THE DRILL BITS ARE DULL, AND NEED TO BE REPLACED. THIS COMPLAINT INVOLVES SIX (6) DEVICES. THIS REPORT IS FOR ONE (1) 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 2 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11942593 ↗
  • MalfunctionJun 4, 2021Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ALL THE DRILL BITS ARE DULL, AND NEED TO BE REPLACED. THIS COMPLAINT INVOLVES SIX (6) DEVICES. THIS REPORT IS FOR ONE (1) 2.0MM DRILL BIT/QC/100MM. THIS REPORT IS 1 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11942578 ↗
  • MalfunctionFeb 21, 2020Break; Break

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE OF THE PROXIMAL HUMERUS ON (B)(6) 2020, THE 2.0MM DRILL BIT BROKE. THERE WERE FRAGMENTS GENERATED WHICH WERE SUCCESSFULLY REMOVED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT STATUS WAS GOOD. THIS REPORT IS FOR ONE (1) 2.0MM DRILL BIT/QC/100MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9740619 ↗
  • InjuryAug 27, 2019Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2019, DURING AN ORTHOPEDIC PROCEDURE, THE DRILL BIT WAS BROKEN AND UNABLE TO BE RETRIEVED. FRAGMENTS WERE GENERATED AND LEFT INSIDE THE PATIENT AND NO MEDICAL INTERVENTION WAS MADE. NO PLANS TO REMOVE THE FRAGMENTS. PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY. PATIENT WAS FINE AND DOING WELL.

    Patient problems: Foreign Body In Patient

    MAUDE report 8937193 ↗
  • InjuryAug 18, 2017Break; Material Fragmentation; Break; Material Fragmentation

    CUSTOMER QUALITY UNIT RECEIVED A MAUDE REPORT FORWARDED FROM DEPARTMENT OF HUMAN SERVICES. ONLY ADDITIONAL AND/OR CORRECTED INFORMATION WILL BE CONTAINED IN THIS REPORT. IT WAS REPORTED THAT A 2.0 MM DRILL BIT/QUICK COUPLING/100 MM BROKE DURING A TISSUE GRAFTING OF THE PATELLA ON (B)(6) 2017. THE TIP OF THE DRILL BIT BROKE OFF IN THE PATELLA AND WAS LEFT IN THE PATIENT. SURGICAL DELAY AND PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) 2.0 MM DRILL BIT. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 6805681 ↗

Other Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment devices

All Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment devices →