NA 313.99Recalls & reported problems
Synthes GmbH · CRUCIFORM SCREWDRIVER WITH HOLDING SLEEVE · Class 1
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886982188328), received between Aug 22, 2019 and Aug 28, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Contamination /Decontamination Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Patient Involvement2
- Foreign Body In Patient1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024 THE SCREW DRIVER BROKE. THERE WAS PATIENT INVOLVEMENT. NO OTHER INFORMATION REPORTED. THIS REPORT IS FOR ONE (1) CRUCIFORM SCREWDRIVER HOLDING SLEEVE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 20089057 ↗MalfunctionApr 5, 2024Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE HANDLES ARE FALLING APART. THE EVENT OCCURRED DURING STERILE PROCESSING, CALIBRATION AND FIELD INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) CRUCIFORM SCREWDRIVER HOLDING SLEEVE THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19054364 ↗MalfunctionJan 6, 2023Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT THE TIP OF A SCREWDRIVER IS BROKEN. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) CRUCIFORM SCREWDRIVER HOLDING SLEEVE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16118201 ↗MalfunctionApr 15, 2022Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CRUCIFORM SCREWDRIVER HANDLE WAS STARTING TO FLAKE. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14120687 ↗MalfunctionAug 25, 2021Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIP OF THE CRUCIFORM SCREWDRIVER HOLDING SLEEVE BROKE. IT IS UNKNOWN WHEN AND WHERE THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE IS PATIENT NOR PROCEDURE INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) CRUCIFORM SCREWDRIVER HOLDING SLEEVE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12367566 ↗MalfunctionMar 30, 2021Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT A HARDWARE REMOVAL OPERATION. THE TWO (2) DRIVERS, SCREWDRIVER SHAFT AND SCREWDRIVER WITH HOLDING SLEEVE HAD BROKEN DURING SCREW REMOVAL CASES, AND THAT A SCREWDRIVER HANDLE WAS NOT QUICK CONNECTING TO DRIVE SHAFTS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY AND NO FRAGMENTS WERE GENERATED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREW (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN). THIS COMPLAINT INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR (1) CRUCIFORM SCREWDRIVER WITH HOLDING SLEEVE. THIS REPORT IS 3 OF 3 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11592218 ↗MalfunctionApr 29, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CRUCIFORM SCREWDRIVER WITH HOLDING SLEEVE WAS FOUND BROKEN IN THE STERILE PROCESSING DEPARTMENT. THERE WAS NO PATIENT NOR PROCEDURE INVOLVEMENT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10012216 ↗MalfunctionSep 25, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT REMOVAL OF FIVE (5) UNKNOWN SCREWS AND AN UNKNOWN PLATE OF THE FIFTH METACARPAL DUE TO PAIN. HOWEVER, THE TIP OF A CRUCIFORM SCREWDRIVER WITH HOLDING SLEEVE BROKE WHEN TRYING TO REMOVE AN UNKNOWN LAST SCREW. THE BROKEN PIECES OF THE SCREWDRIVER WERE REMOVED WITH EASE. MOREOVER, THE SURGEON HAD DIFFICULTY REMOVING THE LAST SCREW BECAUSE THE PLATE AND THE SCREW WERE WELDED TOGETHER, AND THE SCREW WOULD NOT BUDGE. CONSEQUENTLY, THE PART OF THE PLATE AND A SCREW WERE LEFT IN THE PATIENT'S BODY. RESULTING, THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 20 MINUTES DUE TO AN ADDITIONAL TRAY NEEDED AND CUTTING OF THE PLATE. PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART #, LOT # UNKNOWN, QUANTITY# 4). THIS IS REPORT 1 FOR 3 (B)(4).
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient; Pain
MAUDE report 9117067 ↗MalfunctionAug 22, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT ON (B)(6) 2019, AFTER STERILIZATION, IT WAS NOTICED THAT THE FOLLOWING WOOD HANDLE SCREWDRIVERS HAD A LEAKING DYE; ONE (1) CRUCIFORM SCREWDRIVER WITH HOLDING SLEEVE, TEN (10) SMALL HEXAGONAL SCREWDRIVER WITH HOLDING SLEEVE, ONE (1) LARGE SCREWDRIVER, ONE (1) CANNULATED HEXAGONAL SCREWDRIVER, AND FIVE (5) STAR DRIVE SCREWDRIVER. ALL THESE WOOD HANDLE SCREWDRIVERS ARE LEAKING DYE INTO THE TRAYS AS THEY HAVE BEEN THROUGH COUNTLESS STERILIZATIONS AND BEEN IN CIRCULATION FOR MANY YEARS. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT CAPTURES (8) EIGHT OUT OF 18 DEVICES, REMAINING DEVICES ARE CAPTURED UNDER RELATED COMPLAINT (B)(4). THIS REPORT IS FOR ONE (1) CRUCIFORM SCREWDRIVER WITH HOLDING SLEEVE. THIS IS REPORT 1 OF 8 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8920702 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.