NA 314.118Recalls & reported problems
Synthes GmbH · STARDRIVE SCREWDRIVER T25 · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Synthes STARDRIVE screwdriver is used to tighten surgical screws. The 25 reports mostly describe parts breaking or bending, stripped or poorly engaging screwdriver heads, staining after sterilization, and compatibility or contamination issues. The reports mention no deaths and 1 injury; no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 10886982188557), received between Dec 14, 2017 and Sep 15, 2026 — classified as 1 injury, 25 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Material Twisted/Bent6
- Contamination /Decontamination Problem3
- Device-Device Incompatibility3
- Material Deformation3
- Adverse Event Without Identified Device or Use Problem1
- Crack1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Patient Involvement4
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- No Code Available1
- Pain1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Crack; Material Twisted/Bent; Illegible Information
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50780282 ↗MalfunctionJul 27, 2026Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50038702 ↗MalfunctionMar 10, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING PREOP TESTING/ INSPECTION IT WAS FOUND THAT THE ITEM IS NO LONGER FUNCTIONAL, SHAFT PULLS OUT OF HANDLE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24559296 ↗MalfunctionJan 30, 2026Device-Device Incompatibility; Material Deformation; No Apparent Adverse Event
IT WAS REPORTED THAT AFTER A CASE IT WAS NOTICED THAT THE HEAD OF THE SCREWDRIVER IS BENT AND IT IS CAUSING THE SCREWS TO GET STUCK ON THE END. THERE WAS NO SURGICAL DELAY AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24225744 ↗MalfunctionAug 4, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT WHILE INSPECTING THE INSTRUMENTS AFTER GOING THROUGH THE DECONTAMINATION PROCESS, REP NOTICED THAT THERE WERE SOME THAT WERE DAMAGED. REP DOES NOT KNOW WHEN THIS DAMAGE OCCURRED WHETHER IT WAS POSTOP OR PREOP. THERE WAS NO PATIENT INVOLVEMENT. SOME OF THE INSTRUMENTS HAVE TO BE ABLE TO SEND BACK, BUT THERE ARE A COUPLE THAT WILL NOT BE ABLE TO RETURN BECAUSE THEY WERE DISPOSED OF BY THE FACILITY. REP KNOWS THE LOT NUMBER ON SOME OF THEM, BUT THE ONES THAT WERE DISPOSED OF REP DOES NOT KNOW THE LOT NUMBER. 1. PRODUCT # 314.118, LOT # H887026, QUANTITY 1. THIS ITEM IS A T25 SCREWDRIVER, AND THE TIP OF THE SCREWDRIVER IS STRIPPED. THIS ITEM IS AVAILABLE FOR RETURN. 2. PRODUCT # 03.127.010, LOT # 8391544, QUANTITY 1. THIS ITEM IS A DRILL SLEEVE FOR AN AIMING ARM, AND THE TIP OF IT IS BENT AND NEEDS REPLACED. THIS ITEM IS AVAILABLE FOR RETURN. 3. PRODUCT # 319.006, LOT # H873105, QUANTITY 1. THIS ITEM IS A DEPTH GAUGE, AND THE TIP OF THE DEPTH GAUGE IS BROKE OFF. THIS ITEM IS AVAILABLE FOR RETURN. 4. PRODUCT.# 03.333.301, LOT # UNKNOWN, QUANTITY 2. THESE TWO ITEMS ARE POWER SHAFTS T6 AND THE TIPS OF THEM WERE STRIPPED. THESE ITEMS ARE NOT AVAILABLE FOR RETURN DUE TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22709612 ↗MalfunctionJul 9, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE HANDLES ARE LEAVING A BROWN STAIN IN POUCHES AFTER STERILIZATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22460918 ↗MalfunctionJan 31, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SCREWDRIVER HEADS ARE STRIPPED. THEY DO NOT ENGAGE SCREW HEADS PROPERLY ANYMORE. THE REPLACEMENT SCREWDRIVER WAS USED TO PROCEED WITH THE PROCEDURE. THERE WAS A TWO MINUTES OF SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21285329 ↗MalfunctionMar 25, 2024Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED ON MARCH 7, 2024, THAT THE SCREWDRIVER HANDLE IS CHIPPING AWAY WITH STERILIZATION. THERE WERE NO PATIENT OUTCOME AND CONSEQUENCES. THIS REPORT IS FOR ONE (1) SCRDRIVER T25 SELF-HOLDING L290 THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18967664 ↗MalfunctionJul 26, 2023Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE METAL SHAFT OF THE SCREWDRIVER IN QUESTION HAD BROKEN LOOSE FROM THE WOODEN HANDLE. WHEN TIGHTENING A SCREW, THE HANDLE BROKE LOOSE AND WAS UNABLE TO COMPLETE TIGHTENING OF THE SCREW. THE ISSUE WAS FOUND DURING A QUALITY CHECK AFTER THE SET ARRIVED BACK FROM A CASE. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17395850 ↗InjuryJan 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2023 DURING INTRA-OP PATIENT REQUESTED TO HAVE BIL EX TIBIA NAILS, AND SCREWS, REMOVED DUE TO PAIN AND IRRITATION. THE IMPLANTS ON THE RIGHT SIDE WERE REMOVED WITHOUT INCIDENT. THE IMPLANTS ON THE LEFT WERE DIFFICULT. THE END CAP WAS REMOVED, AND NAIL EXTRACTOR THREADED INTO NAIL. THE DISTAL SCREW WAS REMOVED. WHILE ATTEMPTING TO REMOVE PROXIMAL SCREW, THE SCREW HEAD BROKE OFF OF THE SHAFT, AND LOOKED TO BE COLD WELDED INTO THE HEAD OF THE SCREW. THE BROKEN SCREW REMOVAL SET WAS OPENED AND SURGEON ATTEMPTED TO GET SCREW SHAFT OUT OF NAIL/TIBIA. THE BROKEN SCREW REMOVAL INSTRUMENTS WERE UNABLE TO ASSIST IN REMOVING THE SCREW SHAFT. THE SURGEON HAD TO OPEN SKIN MEDIALLY, AND USE AN OSTEOTOME TO CUT BONE AWAY FROM SCREW ON LATERAL AND MEDIAL SIDES. HE WAS THEN ABLE TO TAMP THE SCREW SHAFT OUT OF PATIENT. HE THEN PROCEEDED TO REMOVE THE NAIL. THERE WAS SURGICAL DELAY OF SIXTY (60) MINUTES. ACTION WAS TAKEN OF OPENED ADDITIONAL SETS/INSTRUMENTS, X-RAY WERE TAKEN AND THERE WERE FRAGMENTS GENERATED AND THE FRAGMENTS WERE REMOVED EASILY WITHOUT ANY ADDITIONAL INTERVENTION. SURGERY WAS COMPLETED SUCCESSFULLY. PATIENT CONSEQUENCES ARE UNKNOWN. THIS COMPLAINT IN
Patient problems: Pain; Skin Inflammation/ Irritation
MAUDE report 16247327 ↗MalfunctionSep 19, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON AN UNKNOWN DATE THAT THE SCREWDRIVER WAS NOT SEATING WELL INTO SCREW HEADS. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) SDDRIVE SCREWDRIVER T25 THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15450973 ↗MalfunctionJul 21, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, SCREWDRIVER WOULD NOT FIT PROPERLY INTO 5.0MM SCREW WITH T25 RECESS. ANOTHER T25 SCREWDRIVER WAS OPENED AND SCREWS WERE INSERTED WITHOUT DELAY. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: UNK SCREWS-TRAUMA (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN). THIS REPORT IS FOR ONE (1) SDDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15080409 ↗MalfunctionJan 31, 2022Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE SDDRIVE SCREWDRIVER WAS BROKEN. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) SCRDRIVER T25 SELF-HOLDING L290 THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13392508 ↗MalfunctionJan 10, 2022Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON DECEMBER 13, 2021, FOUR (4) SCREWDRIVERS WERE STAINING THE PLASTIC STERILE PEEL POUCHES. THERE WAS NO PATIENT CONSEQUENCE. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) SDDRIVE(TM) SCREWDRIVER T25. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13204300 ↗MalfunctionJun 29, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING FIELD INSPECTION AND STERILE PROCESSING, ALL THE SIMILI WOOD HANDLE OF THE SCREWDRIVER ARE LEAVING RESIDUE IN THE PEEL PACK. A BROWN SPOT MAKING THE RESPONSIBLE THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR ONE (1) STARDRIVE SCREWDRIVER T25. THIS REPORT IS 4 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12084058 ↗MalfunctionApr 5, 2021Break; Break
VETERINARY COMPLAINT: IT WAS REPORTED THAT ON (B)(6) 2021, DURING A HARDWARE REMOVAL ON A DEAD HORSE, THE STARDRIVE SCREWDRIVER T5 BROKE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. PROCEDURE OUTCOME IS UNKNOWN. NO PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) SD DRIVE SCREWDRIVER T25. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11613904 ↗MalfunctionNov 11, 2020Break; Break
IT WAS REPORTED THAT WHILE INSPECTING BACK UP ITEMS ON AN UNKNOWN DATE, SOME WERE SET ASIDE FOR COMPLAINTS. ONE (1) STARDRIVE SCREWDRIVER TEETH ARE CHIPPED. TWO (2) DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS HAD WIRES LOOSE. ONE (1) STARDRIVE SHAFT TEETH WAS CHIPPED. AND ONE (1) INTERLOCKING SCREWDRIVER IS WARPED AND SHIPPED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10825186 ↗MalfunctionOct 8, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2020, THE FINISH ON HANDLE WAS COMING OFF AND MAKING THE HANDLE NOT SAFE TO STERILIZE. THERE WAS NO REPORTED PATIENT CONSEQUENCE. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10652184 ↗MalfunctionJan 17, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON AN UNKNOWN DATE, TWO (2) SYNTHES MULTI LOC GRAPHIC CASES WERE NOTICED TO HAVE PEELED OFF. THE REPORTED GRAPHIC CASES WERE NOT USED IN SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9603125 ↗MalfunctionNov 15, 2019Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN INSPECTION, A STARDRIVE SCREWDRIVER WAS NOTICED TO BE STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9328446 ↗MalfunctionApr 25, 2019Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE, CUSTOMER REPORTED DISSATISFACTION WITH THE PHENOLIC HANDLES OF THE FOLLOWING DEVICES: ONE (1) STARDRIVE SCREWDRIVER T25, ONE (1) SCREWDRIVER HEXAGONAL WITH HOLDING SLEEVE, ONE (1) SCREWDRIVER STARDRIVE, ONE (1) CANNULATED 4.0MM HEXAGONAL SCREWDRIVER. THE CUSTOMER HAS A MANDATE TO ELIMINATE ALL "WOODEN STYLE" HANDLES FROM THEIR FACILITIES WHEREVER POSSIBLE. UPON INITIAL INSPECTION OF THE RETURNED DEVICES, THESE DEVICES WERE NOTED TO HAVE DEFECTS. THIS REPORT IS FOR ONE (1) STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8551196 ↗MalfunctionOct 3, 2018Break; Break
IT WAS REPORTED THAT ON (B)(6) 2018, THE T25 STARDRIVE SCREWDRIVER WAS NOTICED TO BE LOOSE AND THE SHAFT PORTION DISCONNECTED OR BROKE OFF FROM THE HANDLE. THE ISSUE WAS IDENTIFIED AT THE BACK TABLE WHEN THE SURGICAL TECHNICIAN WAS PREPARING FOR THE CASE WHILE THE OPERATING ROOM (OR) TEAM WAS POSITIONING THE PATIENT. THE PROCEDURE WAS A SIMPLE IMPLANT REMOVAL OF A COUPLE SCREWS. THE PROCEDURE WAS COMPLETED UNEVENTFULLY WITHOUT ANY DELAY. THE ISSUE DID NOT IMPACT THE PROCEDURE OR THE PATIENT. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS REPORT IS FOR ONE (1) STARDRIVE SCREWDRIVER T25. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7929703 ↗MalfunctionSep 14, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2018, THAT THE HANDLE OF THE FOLLOWING SCREWDRIVERS; (2) BALL HEX SCREWDRIVER, (2) FLEXIBLE HEXAGONAL SCREWDRIVER, SCREWDRIVER T25 L380, STARDRIVE SCREWDRIVER T25, HAMMER (2) AND SLIDE HAMMER FOR HAMMER GUIDE (2) LEAVES AN OILY DEPOSIT IN THE BOTTOM OF THE BOXES. THE STERILIZATION EQUINE HAD THE SCREWDRIVERS CLEANED ALONE AND PACK THEM INTO INDIVIDUAL BAGS. THE OILY STAINS ARE REAPPEARED IN A FEW HOURS. NO PATIENT CONSEQUENCE REPORTED. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS 10 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7879032 ↗MalfunctionFeb 23, 2018Dull, Blunt; Dull, Blunt
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE TIP OF ONE (1) STARDRIVE SCREWDRIVER WAS STRIPPED AND WAS NO LONGER FUNCTIONAL. THIS WAS IDENTIFIED BY THE MEDICAL DEVICE REPROCESSING (MDR, LIKE STERILE PROCESSING DEPARTMENT) STAFF DURING THE CLEANING PROCESS. THE MDR STAFF NOTED THAT THE SCREWDRIVER WAS BADLY STRIPPED AND UPON TESTING WITH UNKNOWN QUANTITY OF UNKNOWN SCREWS, IT DID NOT WORK AS INTENDED. IT IS UNKNOWN WHEN THE MALFUNCTION OCCURRED, HOWEVER, AT THE TIME OF DISCOVERING THE FAULT, THERE WAS NO PATIENT OR PROCEDURAL INVOLVEMENT. CONCOMITANT DEVICES REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# UNKNOWN). THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Code Available
MAUDE report 7293756 ↗MalfunctionDec 14, 2017Break; Break
FURTHER IT WAS CLARIFIED THAT THE PATIENT HAD A FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE AND THE PLAN WAS TO REVISE THE PATIENT TO ANOTHER ZIMMER PLATE. HENCE THE REVISION SURGERY WAS CONDUCTED. HE ALSO SAID THAT ALL PARTS THAT WERE IMPLANTED EARLIER WERE INTACT WHEN THE SURGERY BEGAN AND THE SCREWS THAT WERE COLD WELDED WERE THE ONLY ONES THAT BROKE / GOT DAMAGED TO REMOVE THEM DURING THE SURGERY. THIS REPORT ADDRESSES INTRAOPERATIVE ISSUE OF THREE (3) BROKEN SCREWDRIVERS AND THREE COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. THE POSTOPERATIVE REVISION DUE TO PATIENT¿S FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE HAS BEEN CAPTURED UNDER LINKED COMPLAINT (B)(4). CONCOMITANT DEVICES REPORTED: SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 10), LISS PLATE (PART # 422.348, LOT # 2142869, QUANTITY # 1). THIS IS REPORT 2 OF 6 FOR COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7117431 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page