Skip to content

NA 314.67.96Recalls & reported problems

Synthes GmbH · 1.5MM/2.0MM CRUCIFORM SCREWDRIVER BLD W/HOLDING SLV · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10886982188991), received between Feb 21, 2017 and Aug 18, 2025 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Material Twisted/Bent4
  • Break3
  • Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Patient Involvement3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT WHEN OPENING THE MOD HAND FOR A CASE DURING SURGERY, IT WAS NOTED THAT THE CAPTURES HAVE BEEN DAMAGED SO THEY NO LONGER WILL PICKUP THE CRUCIFORM SCREWS. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22826708 ↗
  • MalfunctionAug 4, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREW CATCHER PART IS NOT GRABBING THE SCREWS AND IT DOES NOT ASSEMBLE. IT WAS DEFINITELY BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698139 ↗
  • MalfunctionMay 30, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2024 DURING A PROCEDURE, A HOLDING SLEEVE WAS STUCK IN THE SHAFT OF THE DRIVER AS THE LITTLE LEGS SPREAD AND WE COULD NOT REMOVE IT DURING THE CASE. LEGS ARE NOW DAMAGED TO USE. THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) 1.5MM/2.0MM CRUCIFORM SCREWDRIVER BLD W/HOLDING SLV THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19431071 ↗
  • MalfunctionOct 18, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE SURGEON COMPLAINED THAT THE FOLLOWING SCREWDRIVERS WERE WORN OUT AND NOT RETAINING THE SCREW WELL AND BELIEVES IT TO BE A MANUFACTURING ISSUE AND WANTS THEM REPLACED. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 1.5/2.0 CRUC SCREWDRIVER BLD HOLDING SLV THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17956745 ↗
  • MalfunctionAug 16, 2022Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT THE SCREWDRIVERS ARE STRIPPED. THERE WAS NO PATIENT OUTCOME/CONSEQUENCES. THIS REPORT IS FOR ONE 1.5/2.0 CRUC SCREWDRIVER BLD HOLDING SLV. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15234313 ↗
  • MalfunctionJul 28, 2022Break; Break

    IT WAS REPORTED THE DEPTH GAUGE WAS BROKEN. NO PATIENT CONSEQUENCES WERE REPORTED. THIS REPORT IS FOR A SCREWDRIVER. THIS IS REPORT 2 OF 5 FOR (B)(4). THIS COMPLAINT IS RELATED TO ANOTHER COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15120580 ↗
  • MalfunctionOct 4, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN PROCEDURE, THE SCREW DRIVER SHAFT WOULD NOT BE TAKEN OFF FROM THE HANDLE. THERE WAS NO SURGICAL DELAY. THE SURGERY WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) 1.5MM/2.0MM CRUCIFORM SCREWDRIVER BLD W/HOLDING SLV. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12569625 ↗
  • MalfunctionOct 8, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CRUCIFORM SCREWDRIVER BLADE WAS STUCK TO THE HANDLE WITH MINI QUICK COUPLING. IT WAS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) 1.5MM/2.0MM CRUCIFORM SCREWDRIVER BLD W/HOLDING SLV. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9166074 ↗
  • MalfunctionSep 19, 2019Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON UNKNOWN DATE, DURING ROUTINE CHECK, ALL OF THE SCREWDRIVER SHAFTS IN MODULAR HAND SET WERE STRIPPED AND UNABLE TO PLACE SCREWS. IF ATTEMPTED TO USE IN SURGERY, THESE SHAFTS WOULD BE USELESS. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT # UNKNOWN, QUANTITY# UNKNOWN). THIS IS REPORT 1 FOR 2 (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9087757 ↗
  • MalfunctionJul 1, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING ROUTINE INCOMING INSPECTION, IT WAS DISCOVERED THAT THE SCREWDRIVER BLADE WITH HOLDING SLEEVE WAS BROKEN. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8750341 ↗
  • MalfunctionFeb 21, 2017Bent; Bent

    IT WAS REPORTED THAT WHILE CLEANING A SYNTHES 1.5MM / 2.0MM CRUCIFORM SCREWDRIVER BLADE WITH HOLDING SLEEVE, THE STERILE PROCESSING STAFF WAS UNABLE TO REMOVE THE INNER "GRASPING" PART FROM THE OUTER TUBE. AFTER SEVERAL ATTEMPTS AND WITH THE AID OF A PAIR OF PLIERS, THE STAFF WAS ABLE TO SEPARATE THE PARTS AND REALIZED THAT THE INNER PORTION OF THE DEVICE WAS BENT. THE PIECES CAN NO LONGER BE ASSEMBLED. THERE WAS NO PATIENT INVOLVEMENT OR PROCEDURE ONGOING AND THE ISSUE WAS DISCOVERED OUTSIDE OF THE OPERATING ROOM. THIS COMPLAINT INVOLVES ONE (1) DEVICE: 1.5MM / 2.0MM CRUCIFORM SCREWDRIVER WITH ONE PART DATA. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 6347986 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates