NA 319.091Recalls & reported problems
Synthes GmbH · DEPTH GAUGE FOR 3.5MM CORTEX SCREWS · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Synthes depth gauge is a surgical measuring tool used with 3.5 mm bone screws. The 64 reports mostly describe the tool breaking or deforming during use — hooks snapping off, ball bearings falling out, sleeves and threads wearing down — with a few noting it gives wrong measurements or jams. No deaths and no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 10886982190000), received between Mar 24, 2017 and Aug 13, 2026 — classified as 64 malfunctions.
The most frequently reported specific device problem is “Break” (32 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break32
- Material Deformation29
- Incorrect Measurement3
- Mechanical Jam2
- Use of Device Problem2
- Device Operates Differently Than Expected1
- Device-Device Incompatibility1
- Material Discolored1
Patient problems
- No Clinical Signs, Symptoms or Conditions54
- No Patient Involvement5
- No Consequences Or Impact To Patient3
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2026Incorrect Measurement; Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50237630 ↗MalfunctionApr 6, 2026Use of Device Problem
IT WAS REPORTED THAT ON (B)(6) 2026 THE DEPTH GAUGE WAS FOUND TO BE BROKEN BY THE SPD STAFF. THE BALL BEARING HAS FALLEN OUT (UNABLE TO LOCATE), SO THE OUTER SLEEVE JUST FALLS OFF OF THE DEPTH GAUGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24794770 ↗MalfunctionMar 9, 2026Use of Device Problem
IT WAS REPORTED THAT THE DEPTH GAUGE IS MISSING SMALL SILVER BALL ON SIDE OF DEVICE. THERE WAS NO OPERATION AND PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24545312 ↗MalfunctionJan 26, 2026Break
IT WAS REPORTED MULTIPLE DEVICES ARE NO LONGER FUNCTIONAL. THESE ISSUES DID NOT IMPACT PATIENT OUTCOMES. BELOW DEVICES WERE FOUND TO BE "RELEASE MECHANISM BROKEN" DURING INSPECTION: 03.043.028. BELOW DEVICES WERE FOUND TO BE "STRIPPED THREADS" DURING INSPECTION: 03.120.022, 312.648, 323.029. BELOW DEVICES WERE FOUND TO BE "NUT STUCK ON SLEEVE" DURING INSPECTION: 03.037.017, 03.037.016. BELOW DEVICES WERE FOUND TO BE "HANDLE LEACHING/SPD WIL NOT STERILIZE" DURING INSPECTION: 03.100.018. BELOW DEVICES WERE FOUND TO BE "BROKEN HANDLE" DURING INSPECTION: 311.43. BELOW DEVICES WERE FOUND TO BE "STRIPPED DRILL CONNECTION AND 2ND IS COLD WELDED TO A SCHANZ PIN" DURING INSPECTION: 393.103. BELOW DEVICES WERE FOUND TO BE "BROKEN SLIDING MECHANISM" DURING INSPECTION: 319.091. BELOW DEVICES WERE FOUND TO BE "NOT SNAPPING IN" DURING INSPECTION: 03.120.018, 03.120.017. BELOW DEVICES WERE FOUND TO BE "STRIPPED" DURING INSPECTION: 03.010.151, 314.467, 03.900.042, 03.130.010, 314.453, 03.010.150, 03.010.152. BELOW DEVICES WERE FOUND TO BE "NOT WORKING PROPERLY" DURING INSPECTION: 329.15. BELOW DEVICES WERE FOUND TO BE "NOT LOCKING TENSION BETWEEN TRIGGER PULLS" DURING INSPECTION: 03.221.015. BELOW
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24179477 ↗MalfunctionDec 1, 2025Break; Break
IT WAS REPORTED THAT ONE DEPTH GAUGE BROKE DURING SURGERY AND THE OTHER DID NOT FIT THROUGH DRILL HOLE. NO SURGICAL DELAY. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23680923 ↗MalfunctionAug 8, 2025Break; Break
IT WAS REPORTED THAT THE DEPTH GAUGE HOOK IS BROKEN FROM USAGE IN PELVIS. THE ISSUE IDENTIFIED DURING INSPECTION. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22751311 ↗MalfunctionAug 8, 2025Break; Break
IT WAS REPORTED THAT THE DEPTH GAUGE HOOK IS BROKEN FROM USAGE IN PELVIS. THE ISSUE IDENTIFIED DURING INSPECTION. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22751308 ↗MalfunctionAug 8, 2025Break; Break
THE FOLLOWING PRODUCT ISSUES WERE REPORTED TO HAVE OCCURRED ON AN UNKNOWN DATE: DEPTH GAUGE FOR 3.5MM CORTEX SCREWS (X 3) - HOOK IS BROKEN FROM USAGE IN PELVIS. RATCHET WRENCH-11MM WIDTH ACROSS FLATS (X 3) - RATCHET DOESN¿T WORK. 2.4/2.7 VOLT(TM) SCR RACK 6-90 2 HIGH 1/3 WIDTH ALUM FRAME (X 2) - RACKS ARE FADED FROM WASH AND SOAP. 1.6MM/0.8MM DOUBLE DRILL GUIDE (X 2) - GUIDE IS BENT. 2.0MM/1.1MM DOUBLE DRILL GUIDE (X 2) - GUIDE IS BENT. 2.7MM COMPRESSION/DISTRACTION POST/FOR VA-LOCKING HOLE X 6- THREADS ARE STRIPPED. 2.8MM GUIDE WIRE/ 300MM LENGTH FLUTED TIP (X 6)- THREADS STRIPPED. 2.7MM VOLT(TM) CONDYLAR PLATE 6H/58MM (X 1) - PLATE DID NOT LOCK WITH LOCKING SCREW. 2.7MM VOLT(TM) CONDYLAR PLATE 10H/95MM (X 1) PLATE DID NOT LOCK WITH LOCKING SCREW. 15.5MM MONOBLOC FLEXIBLE REAMER - LONG (X 2) - REAMERS ARE DULL. PERIOSTEAL ELEVATOR 14MM STRAIGHT BLADE-ROUND EDGE (X 3)- RUST ON ELEVATOR. THERE WAS NO PATIENT INVOLVEMENT; ISSUES WERE NOTED IN STERILE PROCESSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22748421 ↗MalfunctionAug 1, 2025Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT BENT/BROKEN DEPTH GAUGES. THERE WAS NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22686686 ↗MalfunctionApr 21, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE 2025, DURING A SURGERY, FIVE PROXIMAL HUMERUS SLEEVES WERE USED. THE DRILL BIT WAS USED THROUGH FIVE SLEEVES AND SEEMED TO BEND OR WARP THE SLEEVES. THE POWER SHAFT NEEDED TO BE REMOVED THROUGH THE SLEEVE. FIVE WERE USED. THE DEPTH GAUGE WAS ALSO USED AND THE HOOK WAS SUPER BENT ON TWO OF THE DEPTH GAUGES. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21873284 ↗MalfunctionApr 21, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE 2025, DURING A SURGERY, FIVE PROXIMAL HUMERUS SLEEVES WERE USED. THE DRILL BIT WAS USED THROUGH FIVE SLEEVES AND SEEMED TO BEND OR WARP THE SLEEVES. THE POWER SHAFT NEEDED TO BE REMOVED THROUGH THE SLEEVE. FIVE WERE USED. THE DEPTH GAUGE WAS ALSO USED AND THE HOOK WAS SUPER BENT ON TWO OF THE DEPTH GAUGES. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21873262 ↗MalfunctionFeb 11, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THERE WAS PITTING/CHIPPING ON THE TIP OF THE MONOBLOC FLEXIBLE REAMER. ALSO, THE DEPTH GAUGE WAS BENT. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21354897 ↗MalfunctionDec 6, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT ORIF SURGERY FOR A PELVIC FRACTURE ON PELVIC WITH THE DEPTH GAUGE IN QUESTION. IN THE SURGERY, THE TIP OF THE PRODUCT BROKE OFF. THE BROKEN END WAS QUICKLY RETRIEVED AND SURGERY RESUMED USING ANOTHER DEPTH GAUGE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. THE SURGEON CONFIRMED BY X-RAY THAT THERE WERE NO PIECES LEFT IN THE PATIENT'S BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20867178 ↗MalfunctionSep 11, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DEPTH GAUGE FROM THE PELVIS SET IS BENT AND DAMAGED. IT WAS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT/CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20189395 ↗MalfunctionJul 10, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON (B)(6) 2024, DURING A TNA, POLYMER SHIFTED, SURGEON TRIED TO REMOVE IT WITH A DEPTH GAUGE. DURING REMOVAL, TWO IM NAIL DEPTH GAUGES AND A PELVIC DEPTH GAUGE WERE EXTREMELY BENT. THE WRENCH WAS ALSO USED TO HIT THE NAIL TO BANG IT OUT BUT IT SCARIFIED THE WRENCH. THERE WAS NO SURGICAL DELAY. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715509 ↗MalfunctionJun 27, 2024Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIPS ON THE DEPTH GAUGE ARE BENT BACKWARDS OR BROKEN. THERE WERE NO PATIENT INVOLVEMENT OR PATIENT STATUS/OUTCOME/CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 3 OF 3 FOR COMPLAINT(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19625036 ↗MalfunctionJun 27, 2024Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIPS ON THE DEPTH GAUGE ARE BENT BACKWARDS OR BROKEN. THERE WERE NO PATIENT INVOLVEMENT OR PATIENT STATUS/OUTCOME/CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 3.5MM CORTEX SCREWS THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624964 ↗MalfunctionJun 27, 2024Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIPS ON THE DEPTH GAUGE ARE BENT BACKWARDS OR BROKEN. THERE WERE NO PATIENT INVOLVEMENT OR PATIENT STATUS/OUTCOME/CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 3.5MM CORTEX SCREWS THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19624958 ↗MalfunctionJun 26, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, DURING A PELVIC PLATING CASE, THE DEPTH GAUGE HOOK WAS NOT CATCHING ON FAR CORTEX BECAUSE THE TIP BROKE OFF. THIS REPORT IS FOR A DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 19614264 ↗MalfunctionMar 22, 2024Break; Incorrect Measurement; Material Deformation; Break; Incorrect Measurement; Material Deformation
DEVICE REPORT FROM CANADA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2024, THE TIPS OF THE DEPTH GAUGES WERE FOUND TO BE BENT AND/OR BROKEN AT THE SHAFT. THE PRODUCTS WERE UNABLE TO GIVE ACCURATE LENGTH READINGS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 2 MINUTES BY USING REPLACEMENT DEPTH GAUGES. NO FRAGMENTS WERE GENERATED, NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. THERE WAS NO ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18960571 ↗MalfunctionJan 31, 2024Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE SALES REP HAD BROKEN INSTRUMENTS. THE SALES REP WAS TRYING TO EXCHANGE ON WARRANTY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 3 OF 5 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18618739 ↗MalfunctionJan 17, 2024Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLUMBIA AS FOLLOWS: IT WAS REPORTED THAT THE 11/8 SLEEVE PROTECTION GETS STUCK IN SURGERY WITH DRILL SL 4.2 AND SYREAM 2.5 L950 DRILL GUIDE AND BENDS IN SURGERY. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) DEPTH GAUGE F/CORTSCR Ø3.5 MEAS-RANGE UP THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18531314 ↗MalfunctionJan 12, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DEPTH GAUGE WAS BENT, AND THE STRIKING CAP THREADS WERE STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 1 OF 1 FOR: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18503312 ↗MalfunctionDec 12, 2023Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PIECE OF DEPTH GAUGE WAS BROKEN OFF, RENDERING IT DIFFICULT TO USE, AND "SLEEVE" OF DEVICE TO SLIP OFF (SLEEVE OF DEVICE THEREFORE WAS LOST). NO PATIENT CONSEQUENCE WAS REPORTED. THIS REPORT INVOLVES ONE (1) DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18310836 ↗MalfunctionDec 5, 2023Break; Break
DEVICE REPORT FROM CANADA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2023, THE PRODUCTS IN QUESTION WERE BROKEN DURING ORTHO CASES. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR A DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 3 OF 3 FOR (B)(4)..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18267659 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 319.091
Has the NA 319.091 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982190000), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 319.091?
The FDA's MAUDE database lists 64 adverse event reports for this device, received between Mar 24, 2017 and Aug 13, 2026. In those reports, the most common specific device problems are “Break” (32), “Material Deformation” (29) and “Incorrect Measurement” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 319.091?
None of the reports are classified as deaths, 0 as injuries and 64 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 319.091 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 319.091, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 319.091). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.09Synthes GmbH · GAUGE, DEPTH63 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record