Synthes GmbH · RATCHET WRENCH-11MM WIDTH ACROSS FLATS · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886982190444), received between Feb 27, 2020 and Dec 1, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 1, 2025Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2025 THE ARTHREX PULLED A LARGE EXFIX SET TO USE AND 2 RATCHETING WRENCHES WERE BROKEN DURING CASE TRYING TO REMOVE AN EXFIX. THEN THE MANUAL WRENCH WAS USED WHICH WAS KEPT AS A BACKUP IN THE HOUSE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO GENERATION OF FRAGMENTS. THERE WAS NO PATIENT CONSEQUENCES BUT A SLIGHT DELAY DURING CASE BUT NOTHING SIGNIFICANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23679043 ↗
  • MalfunctionDec 1, 2025Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2025 THE ARTHREX PULLED A LARGE EXFIX SET TO USE AND 2 RATCHETING WRENCHES WERE BROKEN DURING CASE TRYING TO REMOVE AN EXFIX. THEN THE MANUAL WRENCH WAS USED WHICH WAS KEPT AS A BACKUP IN THE HOUSE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO GENERATION OF FRAGMENTS. THERE WAS NO PATIENT CONSEQUENCES BUT A SLIGHT DELAY DURING CASE BUT NOTHING SIGNIFICANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23679014 ↗
  • MalfunctionFeb 14, 2024Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SALES CONSULTANT FOUND A COUPLE BROKEN INSTRUMENTS. THE SALES ASSOCIATE WAS UNSURE AS TO HOW OR WHEN THE INSTRUMENTS BROKE. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) RATCHET WRENCH-11MM WIDTH ACROSS FLATS. THIS IS REPORT 3 OF 5 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18702810 ↗
  • MalfunctionJul 6, 2023Break; Break

    IT WAS REPORTED THAT ON (B)(6), 2023, THE RATCHET BROKE DURING SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY, AND NO BROKEN FRAGMENTS WERE RETAINED IN THE PATIENT. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE RATCHET WRENCH-11MM WIDTH ACROSS FLATS. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17266011 ↗
  • MalfunctionSep 19, 2022Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SURGEON WAS TIGHTENING EX FIX CLAMP WITH WRENCH WHEN RATCHETING MECHANISM BROKE. USED DIFFERENT WRENCH IN SET. NO FRAGMENTS WERE GENERATED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNK - EX-FIX CLAMPS (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS REPORT IS FOR ONE (1) RATCHET WRENCH-11MM WIDTH ACROSS FLATS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15449325 ↗
  • MalfunctionJun 29, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, RATCHETING WRENCH FOR THE LARGE EXTERNAL FIXATION SET BROKE DURING SURGERY. IT WOULD NOT TIGHTEN THE NUTS ON THE LARGE COMBINATION CLAMPS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT OUTCOME OR CONSEQUENCES. THIS COMPLAINT INVOLVES ONE (1) DEVICE RATCHET WRENCH-11MM WIDTH ACROSS FLATS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14854370 ↗
  • MalfunctionJan 7, 2022Break; Break

    IT WAS REPORTED THAT ON UNKNOWN DATE, PRODUCTS FOUND BROKEN AFTER SURGERY AND ARE BEYOND REPAIR. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES EIGHTEEN (18) DEVICES. THIS REPORT IS FOR (1) RATCHET WRENCH-11MM WIDTH ACROSS FLATS. THIS REPORT IS 10 OF 14 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13196384 ↗
  • MalfunctionMay 25, 2021Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE RATCHET WRENCH AND THE UNIVERSAL BENDING PLIERS WERE BROKEN. IT IS UNKNOWN WHEN AND WHERE THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE IS PATIENT OR PROCEDURE INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICES. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11877930 ↗
  • MalfunctionFeb 27, 2020Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SALES CONSULTANT WAS GIVEN A BROKEN ITEM BY STERILE PROCESSING DEPARTMENT (SPD). THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9765771 ↗

Other WRENCH devices

All WRENCH devices →