NA 391.962Recalls & reported problems
Synthes GmbH · BENDING/CUTTING PLIERS · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Synthes bending/cutting pliers is a surgical tool used to shape and cut implant rods and plates during bone surgery. The 36 reports mostly describe the tool breaking — snapping carbide tips, cracking cutting blades or slots, small pieces chipping off — along with rusting, pitting, and jamming so the jaws no longer open and close. Most involved no patient at all, and no deaths or injuries are reported. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10886982198525), received between Jan 27, 2017 and Mar 23, 2026 — classified as 36 malfunctions.
The most frequently reported specific device problem is “Break” (32 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break32
- Corroded2
- Difficult to Open or Close2
- Failure to Cut2
- Material Fragmentation2
- Mechanical Jam2
- Component Missing1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- No Consequences Or Impact To Patient8
- No Patient Involvement5
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 23, 2026Break
IT WAS REPORTED THERE WERE BROKEN BENDERS, MISSING SCREWS/PLATES FROM VOLT MINI HAVING INADEQUATE STORAGE, AND THE TROCAR/HAMMER CRACKED/TWISTED OFF FROM HINDFOOT NAIL. IT IS UNKNOWN WHEN THE EVENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663860 ↗MalfunctionOct 8, 2025Break; Break
IT WAS REPORTED THAT THE CUTTING SLOT OF THE PLIERS ARE BROKEN. A SMALL PIECE FELL OFF. A REPLACEMENT CUTTER WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23248429 ↗MalfunctionAug 14, 2025Break; Corroded; Break; Corroded
IT WAS REPORTED THAT ON JULY 30, 2025, THERE WAS PITTING, SCORING, CHIPPING ON DEVICES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22803241 ↗MalfunctionApr 24, 2025Break; Break
IT WAS REPORTED THAT THE CHUNK WAS MISSING FROM TIP OF DEVICE. THERE WAS NO OPERATION INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21902613 ↗MalfunctionOct 17, 2024Difficult to Open or Close; Mechanical Jam; Difficult to Open or Close; Mechanical Jam
IT WAS REPORTED DURING A PROCEDURE ON (B)(6) 2024 THE PLIERS DO NOT OPEN AND CLOSE PROPERLY ANYMORE. THEY ARE JAMMED. NO SURGICAL DELAY. PROCEDURE COMPLETED SUCCESSFULLY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20471762 ↗MalfunctionOct 15, 2024Break; Break
IT WAS REPORTED THAT THE HOSPITAL RECEIVED THE DEVICE WITH A BROKEN CARBIDE TIP AND IS UNUSABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20451842 ↗MalfunctionJun 6, 2024Break; Corroded; Break; Corroded
IT WAS REPORTED THAT ON (B)(6) 2024 THE CUTTING BLADES BROKEN, DEVICE IS ALSO PITTING AND RUSTING. THERE WAS NO PATIENT AND OPERATION INVOLVED. THIS REPORT INVOLVES ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19476786 ↗MalfunctionMar 21, 2024Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE BENDING/CUTTING PLIERS WAS BROKE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18953123 ↗MalfunctionFeb 22, 2024Break; Break
IT WAS FURTHER NOTED THE ISSUE WAS DISCOVERED DURING INSPECTION. THE FACILITY IS A VETERINARY HOSPITAL SO THERE WAS NO HUMAN PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18761402 ↗MalfunctionJun 29, 2023Break; Failure to Cut; Break; Failure to Cut
IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2023, THE TIP OF THE INSTRUMENT WAS SLIGHTLY BROKEN. THE CUTTER WOULDN'T CUT, IT JUST BENT THEM. THERE WAS NO DELAY IN SURGERY, AND NO OTHER MEDICAL INTERVENTION WAS REQUIRED. THERE WERE NO BROKEN FRAGMENTS RETAINED IN THE PATIENT¿S BODY. THIS REPORT INVOLVES ONE BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17223950 ↗MalfunctionAug 5, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, THE PLIERS IN QUESTION WERE RECEIVED BY THE STERILE PROCESSING DEPARTMENT WITH A BROKEN CARBON FIBER INSERT (IN THE PLIERS PORTION OF THE TIP). IT IS UNKNOWN HOW OR WHEN THIS ITEM WAS BROKEN). THERE WAS NO REPORTED PATIENT INTERACTION. THIS REPORT IS FOR BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15178542 ↗MalfunctionJun 27, 2022Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE WHILE INSPECTING THE INSTRUMENTS AFTER GOING THROUGH THE DECONTAMINATION PROCESS, THE SALES REP NOTICED THAT THERE WERE SOME THAT WERE DAMAGED AND NEEDED TO BE REPLACED. BENDING PLIERS FOR 2.7MM & 3.5MM PLATES'S BENDING EDGE WAS BENT. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 4 OF 5 FOR COMPLAINT PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14836027 ↗MalfunctionMay 25, 2022Break; Break
IT WAS REPORTED ON (B)(6) 2022, THE BENDING/CUTTING PLIER DEVICE WAS FOUND TO BE BROKEN. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIER DEVICE. THIS COMPLAINT INVOLVES ONE (1) DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14497870 ↗MalfunctionJul 22, 2021Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THE CUTTING BLADE FELL OUT OF THE CUTTING BENDING PLIERS. THIS REPORT IS FOR A CUTTING BENDING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12212164 ↗MalfunctionMay 28, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT ON UNKNOWN DATE, DURING AN INSPECTION, IT WAS DISCOVERED THAT THE SCREWDRIVER HAS TWISTED TIP, PLIERS HAS BENT CUTTING EDGES, AND ADAPTOR HAS DEFORMED QUICK CONNECT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11904380 ↗MalfunctionMay 5, 2021Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE CARBIDE CUTTING/BENDING PLIERS WERE REPORTED TO HAVE THE CARBIDE CHIPPED OFF. THERE WAS NO PATIENT AND SURGICAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11777433 ↗MalfunctionFeb 11, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 IN THE INSPECTION TIME IT WAS DISCOVERED THAT THE POINTED TOOTH OF PLIER HAS CHIPPED OFF. THERE WAS NO SURGERY AND PATIENT INVOLVEMENT REPORTED. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) UNK - PLATES. THIS IS REPORT 1 OF 1 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11315777 ↗MalfunctionFeb 5, 2021Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE BENDING/ CUTTING PLIERS ARE STICKING AND TIP IS BROKEN. IT IS UNKNOWN WHEN WAS THE ISSUE DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. CONCOMITANT PRODUCT REPORTED: BENDING/CUTTING PLIERS: (PART #391.962-US, LOT #A7QA47, QTY UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11285882 ↗MalfunctionJan 21, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 DURING INSPECTION, THE BENDING/CUTTING PLIERS WERE FOUND TO BE BROKEN. THERE WAS NO PATIENT CONSEQUENCE. THERE IS NO FURTHER INFORMATION AVAILABLE. UPDATED EVENT DESCRIPTION: CONCOMITANT DEVICE REPORTED: UNKNOWN PLATE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11208215 ↗MalfunctionDec 28, 2020Failure to Cut; Failure to Cut
IT WAS REPORTED ON (B)(6) 2020, DURING AN INSPECTION, IT WAS FOUND THAT THE BENDING/CUTTING PLIERS FAILED TO CUT THE MAXFRAME WIRE. THERE WAS NO PATIENT INVOLVEMENT. DURING INVESTIGATION OF THE RETURNED DEVICE BY THE MANUFACTURER, IT WAS NOTED THE CARBIDE INSERT BROKEN OFF AT THE CUTTING SLOT OF THE DEVICE AND A RESISTANCE WAS OBSERVED IN OPENING /CLOSING DUE TO FRICTION. CONCOMITANT DEVICE REPORTED: MAXFRAME WIRE (PART UNKNOWN, LOT UNKNOWN, QUANTITY 1). THIS REPORT IS FOR A BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11077999 ↗MalfunctionDec 8, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE BENDING/ CUTTING PLIERS HAVE BROKEN CUTTING SURFACE ON PLATE CUTTING SIDE. IT WAS DISCOVERED BY EQUIPMENT COORDINATOR DURING SET ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10967628 ↗MalfunctionNov 20, 2020Break; Break
IT WAS REPORTED THAT THE BENDING/CUTTING PLIERS WERE FOUND IN THE CENTRAL PROCESSING DEPARTMENT DAMAGED. ONE OF THE CARBON FIBER PIECES IN THE CUTTING PORTION OF THE PLIERS IS MISSING. THERE IS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10880013 ↗MalfunctionSep 8, 2020Break; Difficult to Open or Close; Break; Difficult to Open or Close
IT WAS REPORTED THAT ON (B)(6) 2020, DURING AN UNKNOWN PROCEDURE, THE BENDING/ CUTTING PLIERS FAILED TO OPEN AND CUT AN UNKNOWN MINI FRAGMENT PLATE. AS PER THE REPORTER, THERE WAS A MISSING TOOTH. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING AN ALTERNATIVE UNKNOWN CUTTING DEVICE. THERE WAS NO SURGICAL DELAY REPORTED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN MINI FRAGMENT PLATE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1) THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10502008 ↗MalfunctionJun 12, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING ROUTINE INSPECTION, IT WAS NOTICED THAT THE CUTTING PORTION OF THE BENDING CUTTING PLIERS INSERT BROKE OFF AND WILL NO LONGER CUT. THERE IS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) BENDING/CUTTING PLIERS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10149058 ↗MalfunctionJun 1, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE BENDING CUTTING PLIERS WAS BROKEN. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10106560 ↗
FDA clearance and classification
The FDA regulates product code HTC (pliers, surgical) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 391.962
Has the NA 391.962 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982198525), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 391.962?
The FDA's MAUDE database lists 36 adverse event reports for this device, received between Jan 27, 2017 and Mar 23, 2026. In those reports, the most common specific device problems are “Break” (32), “Corroded” (2) and “Difficult to Open or Close” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 391.962?
None of the reports are classified as deaths, 0 as injuries and 36 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 391.962 FDA approved?
Devices of this type (product code HTC) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 391.962, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 391.962). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PLIERS, SURGICAL devices
- NA 03.130.270Synthes GmbH · PLIERS, SURGICAL34 reports
- NA 388.50Synthes GmbH · PLIERS, SURGICAL23 reports
- Axle Interspinous Fusion System X060-1520Xtant Medical Holdings, Inc. · Pliers, Surgical19 reports
- NA 388.509Synthes GmbH · PLIERS, SURGICAL14 reports
- VA-LCP 03.211.005Synthes GmbH · PLIERS, SURGICAL6 reports
- NA 391.82Synthes GmbH · PLIERS, SURGICAL5 reports
- NA 388.471Synthes GmbH · PLIERS, SURGICAL4 reports
- NA 391.88Synthes GmbH · PLIERS, SURGICAL4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page