Skip to content

NA 394.86Recalls & reported problems

Synthes GmbH · LARGE EX-FIX 11MM CRBN FBR ROD 350MM / MR-CONDITIONAL · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982199911), received between Sep 18, 2018 and Oct 31, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

5 other recalls cover devices cleared under the same FDA submission (K033158, K082650, K122455, K950384); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 4 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Crack4
  • Adverse Event Without Identified Device or Use Problem1
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 31, 2025Break; Crack; Break; Crack

    IT WAS REPORTED ON OCTOBER 8, 2025 THAT A LARGE EX FIX SET WAS INSPECTED AND IT WAS NOTICED MULTIPLE OF THE CARBON FIBER RODS OF SPLINTERED AND CRACKED. IT WAS NOT USED IN SURGERY OR HAD ANY ADVERSE EVENTS SO NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435786 ↗
  • MalfunctionOct 31, 2025Break; Crack; Break; Crack

    IT WAS REPORTED ON OCTOBER 8, 2025 THAT A LARGE EX FIX SET WAS INSPECTED AND IT WAS NOTICED MULTIPLE OF THE CARBON FIBER RODS OF SPLINTERED AND CRACKED. IT WAS NOT USED IN SURGERY OR HAD ANY ADVERSE EVENTS SO NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435718 ↗
  • MalfunctionOct 31, 2025Break; Crack; Break; Crack

    IT WAS REPORTED ON OCTOBER 8, 2025 THAT A LARGE EX FIX SET WAS INSPECTED AND IT WAS NOTICED MULTIPLE OF THE CARBON FIBER RODS OF SPLINTERED AND CRACKED. IT WAS NOT USED IN SURGERY OR HAD ANY ADVERSE EVENTS SO NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435575 ↗
  • MalfunctionOct 31, 2025Break; Crack; Break; Crack

    IT WAS REPORTED ON (B)(6) 2025 THAT A LARGE EX FIX SET WAS INSPECTED AND IT WAS NOTICED MULTIPLE OF THE CARBON FIBER RODS OF SPLINTERED AND CRACKED. IT WAS NOT USED IN SURGERY OR HAD ANY ADVERSE EVENTS SO NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23435249 ↗
  • MalfunctionJul 18, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A LARGE EX FIX REVISION PROCEDURE ON (B)(6) 2024, THE STAFF COULD NOT GET THE CLAMP OFF THE 350MM CARBON FIBER ROD AND USED A NEW BAR AND CLAMP. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO CASE DELAY. THIS REPORT IS FOR AN UNKNOWN LARGE EX-FIX CLAMP. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19772040 ↗
  • InjurySep 18, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE TIBIA ON (B)(6) 2018. THE INITIAL NONUNION WAS CAUSED BY THE SEVERITY OF INJURY AND COMMINUTION OF BONE FRAGMENTS. NO IMPLANTS USED IN FIRST STAGE FIXATION CAUSED NONUNION. FIRST STAGE FIXATION WAS PERFORMED WITH (9)SYNTHES LARGE EXTERNAL FIXATOR ON (B)(6) 2018. BONE GRAFT WAS HARVESTED ON (B)(6) 2018 USING THE RIA SYSTEM OF WHICH THE RIA DRIVE SHAFT IS A PART. THE CHOICE TO HARVEST BONE GRAFT FROM FEMUR TO APPLY IT TO TIBIA TO PROMOTE FRACTURE HEALING. NONE OF THE EXTERNAL FIXATOR PARTS FAILED. NO PARTS ARE AVAILABLE FOR RETURN. THIS COMPLAINT INVOLVES NINE (9) DEVICES. THIS REPORT IS 8 OF 9 FOR (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 7887347 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033158, K082650, K122455, K950384. One submission often covers many models, so check the product and lot details in each record.

  • Z-2451-2015 · Jul 29, 2015 · Terminated
    Synthes Trauma Ex-Fix System (Small, Medium, Distraction Osteogenesis (DO) and Large) ; (S) intended to stabilize and provide treatment for fractures of the small bones, such as th
  • Z-1833-2014 · Apr 10, 2014 · Terminated
    Synthes External Fixation Systems (Small, Medium, Large and DO), bone fixation appliance components.
  • Z-0941-2013 · Jan 11, 2013 · Terminated
    Large Combination Clamp/Large External Fixation System, Part Number 390.005 Product Usage: Synthes Large External Fixation is intended for use to provide treatment for long bone
  • Z-1816-2013 · Oct 12, 2012 · Open, Classified
    Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatme
  • Z-0277-2013 · Sep 28, 2012 · Terminated
    Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation.

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.