NA 398.754Recalls & reported problems
Synthes GmbH · PERCUTANEOUS ARM 255MM · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Synthes percutaneous arm is a surgical forceps component. The 24 reports mostly describe parts that do not fit together, stick and cannot be separated, or break or bend during handling or use. The reports include no deaths or injuries. No recalls name this device, and reports rose from 1 in the previous 12 months to 4 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10886982201584), received between Oct 11, 2018 and Mar 26, 2026 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility20
- Break3
- Contamination /Decontamination Problem1
- Detachment of Device or Device Component1
- Material Deformation1
- Material Twisted/Bent1
- Mechanical Jam1
- Misassembly by Users1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Patient Involvement3
- No Consequences Or Impact To Patient2
- Insufficient Information1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2026Material Deformation; Material Twisted/Bent
IT WAS REPORTED THAT THE DEVICE IS BENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24700465 ↗MalfunctionMar 11, 2026Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE TWO PIECES OF THE CLAMP WERE ASSEMBLED AND ARE NOW STUCK TOGETHER AND CANNOT BE DISSEMBLED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24569161 ↗MalfunctionJan 9, 2026Device-Device Incompatibility; Mechanical Jam
IT WAS REPORTED THAT WHEN VALIDATING THE EQUIPMENT BEFORE SENDING IT TO THE INSTITUTION AND WHEN MOUNTING THE COLLINEAR GRIPPER, IT WAS OBSERVED THAT THE HOOK PRESENTS RESISTANCE WHEN MOUNTED; ONCE MOUNTED, IT IS NOT POSSIBLE TO SEPARATE IT FROM THE CLAMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24019760 ↗MalfunctionDec 8, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED ON UNKNOWN DATE THAT COLINEAR ATTACHMENT BROKEN. WOULD NOT CONNECT TO THE CLAMP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23744801 ↗MalfunctionJun 5, 2025Break; Break
IT WAS REPORTED THAT USER HAD TWO ITEMS THAT BROKE POSTOPERATIVELY. THE ITEMS WON'T CLICK ONTO THE HANDLE ANYMORE. THIS WAS NOTED IN THE STERILE PROCESSING DEPARTMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22153171 ↗MalfunctionAug 20, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, MULTIPLE ITEMS FROM THE SYNTHES COLINEAR CLAMP SET WERE SENT IN FOR REPAIR. IT WAS NOTICED THAT THE ITEMS NEEDED TO BE REPLACED DUE TO THE ITEMS BECOMING STUCK. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20007884 ↗MalfunctionApr 9, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2024, IN A SURGERY ON THE MORNING, THE STAFF USED THE COLLINEAR CLAMP WITH PERCUTANEOUS ARM TO REDUCE A FRACTURE. AFTER THE CASE, THE ARM WOULD NOT COME OFF OF THE COLLINEAR CLAMP HAND PIECE AND WAS DAMAGED DURING ROUTINE ON-LABEL USE OF THE INSTRUMENT. IT WAS UNKNOWN IF THERE WAS ANY PATIENT CONSEQUENCES/OUTCOME. THIS REPORT PERCUTANEOUS ARM 255MM FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 19074029 ↗MalfunctionNov 29, 2023Device-Device Incompatibility; Physical Resistance/Sticking; Device-Device Incompatibility; Physical Resistance/Sticking
DEVICE REPORT FROM CANADA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023 WHILE THE PRODUCT IN QUESTION WAS BEING DEMOED, THE ARM RELEASE BUTTON WOULD NOT DEPRESS. THIS PREVENTED ATTACHMENTS FROM BEING LOADED ONTO THE TRIGGER HANDLE AS INTENDED. THERE WAS NO PATIENT OR PROCEDURE INTERACTION. NO FURTHER INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR A PERCUTANEOUS ARM 255MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18232080 ↗MalfunctionJul 7, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2023, PRIOR TO STARTING A SURGERY, IT WAS NOTICED THAT AN ATTACHMENT FOR THE COLLINEAR CLAMP DID NOT PROPERLY CONNECT. IT NEEDS TO BE REPAIRED OR REPLACED. NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17277194 ↗MalfunctionApr 20, 2023Device-Device Incompatibility; Device-Device Incompatibility
FRACTURE / LOOSENING OF STEM. UNSURE WHY WIRE MISSED, THE PERCUTANEOUS ARM WAS NOT ABLE TO SNAPPED ONTO CLAMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16775907 ↗MalfunctionMar 21, 2023Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2023, THE INSTRUMENTS IN QUESTION WERE FOUND TO BE BROKEN. THE INSTRUMENTS WERE DISCOVERED AFTER SURGERY. NO PATIENT INVOLVEMENT WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A PERCUTANEOUS ARM 255MM. THIS IS REPORT 5 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16586834 ↗MalfunctionJan 4, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2022, THE HOOKS DID NOT STAY COMPRESSED WHEN CLAMPED. THE SURGICAL DELAY, ADDITIONAL INTERVENTION AND PATIENT CONSEQUENCE WAS UNKNOWN. THE SURGERY WAS SUCCESSFULLY COMPLETED. THIS COMPLAINT INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16096798 ↗MalfunctionDec 15, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT THE SURGERY WITH THE PRODUCTS IN QUESTION. DURING THE OPERATION CHECK OF THE EQUIPMENT, THE CONNECTION BETWEEN THE MAIN UNIT AND VARIOUS ARMS OF THE SLIDING REDUCTION CLAMP IN QUESTION COULD NOT BE CONNECTED BECAUSE THE CONNECTION PART OF THE MAIN UNIT AND THE VARIOUS ARMS DID NOT FIT. THE DIAMETER OF THE BODY SIDE OF THE CONNECTION PART AND THE DIAMETER OF THE ARM SIDE DID NOT SEEM TO MATCH IN THE FIRST PLACE. THERE ARE THREE TYPES OF ARMS, BUT ALL THREE HAVE MISMATCHED DIAMETERS AND CANNOT BE CONNECTED TO THE MAIN UNIT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A PERCUTANEOUS ARM 255MM. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15988963 ↗MalfunctionOct 14, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE INSTRUMENTS WERE UNABLE TO BE DISASSEMBLED. THIS REPORT INVOLVES ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15605784 ↗MalfunctionMar 4, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2022, UPON ASSEMBLY OF COLINEAR CLAMP SURGEON COULD NOT GET BUTTON TO UNJAM. PROCEDURE WAS COMPLETED USING ANOTHER REDUCTION CLAMP USED AND WHILE ASSEMBLING THE PULL OUT CONNECTOR BUTTON WAS JAMMED WHICH DID NOT MOVE TO FINISH ASSEMBLY. THIS REPORT IS FOR ONE (1) PERCUT-ARM F/COLLIN REDUC CLAMP. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13668569 ↗MalfunctionAug 16, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PERCUTANEOUS ARM WAS UNABLE TO CONNECT WITH THE HOHMANN-STYLE ARM. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12329310 ↗MalfunctionJun 2, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN IN-SERVICE, THE PERCUTANEOUS ARM BUTTON DOES NOT WORK. TO HOOK IT ON TO THE ACTUAL COLLINEAR CLAMP, THE SLIDING MECHANISM PRESS THE TRIGGER ON THIS ITEM, THE ROD WOULD ADVANCE UNTIL IT NEEDS RESISTANCE AND THEN IT WILL STOP ADVANCING. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) PERCUTANEOUS ARM 255MM. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921156 ↗MalfunctionApr 16, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT IT WAS IMPOSSIBLE TO CONNECT ATTACHMENT TO THE DEVICES. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS REPORT IS FOR ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11681403 ↗MalfunctionNov 24, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICES ARE UNABLE TO CONNECT SMOOTHLY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE PERCUTANEOUS ARM 255MM. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10898821 ↗MalfunctionJun 19, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THAT THE CLAMPS WERE DIFFICULT TO ASSEMBLE AND DISASSEMBLE. THERE WAS NO REPORTED PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) PERCUTANEOUS ARM 255MM. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10173894 ↗MalfunctionJun 3, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020,TWO OF THE ARMS ON THE COLLINEAR REDUCTION CLAMP WERE MISSING THE SPRINGS (NOT NOTED ON THE CHECK SHEETS) THE REDUCTION GUN WAS ALSO NOT ENGAGING THE ARM THAT DID HAVE A SPRING, AND WHEN IT FINALLY DID ENGAGE IT COULDN¿T BE REMOVED EASILY. THIS COMPLAINT INVOLVES (4) DEVICES. THIS REPORT IS FOR ONE (1) PELVIC ARM 225MM THIS REPORT IS 4 OF 4 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10115545 ↗MalfunctionMay 15, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN (B)(6) 2020,DURING ROUTINE INSPECTION, A PERCUTANEOUS ARM AND A HOHMANN-STYLE ARM BUTTONS ON THE SIDE FOR THE RELEASE TO ATTACH TO THE HANDLE DOES NOT FUNCTION.. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE REPORTED: HANDLE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR 1 PERCUTANEOUS ARM 255MM. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10063600 ↗MalfunctionFeb 27, 2020Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2020 WHEN WRAPPING THE SET, THE PELVIC ARM, HOHMANN STYLE ARM AND PERCUTANEOUS ARM ATTACHMENTS WERE STICKING. THE BUTTONS THAT ARE USED TO ATTACH AND DETACH THE SEPARATE ARMS ONTO THE SLIDING MECHANISM STUCK AND ONCE ATTACHED THEY WOULD NOT COME APART EASILY. FORCE WAS NEEDED TO DETACH THEM. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE PERCUTANEOUS ARM 255MM. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9763284 ↗MalfunctionOct 11, 2018Detachment of Device or Device Component; Misassembly by Users; Detachment of Device or Device Component; Misassembly by Users
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, DURING AN UNKNOWN PROCEDURE, THE PELVIC ARM COULD ONLY BE REMOVED BY FORCE AFTER CONNECTING THE COLLINEAR REDUCTION CLAMP SLIDING MECHANISM. IT WAS UNKNOWN IF THERE WAS A PROCEDURE OR PATIENT INVOLVEMENT. CONCOMITANT DEVICES REPORTED: PERCUTANEOUS ARM, 255MM (PART #: 398.754, LOT #: 1515783, QUANTITY: 1), BONE HOOK-SAPE ARM, 206MM (PART #: 398.756, LOT #: 10-9537, QUANTITY: 1) AND HOHMANN-STYLE ARM, 183MM (PART #: 398.752, LOT #: 1515781, QUANTITY: 1). THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS 2 OF 4 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7958417 ↗
FDA clearance and classification
The FDA regulates product code HTD (forceps) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 398.754
Has the NA 398.754 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982201584), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 398.754?
The FDA's MAUDE database lists 24 adverse event reports for this device, received between Oct 11, 2018 and Mar 26, 2026. In those reports, the most common specific device problems are “Device-Device Incompatibility” (20), “Break” (3) and “Contamination /Decontamination Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 398.754?
None of the reports are classified as deaths, 0 as injuries and 24 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 398.754 FDA approved?
Devices of this type (product code HTD) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 398.754, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 398.754). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FORCEPS devices
- NA 314.291Synthes GmbH · FORCEPS176 reports
- NA 399.99Synthes GmbH · FORCEPS111 reports
- NA PDL114Synthes GmbH · FORCEPS81 reports
- GEM COUPLER Forceps GEM4183CSynovis Micro Companies Alliance, Inc · FORCEPS1 recall4 reports
- AESCULAP BJ061RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ096RAesculap AG · FORCEPS1 recall0 reports
- ATRAUMATA FB527RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ016RAesculap AG · FORCEPS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page