Skip to content

NA 399.091Recalls & reported problems

Synthes GmbH · BONE HOLDING FORCEPS-SOFT RATCHET F/PLATES TO 9MM WIDE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886982202314), received between Apr 20, 2017 and May 6, 2024 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Break” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break10
  • Material Twisted/Bent1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient3
  • No Patient Involvement2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2024Break; Break

    THIS REPORT IS FOR ONE (1) BONE HOLDING FORCEPS-SOFT RATCHET F/PLATES TO 9MM WIDE.THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19250958 ↗
  • MalfunctionDec 9, 2021Break

    IT WAS REPORTED THAT FOLLOWING WASH ON AN UNKNOWN DATE, SEVERAL PRODUCTS WERE FOUND TO BE FAULTY. UPON MANUFACTURER INVESTIGATION, THE DEVICE WAS DISCOVERED TO BE BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A BONE HOLDING FORCEPS. THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12970234 ↗
  • MalfunctionApr 19, 2021Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SPRING BLADE INSIDE THE CLAMP BROKE. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) BONE HOLDING FORCEPS-SOFT RATCHET F/PLATES TO 9MM WIDE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11684760 ↗
  • MalfunctionJan 4, 2021Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, THE BONE HOLDING FORCEPS - SOFT RATCHET F/PLATES TO 9MM WIDE WAS BROKEN. THERE WAS NO PATIENT AND SURGICAL INVOLVEMENT. THIS REPORT IS FOR 1 BONE HOLDING FORCEPS-SOFT RATCHET F/PLATES TO 9MM WIDE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11113360 ↗
  • MalfunctionNov 13, 2019Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IN WAS REPORTED THAT A BONE HOLDING FORCEPS WAS BROKEN. NO FURTHER INFORMATION IS AVAILABLE. THIS IS 1 OF 1 FOR REPORT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9317539 ↗
  • MalfunctionJun 28, 2019Break; Break

    THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8743819 ↗
  • MalfunctionFeb 11, 2019Break; Material Twisted/Bent; Mechanical Jam; Break; Material Twisted/Bent; Mechanical Jam

    IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING STERILIZATION AND INSPECTION, THE JAW PART OF A BONE HOLDING FORCEPS WAS FOUND OUT NOT ALIGNING OR CLOSING PROPERLY AND DOES NOT CLAMP DOWN ALL THE WAY. IT WAS NOTED THAT THE INSTRUMENT WAS DIFFICULT TO USE AND THE MECHANISM STICKS. THERE WAS NO PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) BONE HOLDING FORCEPS-SOFT RATCHET F/PLATES TO 9MM WIDE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8325545 ↗
  • MalfunctionJan 24, 2019Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2019, A BONE HOLDING FORCEPS SOFT RATCHET CLAMP WAS BROKEN WHEN ATTEMPTED TO USE . THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) BONE HOLDING FORCEPS-SOFT RATCHET. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8274937 ↗
  • MalfunctionDec 7, 2017Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE TO REPAIR A SPIRAL FRACTURE OF THE HUMERUS, IT WAS NOTED THE BONE HOLDING FORCEPS IN THE SET WAS BROKEN INTO 2 PIECES. ANOTHER FORCEPS WAS READILY AVAILABLE AND USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN SURGERY AND PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT PATIENT HARM. PATIENT WAS REPORTED STABLE POST-OPERATIVELY. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7097094 ↗
  • MalfunctionApr 20, 2017Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2017 THE PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE TIBIA SHAFT TO TREAT A FRACTURE. WHILE THE SURGEON WAS USING THE REDUCTION FORCEPS ONE OF THE HANDLES SNAPPED AND WAS PARTIALLY DETACHED. THAT INSTRUMENT WAS SET ASIDE AND AN ALTERNATE WAS AVAILABLE FOR THE SURGEON TO USE AND SUCCESSFULLY COMPLETE THE PROCEDURE. AFTER THE INSTRUMENT WENT THROUGH STERILIZATION PROCESS IT BROKE INTO (2) SEPARATE PIECES. THERE WAS NO SURGICAL DELAY TO THE PROCEDURE. NO HARM WAS REPORTED TO THE PATIENT. THE PATIENT OUTCOME WAS STABLE. REPORTEDLY, THE INSTRUMENT HAD BEEN IN USE A LONG TIME. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6508344 ↗

Other FORCEPS devices

All FORCEPS devices →