NA 399.092Recalls & reported problems
Synthes GmbH · REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10886982202321), received between Jul 17, 2017 and Aug 19, 2025 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Break” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break16
- Difficult to Open or Close1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Patient Involvement3
- No Code Available1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 19, 2025Break; Break
IT WAS REPORTED THAT THE SOFT LOCK PART OF THE FORCEPS WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22841874 ↗MalfunctionApr 21, 2025Break; Break
IT WAS REPORTED THAT THE ITEM HAS BROKEN ON THE HANDLER WHICH CLOSES THE FORCEPS. SURGERY WAS NOT DELAYED DUE TO THE EVENT AND WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21872688 ↗MalfunctionApr 21, 2025Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE ITEM BROKE ON THE HANDLER WHICH CLOSES THE FORCEPS. SURGERY WAS NOT DELAYED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT CONSEQUENCE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21872633 ↗MalfunctionJul 19, 2023Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN ITALY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE FORCEPS WAS RUPTURED. THE BREAKAGE OCCURRED WHEN THE STERILIZER ACCIDENTALLY DROPPED THE PINCER ONTO THE GROUND. NO ADVERSE EVENT REPORTED. THIS REPORT IS FOR ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17356064 ↗MalfunctionDec 15, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN PORTUGAL AS FOLLOWS: IT WAS REPORTED THAT THE ORTHOKIT ARRIVED AT THE LOCAL WAREHOUSE WITH A DEVICE BROKE. THIS REPORT IS FOR ONE (1) REDUCT-FORCEPS-LRG W/POINT SOFTLO L222 THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15990604 ↗MalfunctionDec 8, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIP PART OF THE PRODUCT WAS BROKEN. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT INVOLVES ONE REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15942191 ↗MalfunctionJul 28, 2022Break; Break
IT WAS REPORTED THE DEPTH GAUGE WAS BROKEN. NO PATIENT CONSEQUENCES WERE REPORTED. THIS REPORT IS FOR REDUCTION FORCEPS. THIS IS REPORT 5 OF 5 FOR (B)(4). THIS COMPLAINT IS RELATED TO ANOTHER COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15120670 ↗MalfunctionFeb 16, 2022Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PRODUCTS BROKE DUE TO USAGE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 9 OF 9 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13543915 ↗MalfunctionFeb 16, 2022Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PRODUCTS BROKE DUE TO USAGE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 8 OF 9 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13543884 ↗MalfunctionFeb 16, 2022Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PRODUCTS BROKE DUE TO USAGE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 6 OF 9 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13543817 ↗MalfunctionApr 7, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE SOFT LOCK PART WAS FOUND BROKEN ON AN UNKNOWN DATE. IT WAS UNKNOWN WHEN AND HOW IT WAS OCCURRED. THERE WAS NO PROCEDURE AND PATIENT INFORMATION ARE REPORTED. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) REDUCTION FORCEPS WITH POINTS. THIS REPORT IS 1 OF 1 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11630238 ↗MalfunctionJul 22, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE FORCEPS LOCK DEVICE WAS BROKEN AND UNABLE TO HOLD. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308457 ↗MalfunctionNov 14, 2019Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING STERILE PROCESSING, THREE (3) FORCEPS DO NOT STAY CLAMPED, ONE (1) FORCEPS HAS A BROKEN TIP, TWO (2) SCREWDRIVERS HAS A BROKEN TIP, ONE (1) SCREWDRIVER HAS A LOOSE BOTTOM PIECE, TWO (2) LARGE QUICK COUPLING AO ADAPTER WAS BROKEN. THERE IS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES SIX (6) DEVICES. THIS REPORT IS FOR ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT IS 4 OF 6 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9325152 ↗MalfunctionJul 15, 2019Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SCREW OF A REDUCTION FORCEPS WITH POINTS FOR FOOT AND ANKLE WAS NOTED BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8790056 ↗MalfunctionAug 13, 2018Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS DISCOVERED ON AN UNKNOWN DATE THAT A RATCHET WAS BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS REPORT IS FOR A REDUCTION FORCEPS WITH POINTS MEDIUM HANDLE-SOFT RATCHET. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7778365 ↗MalfunctionApr 3, 2018Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE HANDLE BROKE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Code Available
MAUDE report 7391802 ↗MalfunctionJul 17, 2017Break; Break
IT WAS REPORTED THAT A REDUCTION FORCEPS WITH POINTS WAS DISCOVERED TO HAVE BROKEN DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE. A TINE HAS BROKEN OFF OF THE FORCEPS (THE DEVICE IS IN TWO PIECES). THERE IS NO ADDITIONAL INFORMATION ANTICIPATED REGARDING THE PROCEDURE OR PATIENT. THIS REPORT IS FOR ONE (1) REDUCTION FORCEPS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6716938 ↗
FDA clearance and classification
The FDA regulates product code HTD (forceps) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 399.092
Has the NA 399.092 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982202321), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 399.092?
The FDA's MAUDE database lists 17 adverse event reports for this device, received between Jul 17, 2017 and Aug 19, 2025. In those reports, the most common specific device problems are “Break” (16) and “Difficult to Open or Close” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 399.092?
None of the reports are classified as deaths, 0 as injuries and 17 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 399.092 FDA approved?
Devices of this type (product code HTD) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 399.092, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 399.092). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FORCEPS devices
- NA 314.291Synthes GmbH · FORCEPS176 reports
- NA 399.99Synthes GmbH · FORCEPS111 reports
- NA PDL114Synthes GmbH · FORCEPS81 reports
- GEM COUPLER Forceps GEM4183CSynovis Micro Companies Alliance, Inc · FORCEPS1 recall4 reports
- AESCULAP BJ061RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ096RAesculap AG · FORCEPS1 recall0 reports
- ATRAUMATA FB527RAesculap AG · FORCEPS1 recall0 reports
- AESCULAP BJ016RAesculap AG · FORCEPS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page