NA 402.208Recalls & reported problems
Synthes GmbH · 2.7MM TI LOCKING SCR SLF-TPNG WITH T8 STARDRIVE RECESS-8MM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982205704), received between Feb 19, 2026 and Feb 19, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K033995, K050110, K113364); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2026Break
IT WAS REPORTED THAT ON (B)(6) 2026, THE PATIENT UNDERWENT AN ORIF FOR DISTAL FIBULA WITH THE SCREWDRIVER. BEFORE THE SURGERY, THE SALES REP WAS INFORMED IN ADVANCE THAT THE SURGEON'S TECHNIQUE WOULD INVOLVE INSERTING SCREWS OF AROUND 10MM WITH A POWER TOOL. DURING THE SURGERY, THE SURGEON ALSO INFORMED THE SALES REP THAT INSERTING ALL SCREWS WITH A POWER TOOL, ATTACH A TORQUE LIMITATION ATTACHMENT TO THE DRIVER HANDLE, AND PERFORM FINAL TIGHTENING BY HAND. A T8 SCREWDRIVER TIP WAS ATTACHED TO THE POWER TOOL AND A 2.7MM F8MM LONG SCREW WAS INSERTED. WHEN ATTEMPTING FINAL TIGHTENING BY HAND, THE SURGEON REALIZED THAT THE SCREWDRIVER TIP HAD BROKEN AND BECOME LODGED IN THE RECESS. AFTER REMOVING THE KW THAT TEMPORARILY FIXED THE PLATE AND REMOVING THE PLATE, IT WAS DISCOVERED THAT THE SCREW HAD ALSO BROKEN UNDERNEATH THE PLATE, SO THE BROKEN SCREW WAS EXTRACTED WITH A RADIO-PEN. AS THE PLATE AND SCREW HAD BECOME ONE, THEY COULD NOT BE REUSED, AND THE SURGERY WAS COMPLETED USING A 5-HOLE POSTEROLATERAL DFP. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24390463 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033995, K050110, K113364. One submission often covers many models, so check the product and lot details in each record.
- Z-2389-2016 · Jun 28, 2016 · TerminatedSynthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone,
- Z-0323-2016 · Nov 2, 2015 · Terminated2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone frag
- Z-1274-2014 · Jan 7, 2014 · TerminatedDrill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bon
- Z-2278-2013 · Mar 8, 2013 · TerminatedSynthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus, olecranon, and ulna in adults
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page